Verification studies in glucometers: Should we use capillary blood or venous blood for comparison?

Glucometers are widely used in the diagnosis of blood glucose levels in patients with diabetes mellitus. EN ISO 15197 suggests that glucometer comparison studies should have 100 capillary blood samples be worked on at least twice. In this study, we planned on comparing the glucose results measured i...

Full description

Bibliographic Details
Main Authors: Canan Topcuoglu, Fatma Meric Yilmaz, Oguzhan Kaya, Bagdagul Cakir
Format: Article
Language:English
Published: Society of Turaz Bilim 2018-03-01
Series:Medicine Science
Subjects:
Online Access:http://www.ejmanager.com/fulltextpdf.php?mno=278871
_version_ 1797328180302839808
author Canan Topcuoglu
Fatma Meric Yilmaz
Oguzhan Kaya
Bagdagul Cakir
author_facet Canan Topcuoglu
Fatma Meric Yilmaz
Oguzhan Kaya
Bagdagul Cakir
author_sort Canan Topcuoglu
collection DOAJ
description Glucometers are widely used in the diagnosis of blood glucose levels in patients with diabetes mellitus. EN ISO 15197 suggests that glucometer comparison studies should have 100 capillary blood samples be worked on at least twice. In this study, we planned on comparing the glucose results measured in a routine biochemistry analyzer from two different glucometers, capillary and venous blood samples, and aimed to discuss the effects of blood taking systems on the glucometer validation studies. Capillary and venous blood samples were taken from 101 individuals and their glucose concentrations measured simultaneously using two different glucometers (Accu-chek and GlucoMax). Capillary and venous blood samples were centrifuged after clotting and analyzed in the Roche P modular system. In the fasting condition, the equations for regression analysis that were found y=0,873x+24,32 (r=0,857) in between Accu-chek and venous blood glucose, y=0,9x+16,15 (r=0,920) in between Accu-chek and capillary blood glucose, y=0,811x+20,94 (r=0,776) in between GlucoMax and venous blood glucose, and y=0,851x+12,28 (r=0,863) in between GlucoMax and capillary blood glucose.In the postprandial state, the equations were y=0,713x+48,46 (r=0,258) in between Accu-chek and venous blood glucose, y=0,981x+11,77 (r=0,718) in between Accu-chek and capillary blood glucose, y=0,706x+39,12 (r=0,453) in between GlucoMax and venous blood glucose, and y=0,790+22,35 (r=0,787) in between GlucoMax and capillary blood glucose. In the fasting and postprandial state, the capillary glucose levels showed better correlation with glucometer measurements than venous blood glucose levels. In glucometer verification studies, capillary blood obtained with capillary blood sampling systems and used instead of venous blood should be the preferred sample. [Med-Science 2018; 7(1.000): 218-221]
first_indexed 2024-03-08T06:48:10Z
format Article
id doaj.art-6c616e8c2a5e4453866778383ac782c0
institution Directory Open Access Journal
issn 2147-0634
language English
last_indexed 2024-03-08T06:48:10Z
publishDate 2018-03-01
publisher Society of Turaz Bilim
record_format Article
series Medicine Science
spelling doaj.art-6c616e8c2a5e4453866778383ac782c02024-02-03T07:29:03ZengSociety of Turaz BilimMedicine Science2147-06342018-03-017121822110.5455/medscience.2017.06.8696278871Verification studies in glucometers: Should we use capillary blood or venous blood for comparison?Canan Topcuoglu0Fatma Meric YilmazOguzhan KayaBagdagul CakirAnkara Numune Training and Research Hospital, Clinical Biochemistry Laboratory, Ankara, Turkey Yildirim Beyazit University Medical Faculty Biochemistry Department, Ankara, TurkeyGlucometers are widely used in the diagnosis of blood glucose levels in patients with diabetes mellitus. EN ISO 15197 suggests that glucometer comparison studies should have 100 capillary blood samples be worked on at least twice. In this study, we planned on comparing the glucose results measured in a routine biochemistry analyzer from two different glucometers, capillary and venous blood samples, and aimed to discuss the effects of blood taking systems on the glucometer validation studies. Capillary and venous blood samples were taken from 101 individuals and their glucose concentrations measured simultaneously using two different glucometers (Accu-chek and GlucoMax). Capillary and venous blood samples were centrifuged after clotting and analyzed in the Roche P modular system. In the fasting condition, the equations for regression analysis that were found y=0,873x+24,32 (r=0,857) in between Accu-chek and venous blood glucose, y=0,9x+16,15 (r=0,920) in between Accu-chek and capillary blood glucose, y=0,811x+20,94 (r=0,776) in between GlucoMax and venous blood glucose, and y=0,851x+12,28 (r=0,863) in between GlucoMax and capillary blood glucose.In the postprandial state, the equations were y=0,713x+48,46 (r=0,258) in between Accu-chek and venous blood glucose, y=0,981x+11,77 (r=0,718) in between Accu-chek and capillary blood glucose, y=0,706x+39,12 (r=0,453) in between GlucoMax and venous blood glucose, and y=0,790+22,35 (r=0,787) in between GlucoMax and capillary blood glucose. In the fasting and postprandial state, the capillary glucose levels showed better correlation with glucometer measurements than venous blood glucose levels. In glucometer verification studies, capillary blood obtained with capillary blood sampling systems and used instead of venous blood should be the preferred sample. [Med-Science 2018; 7(1.000): 218-221]http://www.ejmanager.com/fulltextpdf.php?mno=278871Capillary bloodvenous bloodglucometer r
spellingShingle Canan Topcuoglu
Fatma Meric Yilmaz
Oguzhan Kaya
Bagdagul Cakir
Verification studies in glucometers: Should we use capillary blood or venous blood for comparison?
Medicine Science
Capillary blood
venous blood
glucometer r
title Verification studies in glucometers: Should we use capillary blood or venous blood for comparison?
title_full Verification studies in glucometers: Should we use capillary blood or venous blood for comparison?
title_fullStr Verification studies in glucometers: Should we use capillary blood or venous blood for comparison?
title_full_unstemmed Verification studies in glucometers: Should we use capillary blood or venous blood for comparison?
title_short Verification studies in glucometers: Should we use capillary blood or venous blood for comparison?
title_sort verification studies in glucometers should we use capillary blood or venous blood for comparison
topic Capillary blood
venous blood
glucometer r
url http://www.ejmanager.com/fulltextpdf.php?mno=278871
work_keys_str_mv AT canantopcuoglu verificationstudiesinglucometersshouldweusecapillarybloodorvenousbloodforcomparison
AT fatmamericyilmaz verificationstudiesinglucometersshouldweusecapillarybloodorvenousbloodforcomparison
AT oguzhankaya verificationstudiesinglucometersshouldweusecapillarybloodorvenousbloodforcomparison
AT bagdagulcakir verificationstudiesinglucometersshouldweusecapillarybloodorvenousbloodforcomparison