Comparison between noninvasive measurement of central venous pressure using near infrared spectroscopy with an invasive central venous pressure monitoring in cardiac surgical Intensive Care Unit
Introduction: Central venous pressure (CVP) measurement is essential in the management of certain clinical situations, including cardiac failure, volume overload and sepsis. CVP measurement requires catheterization of the central vein which is invasive and may lead to complications. The aim of this...
Main Authors: | , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Wolters Kluwer Medknow Publications
2016-01-01
|
Series: | Annals of Cardiac Anaesthesia |
Subjects: | |
Online Access: | http://www.annals.in/article.asp?issn=0971-9784;year=2016;volume=19;issue=3;spage=405;epage=409;aulast=Sathish |
_version_ | 1811303497681862656 |
---|---|
author | N Sathish Naveen G Singh P S Nagaraja B M Sarala C G Prabhushankar Manasa Dhananjaya N Manjunatha |
author_facet | N Sathish Naveen G Singh P S Nagaraja B M Sarala C G Prabhushankar Manasa Dhananjaya N Manjunatha |
author_sort | N Sathish |
collection | DOAJ |
description | Introduction: Central venous pressure (CVP) measurement is essential in the management of certain clinical situations, including cardiac failure, volume overload and sepsis. CVP measurement requires catheterization of the central vein which is invasive and may lead to complications. The aim of this study was to evaluate the accuracy of measurement of CVP using a new noninvasive method based on near infrared spectroscopy (NIRS) in a group of cardiac surgical Intensive Care Unit (ICU) patients. Methodology: Thirty patients in cardiac surgical ICU were enrolled in the study who had an in situ central venous catheter (CVC). Sixty measurements were recorded in 1 h for each patient. A total of 1800 values were compared between noninvasive CVP (CVPn) obtained from Mespere VENUS 2000 CVP system and invasive CVP (CVPi) obtained from CVC. Results: Strong positive correlation was found between CVPi and CVPn (R = 0.9272, P < 0.0001). Linear regression equation - CVPi = 0.5404 + 0.8875 × CVPn (r2 = 0.86, P < 0.001), Bland-Altman bias plots showed mean difference ± standard deviation and limits of agreement: −0.31 ± 1.36 and − 2.99 to + 2.37 (CVPi-CVPn). Conclusion: Noninvasive assessment of the CVP based on NIRS yields readings consistently close to those measured invasively. CVPn may be a clinically useful substitute for CVPi measurements with an advantage of being simple and continuous. It is a promising tool for early management of acute state wherein knowledge of CVP is helpful. |
first_indexed | 2024-04-13T07:48:01Z |
format | Article |
id | doaj.art-6757f937c848424ab125412527f613f5 |
institution | Directory Open Access Journal |
issn | 0971-9784 |
language | English |
last_indexed | 2024-04-13T07:48:01Z |
publishDate | 2016-01-01 |
publisher | Wolters Kluwer Medknow Publications |
record_format | Article |
series | Annals of Cardiac Anaesthesia |
spelling | doaj.art-6757f937c848424ab125412527f613f52022-12-22T02:55:37ZengWolters Kluwer Medknow PublicationsAnnals of Cardiac Anaesthesia0971-97842016-01-0119340540910.4103/0971-9784.185520Comparison between noninvasive measurement of central venous pressure using near infrared spectroscopy with an invasive central venous pressure monitoring in cardiac surgical Intensive Care UnitN SathishNaveen G SinghP S NagarajaB M SaralaC G PrabhushankarManasa DhananjayaN ManjunathaIntroduction: Central venous pressure (CVP) measurement is essential in the management of certain clinical situations, including cardiac failure, volume overload and sepsis. CVP measurement requires catheterization of the central vein which is invasive and may lead to complications. The aim of this study was to evaluate the accuracy of measurement of CVP using a new noninvasive method based on near infrared spectroscopy (NIRS) in a group of cardiac surgical Intensive Care Unit (ICU) patients. Methodology: Thirty patients in cardiac surgical ICU were enrolled in the study who had an in situ central venous catheter (CVC). Sixty measurements were recorded in 1 h for each patient. A total of 1800 values were compared between noninvasive CVP (CVPn) obtained from Mespere VENUS 2000 CVP system and invasive CVP (CVPi) obtained from CVC. Results: Strong positive correlation was found between CVPi and CVPn (R = 0.9272, P < 0.0001). Linear regression equation - CVPi = 0.5404 + 0.8875 × CVPn (r2 = 0.86, P < 0.001), Bland-Altman bias plots showed mean difference ± standard deviation and limits of agreement: −0.31 ± 1.36 and − 2.99 to + 2.37 (CVPi-CVPn). Conclusion: Noninvasive assessment of the CVP based on NIRS yields readings consistently close to those measured invasively. CVPn may be a clinically useful substitute for CVPi measurements with an advantage of being simple and continuous. It is a promising tool for early management of acute state wherein knowledge of CVP is helpful.http://www.annals.in/article.asp?issn=0971-9784;year=2016;volume=19;issue=3;spage=405;epage=409;aulast=SathishCardiac preload; Central venous pressure; Near infrared spectroscopy |
spellingShingle | N Sathish Naveen G Singh P S Nagaraja B M Sarala C G Prabhushankar Manasa Dhananjaya N Manjunatha Comparison between noninvasive measurement of central venous pressure using near infrared spectroscopy with an invasive central venous pressure monitoring in cardiac surgical Intensive Care Unit Annals of Cardiac Anaesthesia Cardiac preload; Central venous pressure; Near infrared spectroscopy |
title | Comparison between noninvasive measurement of central venous pressure using near infrared spectroscopy with an invasive central venous pressure monitoring in cardiac surgical Intensive Care Unit |
title_full | Comparison between noninvasive measurement of central venous pressure using near infrared spectroscopy with an invasive central venous pressure monitoring in cardiac surgical Intensive Care Unit |
title_fullStr | Comparison between noninvasive measurement of central venous pressure using near infrared spectroscopy with an invasive central venous pressure monitoring in cardiac surgical Intensive Care Unit |
title_full_unstemmed | Comparison between noninvasive measurement of central venous pressure using near infrared spectroscopy with an invasive central venous pressure monitoring in cardiac surgical Intensive Care Unit |
title_short | Comparison between noninvasive measurement of central venous pressure using near infrared spectroscopy with an invasive central venous pressure monitoring in cardiac surgical Intensive Care Unit |
title_sort | comparison between noninvasive measurement of central venous pressure using near infrared spectroscopy with an invasive central venous pressure monitoring in cardiac surgical intensive care unit |
topic | Cardiac preload; Central venous pressure; Near infrared spectroscopy |
url | http://www.annals.in/article.asp?issn=0971-9784;year=2016;volume=19;issue=3;spage=405;epage=409;aulast=Sathish |
work_keys_str_mv | AT nsathish comparisonbetweennoninvasivemeasurementofcentralvenouspressureusingnearinfraredspectroscopywithaninvasivecentralvenouspressuremonitoringincardiacsurgicalintensivecareunit AT naveengsingh comparisonbetweennoninvasivemeasurementofcentralvenouspressureusingnearinfraredspectroscopywithaninvasivecentralvenouspressuremonitoringincardiacsurgicalintensivecareunit AT psnagaraja comparisonbetweennoninvasivemeasurementofcentralvenouspressureusingnearinfraredspectroscopywithaninvasivecentralvenouspressuremonitoringincardiacsurgicalintensivecareunit AT bmsarala comparisonbetweennoninvasivemeasurementofcentralvenouspressureusingnearinfraredspectroscopywithaninvasivecentralvenouspressuremonitoringincardiacsurgicalintensivecareunit AT cgprabhushankar comparisonbetweennoninvasivemeasurementofcentralvenouspressureusingnearinfraredspectroscopywithaninvasivecentralvenouspressuremonitoringincardiacsurgicalintensivecareunit AT manasadhananjaya comparisonbetweennoninvasivemeasurementofcentralvenouspressureusingnearinfraredspectroscopywithaninvasivecentralvenouspressuremonitoringincardiacsurgicalintensivecareunit AT nmanjunatha comparisonbetweennoninvasivemeasurementofcentralvenouspressureusingnearinfraredspectroscopywithaninvasivecentralvenouspressuremonitoringincardiacsurgicalintensivecareunit |