Patients skin dose measurements during interventional radiological examinations using Gafchromic XR-RV3 FILM
This work is devoted to the assessment of the absorbed dose in the skin of patients undergoing interventional radiological examinations. There is a probability of deterministic effects in the skin of patients due to high doses of radiation during this type of medical examinations. The aim of this wo...
Main Author: | |
---|---|
Format: | Article |
Language: | English |
Published: |
Saint-Petersburg Research Institute of Radiation Hygiene after Professor P.V. Ramzaev
2020-01-01
|
Series: | Радиационная гигиена |
Subjects: | |
Online Access: | https://www.radhyg.ru/jour/article/view/663 |
_version_ | 1797873790618697728 |
---|---|
author | Svetlana S. Sarycheva |
author_facet | Svetlana S. Sarycheva |
author_sort | Svetlana S. Sarycheva |
collection | DOAJ |
description | This work is devoted to the assessment of the absorbed dose in the skin of patients undergoing interventional radiological examinations. There is a probability of deterministic effects in the skin of patients due to high doses of radiation during this type of medical examinations. The aim of this work was to conduct direct measurements of the absorbed dose in the skin of patients undergoing interventional radiological procedures using special dosimetric radiochromic films Gafchromic XR-RV3 to visualize the distribution of radiation over the patient’s skin surface and to study the possibility of exceeding the threshold values for the deterministic effects in the skin. The paper discusses the features of measurements with GafchromiC XR-RV3 films. The method of film digitization on conventional flatbed scanner and image processing with the ImageJ program was tested. The obtained calibration curve for this type of film was presented. The skin dose distributions for several interventional studies obtained with radiochromic films were presented. The measured value of the maximum absorbed dose in the skin for four of the fourteen analyzed procedures exceeded the threshold value of the absorbed dose for the occurrence of skin erythema in 2 Gy. The highest values of the maximum absorbed dose in the skin were obtained for the coronary angioplasty – 3.2 Gy and for the procedure of embolization of the uterine arteries – 2.9 Gy. |
first_indexed | 2024-04-10T01:21:42Z |
format | Article |
id | doaj.art-67c35c2d61b2495a8b6c7a6bb0742c23 |
institution | Directory Open Access Journal |
issn | 1998-426X |
language | English |
last_indexed | 2024-04-10T01:21:42Z |
publishDate | 2020-01-01 |
publisher | Saint-Petersburg Research Institute of Radiation Hygiene after Professor P.V. Ramzaev |
record_format | Article |
series | Радиационная гигиена |
spelling | doaj.art-67c35c2d61b2495a8b6c7a6bb0742c232023-03-13T09:47:10ZengSaint-Petersburg Research Institute of Radiation Hygiene after Professor P.V. RamzaevРадиационная гигиена1998-426X2020-01-01124899510.21514/1998-426X-2019-12-4-89-95625Patients skin dose measurements during interventional radiological examinations using Gafchromic XR-RV3 FILMSvetlana S. Sarycheva0Санкт-Петербургский научно-исследовательский институт радиационной гигиены имени профессора П.В. Рамзаева, Федеральная служба по надзору в сфере защиты прав потребителей и благополучия человекаThis work is devoted to the assessment of the absorbed dose in the skin of patients undergoing interventional radiological examinations. There is a probability of deterministic effects in the skin of patients due to high doses of radiation during this type of medical examinations. The aim of this work was to conduct direct measurements of the absorbed dose in the skin of patients undergoing interventional radiological procedures using special dosimetric radiochromic films Gafchromic XR-RV3 to visualize the distribution of radiation over the patient’s skin surface and to study the possibility of exceeding the threshold values for the deterministic effects in the skin. The paper discusses the features of measurements with GafchromiC XR-RV3 films. The method of film digitization on conventional flatbed scanner and image processing with the ImageJ program was tested. The obtained calibration curve for this type of film was presented. The skin dose distributions for several interventional studies obtained with radiochromic films were presented. The measured value of the maximum absorbed dose in the skin for four of the fourteen analyzed procedures exceeded the threshold value of the absorbed dose for the occurrence of skin erythema in 2 Gy. The highest values of the maximum absorbed dose in the skin were obtained for the coronary angioplasty – 3.2 Gy and for the procedure of embolization of the uterine arteries – 2.9 Gy.https://www.radhyg.ru/jour/article/view/663интервенционная радиологияпленочная дозиметриядетерминированные эффектыпоглощенная доза в коже |
spellingShingle | Svetlana S. Sarycheva Patients skin dose measurements during interventional radiological examinations using Gafchromic XR-RV3 FILM Радиационная гигиена интервенционная радиология пленочная дозиметрия детерминированные эффекты поглощенная доза в коже |
title | Patients skin dose measurements during interventional radiological examinations using Gafchromic XR-RV3 FILM |
title_full | Patients skin dose measurements during interventional radiological examinations using Gafchromic XR-RV3 FILM |
title_fullStr | Patients skin dose measurements during interventional radiological examinations using Gafchromic XR-RV3 FILM |
title_full_unstemmed | Patients skin dose measurements during interventional radiological examinations using Gafchromic XR-RV3 FILM |
title_short | Patients skin dose measurements during interventional radiological examinations using Gafchromic XR-RV3 FILM |
title_sort | patients skin dose measurements during interventional radiological examinations using gafchromic xr rv3 film |
topic | интервенционная радиология пленочная дозиметрия детерминированные эффекты поглощенная доза в коже |
url | https://www.radhyg.ru/jour/article/view/663 |
work_keys_str_mv | AT svetlanassarycheva patientsskindosemeasurementsduringinterventionalradiologicalexaminationsusinggafchromicxrrv3film |