Repeatability and reproducibility of a double-pass optical quality analysis device.
To evaluate the repeatability and reproducibility of a double-pass instrument (OQASII, Visiomereics SL, Spain), which objectively measures overall optical quality of the human eyes.The right eye of 119 healthy subjects with best corrected visual acuity of 20/25 or better was included in this prospec...
Main Authors: | , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Public Library of Science (PLoS)
2015-01-01
|
Series: | PLoS ONE |
Online Access: | http://europepmc.org/articles/PMC4315520?pdf=render |
_version_ | 1811282578647285760 |
---|---|
author | Chen-Chen Xu Ting Xue Qin-Mei Wang Yi-Ni Zhou Jin-Hai Huang A-Yong Yu |
author_facet | Chen-Chen Xu Ting Xue Qin-Mei Wang Yi-Ni Zhou Jin-Hai Huang A-Yong Yu |
author_sort | Chen-Chen Xu |
collection | DOAJ |
description | To evaluate the repeatability and reproducibility of a double-pass instrument (OQASII, Visiomereics SL, Spain), which objectively measures overall optical quality of the human eyes.The right eye of 119 healthy subjects with best corrected visual acuity of 20/25 or better was included in this prospective, comparative, observational study. Two separate tests with OQASII were conducted sequentially on the same day by two different examiners. A week later, the first examiner conducted the third measurement. All subjects underwent three consecutive tests during each session. The repeatability and reproducibility of the modulation transfer function cut off frequency (MTF cutoff), the Strehl ratio, the OQAS values (OVs) at contrasts of 100%, 20% and 9%, and the objective scatter index (OSI) were analyzed.For MTF cutoff, Strehl ratio, OV100%, OV20%, OV9%, and OSI, the mean values were 39.32±9.75cpd, 0.22±0.06, 1.31±0.33, 1.33±0.39, 1.33±0.41, 0.60±0.42, respectively. Repeatability and reproducibility were good with a very low coefficient of variation and high interclass correlation coefficients (>0.88) for all parameters. Bland-Altman plots showed good correlation with 95% limits of agreement ranged from -6.04 to 6.78cpd, -0.05 to 0.05, -0.20 to 0.23, -0.29 to 0.32, -0.40 to 0.42, -0.23 to 0.21 in inter-observer, and -6.56 to 7.42cpd, -0.06 to 0.06, -0.22 to 0.24, -0.30 to 0.32, -0.35 to 0.34, -0.24 to 0.23 in inter-visit, respectively.The OQASII system yields excellent repeatability and good reproducibility for objective measurements of overall optical quality in clinic. |
first_indexed | 2024-04-13T01:53:56Z |
format | Article |
id | doaj.art-6ed639edcdf64294bbe3b78d516ef1f0 |
institution | Directory Open Access Journal |
issn | 1932-6203 |
language | English |
last_indexed | 2024-04-13T01:53:56Z |
publishDate | 2015-01-01 |
publisher | Public Library of Science (PLoS) |
record_format | Article |
series | PLoS ONE |
spelling | doaj.art-6ed639edcdf64294bbe3b78d516ef1f02022-12-22T03:07:48ZengPublic Library of Science (PLoS)PLoS ONE1932-62032015-01-01102e011758710.1371/journal.pone.0117587Repeatability and reproducibility of a double-pass optical quality analysis device.Chen-Chen XuTing XueQin-Mei WangYi-Ni ZhouJin-Hai HuangA-Yong YuTo evaluate the repeatability and reproducibility of a double-pass instrument (OQASII, Visiomereics SL, Spain), which objectively measures overall optical quality of the human eyes.The right eye of 119 healthy subjects with best corrected visual acuity of 20/25 or better was included in this prospective, comparative, observational study. Two separate tests with OQASII were conducted sequentially on the same day by two different examiners. A week later, the first examiner conducted the third measurement. All subjects underwent three consecutive tests during each session. The repeatability and reproducibility of the modulation transfer function cut off frequency (MTF cutoff), the Strehl ratio, the OQAS values (OVs) at contrasts of 100%, 20% and 9%, and the objective scatter index (OSI) were analyzed.For MTF cutoff, Strehl ratio, OV100%, OV20%, OV9%, and OSI, the mean values were 39.32±9.75cpd, 0.22±0.06, 1.31±0.33, 1.33±0.39, 1.33±0.41, 0.60±0.42, respectively. Repeatability and reproducibility were good with a very low coefficient of variation and high interclass correlation coefficients (>0.88) for all parameters. Bland-Altman plots showed good correlation with 95% limits of agreement ranged from -6.04 to 6.78cpd, -0.05 to 0.05, -0.20 to 0.23, -0.29 to 0.32, -0.40 to 0.42, -0.23 to 0.21 in inter-observer, and -6.56 to 7.42cpd, -0.06 to 0.06, -0.22 to 0.24, -0.30 to 0.32, -0.35 to 0.34, -0.24 to 0.23 in inter-visit, respectively.The OQASII system yields excellent repeatability and good reproducibility for objective measurements of overall optical quality in clinic.http://europepmc.org/articles/PMC4315520?pdf=render |
spellingShingle | Chen-Chen Xu Ting Xue Qin-Mei Wang Yi-Ni Zhou Jin-Hai Huang A-Yong Yu Repeatability and reproducibility of a double-pass optical quality analysis device. PLoS ONE |
title | Repeatability and reproducibility of a double-pass optical quality analysis device. |
title_full | Repeatability and reproducibility of a double-pass optical quality analysis device. |
title_fullStr | Repeatability and reproducibility of a double-pass optical quality analysis device. |
title_full_unstemmed | Repeatability and reproducibility of a double-pass optical quality analysis device. |
title_short | Repeatability and reproducibility of a double-pass optical quality analysis device. |
title_sort | repeatability and reproducibility of a double pass optical quality analysis device |
url | http://europepmc.org/articles/PMC4315520?pdf=render |
work_keys_str_mv | AT chenchenxu repeatabilityandreproducibilityofadoublepassopticalqualityanalysisdevice AT tingxue repeatabilityandreproducibilityofadoublepassopticalqualityanalysisdevice AT qinmeiwang repeatabilityandreproducibilityofadoublepassopticalqualityanalysisdevice AT yinizhou repeatabilityandreproducibilityofadoublepassopticalqualityanalysisdevice AT jinhaihuang repeatabilityandreproducibilityofadoublepassopticalqualityanalysisdevice AT ayongyu repeatabilityandreproducibilityofadoublepassopticalqualityanalysisdevice |