A Real-Time Energy Response Correction Method for Cs<sub>3</sub>Cu<sub>2</sub>I<sub>5</sub>:Tl Scintillating Dosimeter

The uneven energy response of radiation detectors severely limits the accuracy of the dose rate meter used for radiation protection. Currently widely used in dose rate meters as a physical method of setting shielding compensation, the energy response correction error of the detector at different ene...

Full description

Bibliographic Details
Main Authors: Jiaming Li, Leilei Zhang, Jiaqi Wang, Hengyi Su, Zungang Wang, Zhiyuan Li
Format: Article
Language:English
Published: MDPI AG 2023-11-01
Series:Sensors
Subjects:
Online Access:https://www.mdpi.com/1424-8220/23/21/8910
Description
Summary:The uneven energy response of radiation detectors severely limits the accuracy of the dose rate meter used for radiation protection. Currently widely used in dose rate meters as a physical method of setting shielding compensation, the energy response correction error of the detector at different energies is mostly between 15 and 25%. This work designs a real-time correction method for energy response based on a novel Cs<sub>3</sub>Cu<sub>2</sub>I<sub>5</sub>:Tl scintillation detector to improve the accuracy of the dose rate meter used for radiation protection. The technique utilizes the idea of pulse amplitude weighting (PAW) to segment the pulse amplitude histogram. This detector achieves an almost constant energy response after our correction. The experimental results show that compared to <sup>137</sup>Cs γ rays, the maximum error of the response is 8.26% in the photon energy ranging from 33 keV to 1.25 MeV, which is much better than ±30% of the recommended IEC 61526:2010, verifying the feasibility of PAW.
ISSN:1424-8220