Low Temperature Microcalorimeters for Decay Energy Spectroscopy

Low Temperature Detectors have been used to measure embedded radioisotopes in a measurement mode known as Decay Energy Spectroscopy (DES) since 1992. DES microcalorimeter measurements have been used for applications ranging from neutrino mass measurements to metrology to measurements for safeguards...

Full description

Bibliographic Details
Main Author: Katrina E. Koehler
Format: Article
Language:English
Published: MDPI AG 2021-04-01
Series:Applied Sciences
Subjects:
Online Access:https://www.mdpi.com/2076-3417/11/9/4044
Description
Summary:Low Temperature Detectors have been used to measure embedded radioisotopes in a measurement mode known as Decay Energy Spectroscopy (DES) since 1992. DES microcalorimeter measurements have been used for applications ranging from neutrino mass measurements to metrology to measurements for safeguards and medical nuclides. While the low temperature detectors have extremely high intrinsic energy resolution (several times better than semiconductor detectors), the energy resolution achieved in practice is strongly dependent on factors such as sample preparation method. This review seeks to present the literature consensus on what has been learned by looking at the energy resolution as a function of various choices of detector, absorber, and sample preparation methods.
ISSN:2076-3417