Eletrodo Modificado com [CoO(OH)] Coloidal Aplicado na Detecção de Ácido Oxálico
In this work, we used a chemically modified electrode with cobalt oxyhydroxide in colloidal form as an electrochemical sensor for oxalic acid. Initial experiments were performed in different electrolytes (LiOH, NaOH e KOH), that better defined electrochemical response in medium NaOH solution [0.5 mo...
Main Authors: | , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Universidade Federal de Mato Grosso do Sul
2015-06-01
|
Series: | Orbital: The Electronic Journal of Chemistry |
Subjects: | |
Online Access: | https://periodicos.ufms.br/index.php/orbital/article/view/17879 |
Summary: | In this work, we used a chemically modified electrode with cobalt oxyhydroxide in colloidal form as an electrochemical sensor for oxalic acid. Initial experiments were performed in different electrolytes (LiOH, NaOH e KOH), that better defined electrochemical response in medium NaOH solution [0.5 mol.L-1]. From measurements of cyclic voltammetry and square wave voltammetry for quantification of oxalic acid responses electroanalytical was obtained for concentrations in order 10-5-10-6 mol L-1. The detection limits (2.52x10-5 mol.L-1) and quantification (7.55x10-6 mol.L-1) were significantly better for measurements performed by square wave voltammetry. The CME-Co has potential application as a sensor for the measurement of oxalic acid and redox species.
DOI: http://dx.doi.org/10.17807/orbital.v7i2.572
|
---|---|