Clinical Validation of Dental Implant Stability by Newly Designed Damping Capacity Assessment Device during the Healing Period

<i>Background and Objectives:</i> To evaluate the stability of a dental implant and the effectiveness of a newly designed damping capacity assessment device by improving the number of blows and strength evaluated by a prospective clinical study. <i>Materials and Method:</i> T...

Full description

Bibliographic Details
Main Authors: Ho-Kyung Lim, Sung-Jae Lee, Yujin Jeong, Jong-Seok Lee, Jae-Jun Ryu, Ji-Suk Shim, In-Seok Song
Format: Article
Language:English
Published: MDPI AG 2022-10-01
Series:Medicina
Subjects:
Online Access:https://www.mdpi.com/1648-9144/58/11/1570
_version_ 1827646143006244864
author Ho-Kyung Lim
Sung-Jae Lee
Yujin Jeong
Jong-Seok Lee
Jae-Jun Ryu
Ji-Suk Shim
In-Seok Song
author_facet Ho-Kyung Lim
Sung-Jae Lee
Yujin Jeong
Jong-Seok Lee
Jae-Jun Ryu
Ji-Suk Shim
In-Seok Song
author_sort Ho-Kyung Lim
collection DOAJ
description <i>Background and Objectives:</i> To evaluate the stability of a dental implant and the effectiveness of a newly designed damping capacity assessment device by improving the number of blows and strength evaluated by a prospective clinical study. <i>Materials and Method:</i> The stability of dental implants was measured in 50 implants in a total of 38 patients. Measurements were performed using Anycheck and Periotest M devices, twice in total, divided into buccal and lingual directions. In addition, measurements were performed on the day of surgery, two weeks, one month, two months, and three months after surgery for a total of five times. After the standardization of the measured values, the differences and changes over time for each device were observed. <i>Result:</i> No difference in standardized values between the two devices was observed at any time point. In both devices, stability decreased at two weeks postoperatively but gradually increased thereafter. No differences were observed in the values according to the measurement direction. <i>Conclusions:</i> The damping capacity of Anycheck was similar to that of Periotest M. After a slight decrease in stability two weeks after implant placement, implant stability increased over time.
first_indexed 2024-03-09T18:51:53Z
format Article
id doaj.art-6a5a397b489a4978af17d92245c2082e
institution Directory Open Access Journal
issn 1010-660X
1648-9144
language English
last_indexed 2024-03-09T18:51:53Z
publishDate 2022-10-01
publisher MDPI AG
record_format Article
series Medicina
spelling doaj.art-6a5a397b489a4978af17d92245c2082e2023-11-24T05:46:57ZengMDPI AGMedicina1010-660X1648-91442022-10-015811157010.3390/medicina58111570Clinical Validation of Dental Implant Stability by Newly Designed Damping Capacity Assessment Device during the Healing PeriodHo-Kyung Lim0Sung-Jae Lee1Yujin Jeong2Jong-Seok Lee3Jae-Jun Ryu4Ji-Suk Shim5In-Seok Song6Department of Oral & Maxillofacial Surgery, Korea University Guro Hospital, Seoul 08308, KoreaDepartment of Oral & Maxillofacial Surgery, Korea University Anam Hospital, Seoul 02841, KoreaDepartment of Biostatistics, Korea University College of Medicine, Seoul 02841, KoreaDepartment of Prosthodontics, Korea University Anam Hospital, Seoul 02841, KoreaDepartment of Prosthodontics, Korea University Anam Hospital, Seoul 02841, KoreaDepartment of Prosthodontics, Korea University Guro Hospital, Seoul 08308, KoreaDepartment of Oral & Maxillofacial Surgery, Korea University Anam Hospital, Seoul 02841, Korea<i>Background and Objectives:</i> To evaluate the stability of a dental implant and the effectiveness of a newly designed damping capacity assessment device by improving the number of blows and strength evaluated by a prospective clinical study. <i>Materials and Method:</i> The stability of dental implants was measured in 50 implants in a total of 38 patients. Measurements were performed using Anycheck and Periotest M devices, twice in total, divided into buccal and lingual directions. In addition, measurements were performed on the day of surgery, two weeks, one month, two months, and three months after surgery for a total of five times. After the standardization of the measured values, the differences and changes over time for each device were observed. <i>Result:</i> No difference in standardized values between the two devices was observed at any time point. In both devices, stability decreased at two weeks postoperatively but gradually increased thereafter. No differences were observed in the values according to the measurement direction. <i>Conclusions:</i> The damping capacity of Anycheck was similar to that of Periotest M. After a slight decrease in stability two weeks after implant placement, implant stability increased over time.https://www.mdpi.com/1648-9144/58/11/1570dental implantsstabilitydentistryanalytic device
spellingShingle Ho-Kyung Lim
Sung-Jae Lee
Yujin Jeong
Jong-Seok Lee
Jae-Jun Ryu
Ji-Suk Shim
In-Seok Song
Clinical Validation of Dental Implant Stability by Newly Designed Damping Capacity Assessment Device during the Healing Period
Medicina
dental implants
stability
dentistry
analytic device
title Clinical Validation of Dental Implant Stability by Newly Designed Damping Capacity Assessment Device during the Healing Period
title_full Clinical Validation of Dental Implant Stability by Newly Designed Damping Capacity Assessment Device during the Healing Period
title_fullStr Clinical Validation of Dental Implant Stability by Newly Designed Damping Capacity Assessment Device during the Healing Period
title_full_unstemmed Clinical Validation of Dental Implant Stability by Newly Designed Damping Capacity Assessment Device during the Healing Period
title_short Clinical Validation of Dental Implant Stability by Newly Designed Damping Capacity Assessment Device during the Healing Period
title_sort clinical validation of dental implant stability by newly designed damping capacity assessment device during the healing period
topic dental implants
stability
dentistry
analytic device
url https://www.mdpi.com/1648-9144/58/11/1570
work_keys_str_mv AT hokyunglim clinicalvalidationofdentalimplantstabilitybynewlydesigneddampingcapacityassessmentdeviceduringthehealingperiod
AT sungjaelee clinicalvalidationofdentalimplantstabilitybynewlydesigneddampingcapacityassessmentdeviceduringthehealingperiod
AT yujinjeong clinicalvalidationofdentalimplantstabilitybynewlydesigneddampingcapacityassessmentdeviceduringthehealingperiod
AT jongseoklee clinicalvalidationofdentalimplantstabilitybynewlydesigneddampingcapacityassessmentdeviceduringthehealingperiod
AT jaejunryu clinicalvalidationofdentalimplantstabilitybynewlydesigneddampingcapacityassessmentdeviceduringthehealingperiod
AT jisukshim clinicalvalidationofdentalimplantstabilitybynewlydesigneddampingcapacityassessmentdeviceduringthehealingperiod
AT inseoksong clinicalvalidationofdentalimplantstabilitybynewlydesigneddampingcapacityassessmentdeviceduringthehealingperiod