Quality assurance procedures based on dosimetric, gamma analysis as a fast reliable tool for commissioning brachytherapy treatment planning systems

Fast and easily repeatable methods for commissioning procedures for brachytherapy (BT) treatment planning systems (TPS) are needed. Radiochromic film dosimetry with gamma analysis is widely used in external beam quality assurance (QA) procedures and planar film dosimetry is also increasingly used fo...

Full description

Bibliographic Details
Main Authors: Zwierzchowski Grzegorz, Bieleda Grzegorz, Skowronek Janusz
Format: Article
Language:English
Published: Sciendo 2017-11-01
Series:Radiology and Oncology
Subjects:
Online Access:https://doi.org/10.1515/raon-2017-0050
_version_ 1818692660436664320
author Zwierzchowski Grzegorz
Bieleda Grzegorz
Skowronek Janusz
author_facet Zwierzchowski Grzegorz
Bieleda Grzegorz
Skowronek Janusz
author_sort Zwierzchowski Grzegorz
collection DOAJ
description Fast and easily repeatable methods for commissioning procedures for brachytherapy (BT) treatment planning systems (TPS) are needed. Radiochromic film dosimetry with gamma analysis is widely used in external beam quality assurance (QA) procedures and planar film dosimetry is also increasingly used for verification of the dose distribution in BT applications. Using the gamma analysis method for comparing calculated and measured dose data could be used for commissioning procedures of the newly developed TG-186 and MBDCA calculation algorithms. The aim of this study was dosimetric verification of the calculation algorithm used in TPS when the CT/MRI ring applicator is used.
first_indexed 2024-12-17T13:01:19Z
format Article
id doaj.art-649fa7e07e67479fa39f37d771ef4b77
institution Directory Open Access Journal
issn 1581-3207
language English
last_indexed 2024-12-17T13:01:19Z
publishDate 2017-11-01
publisher Sciendo
record_format Article
series Radiology and Oncology
spelling doaj.art-649fa7e07e67479fa39f37d771ef4b772022-12-21T21:47:21ZengSciendoRadiology and Oncology1581-32072017-11-0151446947410.1515/raon-2017-0050raon-2017-0050Quality assurance procedures based on dosimetric, gamma analysis as a fast reliable tool for commissioning brachytherapy treatment planning systemsZwierzchowski Grzegorz0Bieleda Grzegorz1Skowronek Janusz2Poznan University of Medical Sciences, Faculty of Health Sciences, Poznana, PolandPoznan University of Medical Sciences, Faculty of Health Sciences, Poznana, PolandPoznan University of Medical Sciences, Faculty of Health Sciences, Poznana, PolandFast and easily repeatable methods for commissioning procedures for brachytherapy (BT) treatment planning systems (TPS) are needed. Radiochromic film dosimetry with gamma analysis is widely used in external beam quality assurance (QA) procedures and planar film dosimetry is also increasingly used for verification of the dose distribution in BT applications. Using the gamma analysis method for comparing calculated and measured dose data could be used for commissioning procedures of the newly developed TG-186 and MBDCA calculation algorithms. The aim of this study was dosimetric verification of the calculation algorithm used in TPS when the CT/MRI ring applicator is used.https://doi.org/10.1515/raon-2017-0050brachytherapyquality assurancefilm dosimetryring applicator
spellingShingle Zwierzchowski Grzegorz
Bieleda Grzegorz
Skowronek Janusz
Quality assurance procedures based on dosimetric, gamma analysis as a fast reliable tool for commissioning brachytherapy treatment planning systems
Radiology and Oncology
brachytherapy
quality assurance
film dosimetry
ring applicator
title Quality assurance procedures based on dosimetric, gamma analysis as a fast reliable tool for commissioning brachytherapy treatment planning systems
title_full Quality assurance procedures based on dosimetric, gamma analysis as a fast reliable tool for commissioning brachytherapy treatment planning systems
title_fullStr Quality assurance procedures based on dosimetric, gamma analysis as a fast reliable tool for commissioning brachytherapy treatment planning systems
title_full_unstemmed Quality assurance procedures based on dosimetric, gamma analysis as a fast reliable tool for commissioning brachytherapy treatment planning systems
title_short Quality assurance procedures based on dosimetric, gamma analysis as a fast reliable tool for commissioning brachytherapy treatment planning systems
title_sort quality assurance procedures based on dosimetric gamma analysis as a fast reliable tool for commissioning brachytherapy treatment planning systems
topic brachytherapy
quality assurance
film dosimetry
ring applicator
url https://doi.org/10.1515/raon-2017-0050
work_keys_str_mv AT zwierzchowskigrzegorz qualityassuranceproceduresbasedondosimetricgammaanalysisasafastreliabletoolforcommissioningbrachytherapytreatmentplanningsystems
AT bieledagrzegorz qualityassuranceproceduresbasedondosimetricgammaanalysisasafastreliabletoolforcommissioningbrachytherapytreatmentplanningsystems
AT skowronekjanusz qualityassuranceproceduresbasedondosimetricgammaanalysisasafastreliabletoolforcommissioningbrachytherapytreatmentplanningsystems