A PLLA Coating Does Not Affect the Insertion Pressure or Frictional Behavior of a CI Electrode Array at Higher Insertion Speeds

To prevent endocochlear insertion trauma, the development of drug delivery coatings in the field of CI electrodes has become an increasing focus of research. However, so far, the effect of a polymer coating of PLLA on the mechanical properties, such as the insertion pressure and friction of an elect...

Full description

Bibliographic Details
Main Authors: Dana Dohr, Katharina Wulf, Niels Grabow, Robert Mlynski, Sebastian P. Schraven
Format: Article
Language:English
Published: MDPI AG 2022-04-01
Series:Materials
Subjects:
Online Access:https://www.mdpi.com/1996-1944/15/9/3049
_version_ 1797504076350488576
author Dana Dohr
Katharina Wulf
Niels Grabow
Robert Mlynski
Sebastian P. Schraven
author_facet Dana Dohr
Katharina Wulf
Niels Grabow
Robert Mlynski
Sebastian P. Schraven
author_sort Dana Dohr
collection DOAJ
description To prevent endocochlear insertion trauma, the development of drug delivery coatings in the field of CI electrodes has become an increasing focus of research. However, so far, the effect of a polymer coating of PLLA on the mechanical properties, such as the insertion pressure and friction of an electrode array, has not been investigated. In this study, the insertion pressure of a PLLA-coated, 31.5-mm long standard electrode array was examined during placement in a linear cochlear model. Additionally, the friction coefficients between a PLLA-coated electrode array and a tissue simulating the endocochlear lining were acquired. All data were obtained at different insertion speeds (0.1, 0.5, 1.0, 1.5, and 2.0 mm/s) and compared with those of an uncoated electrode array. It was shown that both the maximum insertion pressure generated in the linear model and the friction coefficient of the PLLA-coated electrode did not depend on the insertion speed. At higher insertion speeds above 1.0 mm/s, the insertion pressure (1.268 ± 0.032 mmHg) and the friction coefficient (0.40 ± 0.15) of the coated electrode array were similar to those of an uncoated array (1.252 ± 0.034 mmHg and 0.36 ± 0.15). The present study reveals that a PLLA coating on cochlear electrode arrays has a negligible effect on the electrode array insertion pressure and the friction when higher insertion speeds are used compared with an uncoated electrode array. Therefore, PLLA is a suitable material to be used as a coating for CI electrode arrays and can be considered for a potential drug delivery system.
first_indexed 2024-03-10T03:59:29Z
format Article
id doaj.art-6ae9cc8d27604f6bb309557c89564d2d
institution Directory Open Access Journal
issn 1996-1944
language English
last_indexed 2024-03-10T03:59:29Z
publishDate 2022-04-01
publisher MDPI AG
record_format Article
series Materials
spelling doaj.art-6ae9cc8d27604f6bb309557c89564d2d2023-11-23T08:37:18ZengMDPI AGMaterials1996-19442022-04-01159304910.3390/ma15093049A PLLA Coating Does Not Affect the Insertion Pressure or Frictional Behavior of a CI Electrode Array at Higher Insertion SpeedsDana Dohr0Katharina Wulf1Niels Grabow2Robert Mlynski3Sebastian P. Schraven4Department of Otorhinolaryngology, Head and Neck Surgery “Otto Körner”, Rostock University Medical Center, 18057 Rostock, GermanyInstitute for Biomedical Engineering, Rostock University Medical Center, 18119 Rostock, GermanyInstitute for Biomedical Engineering, Rostock University Medical Center, 18119 Rostock, GermanyDepartment of Otorhinolaryngology, Head and Neck Surgery “Otto Körner”, Rostock University Medical Center, 18057 Rostock, GermanyDepartment of Otorhinolaryngology, Head and Neck Surgery “Otto Körner”, Rostock University Medical Center, 18057 Rostock, GermanyTo prevent endocochlear insertion trauma, the development of drug delivery coatings in the field of CI electrodes has become an increasing focus of research. However, so far, the effect of a polymer coating of PLLA on the mechanical properties, such as the insertion pressure and friction of an electrode array, has not been investigated. In this study, the insertion pressure of a PLLA-coated, 31.5-mm long standard electrode array was examined during placement in a linear cochlear model. Additionally, the friction coefficients between a PLLA-coated electrode array and a tissue simulating the endocochlear lining were acquired. All data were obtained at different insertion speeds (0.1, 0.5, 1.0, 1.5, and 2.0 mm/s) and compared with those of an uncoated electrode array. It was shown that both the maximum insertion pressure generated in the linear model and the friction coefficient of the PLLA-coated electrode did not depend on the insertion speed. At higher insertion speeds above 1.0 mm/s, the insertion pressure (1.268 ± 0.032 mmHg) and the friction coefficient (0.40 ± 0.15) of the coated electrode array were similar to those of an uncoated array (1.252 ± 0.034 mmHg and 0.36 ± 0.15). The present study reveals that a PLLA coating on cochlear electrode arrays has a negligible effect on the electrode array insertion pressure and the friction when higher insertion speeds are used compared with an uncoated electrode array. Therefore, PLLA is a suitable material to be used as a coating for CI electrode arrays and can be considered for a potential drug delivery system.https://www.mdpi.com/1996-1944/15/9/3049PLLA coatingcoated cochlear electrode arrayinsertion pressurelinear cochlear modelfriction coefficientinsertion speed
spellingShingle Dana Dohr
Katharina Wulf
Niels Grabow
Robert Mlynski
Sebastian P. Schraven
A PLLA Coating Does Not Affect the Insertion Pressure or Frictional Behavior of a CI Electrode Array at Higher Insertion Speeds
Materials
PLLA coating
coated cochlear electrode array
insertion pressure
linear cochlear model
friction coefficient
insertion speed
title A PLLA Coating Does Not Affect the Insertion Pressure or Frictional Behavior of a CI Electrode Array at Higher Insertion Speeds
title_full A PLLA Coating Does Not Affect the Insertion Pressure or Frictional Behavior of a CI Electrode Array at Higher Insertion Speeds
title_fullStr A PLLA Coating Does Not Affect the Insertion Pressure or Frictional Behavior of a CI Electrode Array at Higher Insertion Speeds
title_full_unstemmed A PLLA Coating Does Not Affect the Insertion Pressure or Frictional Behavior of a CI Electrode Array at Higher Insertion Speeds
title_short A PLLA Coating Does Not Affect the Insertion Pressure or Frictional Behavior of a CI Electrode Array at Higher Insertion Speeds
title_sort plla coating does not affect the insertion pressure or frictional behavior of a ci electrode array at higher insertion speeds
topic PLLA coating
coated cochlear electrode array
insertion pressure
linear cochlear model
friction coefficient
insertion speed
url https://www.mdpi.com/1996-1944/15/9/3049
work_keys_str_mv AT danadohr apllacoatingdoesnotaffecttheinsertionpressureorfrictionalbehaviorofacielectrodearrayathigherinsertionspeeds
AT katharinawulf apllacoatingdoesnotaffecttheinsertionpressureorfrictionalbehaviorofacielectrodearrayathigherinsertionspeeds
AT nielsgrabow apllacoatingdoesnotaffecttheinsertionpressureorfrictionalbehaviorofacielectrodearrayathigherinsertionspeeds
AT robertmlynski apllacoatingdoesnotaffecttheinsertionpressureorfrictionalbehaviorofacielectrodearrayathigherinsertionspeeds
AT sebastianpschraven apllacoatingdoesnotaffecttheinsertionpressureorfrictionalbehaviorofacielectrodearrayathigherinsertionspeeds
AT danadohr pllacoatingdoesnotaffecttheinsertionpressureorfrictionalbehaviorofacielectrodearrayathigherinsertionspeeds
AT katharinawulf pllacoatingdoesnotaffecttheinsertionpressureorfrictionalbehaviorofacielectrodearrayathigherinsertionspeeds
AT nielsgrabow pllacoatingdoesnotaffecttheinsertionpressureorfrictionalbehaviorofacielectrodearrayathigherinsertionspeeds
AT robertmlynski pllacoatingdoesnotaffecttheinsertionpressureorfrictionalbehaviorofacielectrodearrayathigherinsertionspeeds
AT sebastianpschraven pllacoatingdoesnotaffecttheinsertionpressureorfrictionalbehaviorofacielectrodearrayathigherinsertionspeeds