Repeatability and agreement of AOCT-1000 M, RTVue XR and IOL master 500 in measuring corneal thickness mapping and axial length applying principle of optical coherence tomography
Abstract Purpose To evaluate the repeatability and agreement of Fourier-domain optical coherence tomography (AOCT-1000 M and RTVue XR) and partial coherence interferometry biometer (IOL Master 500) in measuring corneal thickness mapping and axial length respectively. Methods Corneal thickness was me...
Main Authors: | , , , , , , , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
BMC
2023-11-01
|
Series: | BMC Medical Imaging |
Subjects: | |
Online Access: | https://doi.org/10.1186/s12880-023-01147-7 |
_version_ | 1797451049248751616 |
---|---|
author | Hailong Ni Suzhong Xu Li Tian Jieli Mao Jing Li Na Lin Peike Hu Zhiyi Wu Xiang Chen Zhishu Bao Jingwei Zheng Peihua Yan Ruzhi Deng |
author_facet | Hailong Ni Suzhong Xu Li Tian Jieli Mao Jing Li Na Lin Peike Hu Zhiyi Wu Xiang Chen Zhishu Bao Jingwei Zheng Peihua Yan Ruzhi Deng |
author_sort | Hailong Ni |
collection | DOAJ |
description | Abstract Purpose To evaluate the repeatability and agreement of Fourier-domain optical coherence tomography (AOCT-1000 M and RTVue XR) and partial coherence interferometry biometer (IOL Master 500) in measuring corneal thickness mapping and axial length respectively. Methods Corneal thickness was measured by AOCT-1000 M and RTVue XR. Axial lengths were measured by AOCT-1000 M and IOL Master 500. The repeatability and agreement of corneal thickness and axial length were calculated in two groups of devices. The intraclass correlation coefficient (ICC) was used to verify the repeatability of the device. The 95% confidence interval of the difference compared to the set cut-off value was used to verify the agreement between the two devices. Results A total of 60 subjects with 58 eyes were included. The central corneal thickness measured by AOCT-1000 M and RTVue XR were 504.46 ± 42.53 μm and 504.43 ± 42.89 μm respectively. The average difference between groups was 0.03 ± 4.58 μm, and the 95% confidence interval was (-1.17, 1.24), which was far less than the set threshold value of 15 μm (P < 0.001). Both RTVue XR and AOCT-1000 M had very good ICC values of central corneal thickness (0.998 and 0.994, respectively). The average axial lengths measured by AOCT-1000 M and IOL Master 500 were 24.28 ± 1.25 mm and 24.29 ± 1.26 mm respectively and the 95% confidence interval was (-0.02, 0.01), which was less than the set threshold value of 0.15 mm (P < 0.001). The ICC for both devices were 1.000. Conclusion Good repeatability and agreement were seen in measurements of central corneal thickness and axial length by AOCT-1000 M. |
first_indexed | 2024-03-09T14:49:20Z |
format | Article |
id | doaj.art-69d1e56ceede489b8bbe1b63b8ba2ad4 |
institution | Directory Open Access Journal |
issn | 1471-2342 |
language | English |
last_indexed | 2024-03-09T14:49:20Z |
publishDate | 2023-11-01 |
publisher | BMC |
record_format | Article |
series | BMC Medical Imaging |
spelling | doaj.art-69d1e56ceede489b8bbe1b63b8ba2ad42023-11-26T14:35:16ZengBMCBMC Medical Imaging1471-23422023-11-0123111110.1186/s12880-023-01147-7Repeatability and agreement of AOCT-1000 M, RTVue XR and IOL master 500 in measuring corneal thickness mapping and axial length applying principle of optical coherence tomographyHailong Ni0Suzhong Xu1Li Tian2Jieli Mao3Jing Li4Na Lin5Peike Hu6Zhiyi Wu7Xiang Chen8Zhishu Bao9Jingwei Zheng10Peihua Yan11Ruzhi Deng12Eye Center, Second Affiliated Hospital, School of Medicine, Zhejiang UniversityEye hospital of Wenzhou Medical UniversityEye hospital of Wenzhou Medical UniversityEye hospital of Wenzhou Medical UniversityEye hospital of Wenzhou Medical UniversityEye hospital of Wenzhou Medical UniversityEye Center, Second Affiliated Hospital, School of Medicine, Zhejiang UniversityEye Center, Second Affiliated Hospital, School of Medicine, Zhejiang UniversityEye Center, Second Affiliated Hospital, School of Medicine, Zhejiang UniversityEye hospital of Wenzhou Medical UniversityEye hospital of Wenzhou Medical UniversityVision X Medical Technology Co., Ltd.Eye hospital of Wenzhou Medical UniversityAbstract Purpose To evaluate the repeatability and agreement of Fourier-domain optical coherence tomography (AOCT-1000 M and RTVue XR) and partial coherence interferometry biometer (IOL Master 500) in measuring corneal thickness mapping and axial length respectively. Methods Corneal thickness was measured by AOCT-1000 M and RTVue XR. Axial lengths were measured by AOCT-1000 M and IOL Master 500. The repeatability and agreement of corneal thickness and axial length were calculated in two groups of devices. The intraclass correlation coefficient (ICC) was used to verify the repeatability of the device. The 95% confidence interval of the difference compared to the set cut-off value was used to verify the agreement between the two devices. Results A total of 60 subjects with 58 eyes were included. The central corneal thickness measured by AOCT-1000 M and RTVue XR were 504.46 ± 42.53 μm and 504.43 ± 42.89 μm respectively. The average difference between groups was 0.03 ± 4.58 μm, and the 95% confidence interval was (-1.17, 1.24), which was far less than the set threshold value of 15 μm (P < 0.001). Both RTVue XR and AOCT-1000 M had very good ICC values of central corneal thickness (0.998 and 0.994, respectively). The average axial lengths measured by AOCT-1000 M and IOL Master 500 were 24.28 ± 1.25 mm and 24.29 ± 1.26 mm respectively and the 95% confidence interval was (-0.02, 0.01), which was less than the set threshold value of 0.15 mm (P < 0.001). The ICC for both devices were 1.000. Conclusion Good repeatability and agreement were seen in measurements of central corneal thickness and axial length by AOCT-1000 M.https://doi.org/10.1186/s12880-023-01147-7RepeatabilityAgreementCentral corneal thicknessAxial lengthRTVue XRAOCT-1000 M |
spellingShingle | Hailong Ni Suzhong Xu Li Tian Jieli Mao Jing Li Na Lin Peike Hu Zhiyi Wu Xiang Chen Zhishu Bao Jingwei Zheng Peihua Yan Ruzhi Deng Repeatability and agreement of AOCT-1000 M, RTVue XR and IOL master 500 in measuring corneal thickness mapping and axial length applying principle of optical coherence tomography BMC Medical Imaging Repeatability Agreement Central corneal thickness Axial length RTVue XR AOCT-1000 M |
title | Repeatability and agreement of AOCT-1000 M, RTVue XR and IOL master 500 in measuring corneal thickness mapping and axial length applying principle of optical coherence tomography |
title_full | Repeatability and agreement of AOCT-1000 M, RTVue XR and IOL master 500 in measuring corneal thickness mapping and axial length applying principle of optical coherence tomography |
title_fullStr | Repeatability and agreement of AOCT-1000 M, RTVue XR and IOL master 500 in measuring corneal thickness mapping and axial length applying principle of optical coherence tomography |
title_full_unstemmed | Repeatability and agreement of AOCT-1000 M, RTVue XR and IOL master 500 in measuring corneal thickness mapping and axial length applying principle of optical coherence tomography |
title_short | Repeatability and agreement of AOCT-1000 M, RTVue XR and IOL master 500 in measuring corneal thickness mapping and axial length applying principle of optical coherence tomography |
title_sort | repeatability and agreement of aoct 1000 m rtvue xr and iol master 500 in measuring corneal thickness mapping and axial length applying principle of optical coherence tomography |
topic | Repeatability Agreement Central corneal thickness Axial length RTVue XR AOCT-1000 M |
url | https://doi.org/10.1186/s12880-023-01147-7 |
work_keys_str_mv | AT hailongni repeatabilityandagreementofaoct1000mrtvuexrandiolmaster500inmeasuringcornealthicknessmappingandaxiallengthapplyingprincipleofopticalcoherencetomography AT suzhongxu repeatabilityandagreementofaoct1000mrtvuexrandiolmaster500inmeasuringcornealthicknessmappingandaxiallengthapplyingprincipleofopticalcoherencetomography AT litian repeatabilityandagreementofaoct1000mrtvuexrandiolmaster500inmeasuringcornealthicknessmappingandaxiallengthapplyingprincipleofopticalcoherencetomography AT jielimao repeatabilityandagreementofaoct1000mrtvuexrandiolmaster500inmeasuringcornealthicknessmappingandaxiallengthapplyingprincipleofopticalcoherencetomography AT jingli repeatabilityandagreementofaoct1000mrtvuexrandiolmaster500inmeasuringcornealthicknessmappingandaxiallengthapplyingprincipleofopticalcoherencetomography AT nalin repeatabilityandagreementofaoct1000mrtvuexrandiolmaster500inmeasuringcornealthicknessmappingandaxiallengthapplyingprincipleofopticalcoherencetomography AT peikehu repeatabilityandagreementofaoct1000mrtvuexrandiolmaster500inmeasuringcornealthicknessmappingandaxiallengthapplyingprincipleofopticalcoherencetomography AT zhiyiwu repeatabilityandagreementofaoct1000mrtvuexrandiolmaster500inmeasuringcornealthicknessmappingandaxiallengthapplyingprincipleofopticalcoherencetomography AT xiangchen repeatabilityandagreementofaoct1000mrtvuexrandiolmaster500inmeasuringcornealthicknessmappingandaxiallengthapplyingprincipleofopticalcoherencetomography AT zhishubao repeatabilityandagreementofaoct1000mrtvuexrandiolmaster500inmeasuringcornealthicknessmappingandaxiallengthapplyingprincipleofopticalcoherencetomography AT jingweizheng repeatabilityandagreementofaoct1000mrtvuexrandiolmaster500inmeasuringcornealthicknessmappingandaxiallengthapplyingprincipleofopticalcoherencetomography AT peihuayan repeatabilityandagreementofaoct1000mrtvuexrandiolmaster500inmeasuringcornealthicknessmappingandaxiallengthapplyingprincipleofopticalcoherencetomography AT ruzhideng repeatabilityandagreementofaoct1000mrtvuexrandiolmaster500inmeasuringcornealthicknessmappingandaxiallengthapplyingprincipleofopticalcoherencetomography |