The use of scanning electron microscopy and fixation methods to evaluate the interaction of blood with the surfaces of medical devices

Abstract Testing the hemocompatibility of medical devices after their interaction with blood entails the need to evaluate the activation of blood elements and the degree of their coagulation and adhesion to the device surface. One possible way to achieve this is to use scanning electron microscopy (...

Full description

Bibliographic Details
Main Authors: Martina Nalezinková, Jan Loskot, Alena Myslivcová Fučíková
Format: Article
Language:English
Published: Nature Portfolio 2024-02-01
Series:Scientific Reports
Online Access:https://doi.org/10.1038/s41598-024-55136-z
_version_ 1797274721780236288
author Martina Nalezinková
Jan Loskot
Alena Myslivcová Fučíková
author_facet Martina Nalezinková
Jan Loskot
Alena Myslivcová Fučíková
author_sort Martina Nalezinková
collection DOAJ
description Abstract Testing the hemocompatibility of medical devices after their interaction with blood entails the need to evaluate the activation of blood elements and the degree of their coagulation and adhesion to the device surface. One possible way to achieve this is to use scanning electron microscopy (SEM). The aim was to develop a novel SEM-based method to assess the thrombogenic potential of medical devices and their adhesiveness to blood cells. As a part of this task, also find a convenient procedure of efficient and non-destructive sample fixation for SEM while reducing the use of highly toxic substances and shortening the fixation time. A polymeric surgical mesh was exposed to blood so that blood elements adhered to its surface. Such prepared samples were then chemically fixed for a subsequent SEM measurement; a number of fixation procedures were tested to find the optimal one. The fixation results were evaluated from SEM images, and the degree of blood elements’ adhesion was determined from the images using ImageJ software. The best fixation was achieved with the May–Grünwald solution, which is less toxic than chemicals traditionally used. Moreover, manipulation with highly toxic osmium tetroxide can be avoided in the proposed procedure. A convenient methodology for SEM image analysis has been developed too, enabling to quantitatively evaluate the interaction of blood with the surfaces of various medical devices. Our method replaces the subjective assessment of surface coverage with a better-defined procedure, thus offering more precise and reliable results.
first_indexed 2024-03-07T15:02:19Z
format Article
id doaj.art-6a6ecfeaa53b48a2b94c095130388d9c
institution Directory Open Access Journal
issn 2045-2322
language English
last_indexed 2024-03-07T15:02:19Z
publishDate 2024-02-01
publisher Nature Portfolio
record_format Article
series Scientific Reports
spelling doaj.art-6a6ecfeaa53b48a2b94c095130388d9c2024-03-05T19:04:52ZengNature PortfolioScientific Reports2045-23222024-02-0114111310.1038/s41598-024-55136-zThe use of scanning electron microscopy and fixation methods to evaluate the interaction of blood with the surfaces of medical devicesMartina Nalezinková0Jan Loskot1Alena Myslivcová Fučíková2Department of Biology, Faculty of Science, University of Hradec KrálovéDepartment of Physics, Faculty of Science, University of Hradec KrálovéDepartment of Biology, Faculty of Science, University of Hradec KrálovéAbstract Testing the hemocompatibility of medical devices after their interaction with blood entails the need to evaluate the activation of blood elements and the degree of their coagulation and adhesion to the device surface. One possible way to achieve this is to use scanning electron microscopy (SEM). The aim was to develop a novel SEM-based method to assess the thrombogenic potential of medical devices and their adhesiveness to blood cells. As a part of this task, also find a convenient procedure of efficient and non-destructive sample fixation for SEM while reducing the use of highly toxic substances and shortening the fixation time. A polymeric surgical mesh was exposed to blood so that blood elements adhered to its surface. Such prepared samples were then chemically fixed for a subsequent SEM measurement; a number of fixation procedures were tested to find the optimal one. The fixation results were evaluated from SEM images, and the degree of blood elements’ adhesion was determined from the images using ImageJ software. The best fixation was achieved with the May–Grünwald solution, which is less toxic than chemicals traditionally used. Moreover, manipulation with highly toxic osmium tetroxide can be avoided in the proposed procedure. A convenient methodology for SEM image analysis has been developed too, enabling to quantitatively evaluate the interaction of blood with the surfaces of various medical devices. Our method replaces the subjective assessment of surface coverage with a better-defined procedure, thus offering more precise and reliable results.https://doi.org/10.1038/s41598-024-55136-z
spellingShingle Martina Nalezinková
Jan Loskot
Alena Myslivcová Fučíková
The use of scanning electron microscopy and fixation methods to evaluate the interaction of blood with the surfaces of medical devices
Scientific Reports
title The use of scanning electron microscopy and fixation methods to evaluate the interaction of blood with the surfaces of medical devices
title_full The use of scanning electron microscopy and fixation methods to evaluate the interaction of blood with the surfaces of medical devices
title_fullStr The use of scanning electron microscopy and fixation methods to evaluate the interaction of blood with the surfaces of medical devices
title_full_unstemmed The use of scanning electron microscopy and fixation methods to evaluate the interaction of blood with the surfaces of medical devices
title_short The use of scanning electron microscopy and fixation methods to evaluate the interaction of blood with the surfaces of medical devices
title_sort use of scanning electron microscopy and fixation methods to evaluate the interaction of blood with the surfaces of medical devices
url https://doi.org/10.1038/s41598-024-55136-z
work_keys_str_mv AT martinanalezinkova theuseofscanningelectronmicroscopyandfixationmethodstoevaluatetheinteractionofbloodwiththesurfacesofmedicaldevices
AT janloskot theuseofscanningelectronmicroscopyandfixationmethodstoevaluatetheinteractionofbloodwiththesurfacesofmedicaldevices
AT alenamyslivcovafucikova theuseofscanningelectronmicroscopyandfixationmethodstoevaluatetheinteractionofbloodwiththesurfacesofmedicaldevices
AT martinanalezinkova useofscanningelectronmicroscopyandfixationmethodstoevaluatetheinteractionofbloodwiththesurfacesofmedicaldevices
AT janloskot useofscanningelectronmicroscopyandfixationmethodstoevaluatetheinteractionofbloodwiththesurfacesofmedicaldevices
AT alenamyslivcovafucikova useofscanningelectronmicroscopyandfixationmethodstoevaluatetheinteractionofbloodwiththesurfacesofmedicaldevices