In vitro Assessment of Electromagnetic Interference due to Low-Band RFID reader/writers on Active Implantable Medical Devices

Introduction: The electromagnetic interference (EMI) of active implantable medical devices (AIMDs) due to 125–135 kHz and HF (13.56MHz) radio-frequency identification (RFID) reader/writers are discussed based on in vitro experiments. The purpose of this paper is to clarify the detailed EMI occurrenc...

Full description

Bibliographic Details
Main Authors: Shunichi Futatsumori, PhD, Yoshifumi Kawamura, Takashi Hikage, PhD, Toshio Nojima, PhD, Ben Koike, Hiroshi Fujimoto, Takeshi Toyoshima
Format: Article
Language:English
Published: Wiley 2009-01-01
Series:Journal of Arrhythmia
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S1880427609800136
Description
Summary:Introduction: The electromagnetic interference (EMI) of active implantable medical devices (AIMDs) due to 125–135 kHz and HF (13.56MHz) radio-frequency identification (RFID) reader/writers are discussed based on in vitro experiments. The purpose of this paper is to clarify the detailed EMI occurrence conditions of both implantable cardiac pacemakers and implantable cardioverter defibrillators (ICDs) when used in proximity to commercial lowband RFID reader/writers. Methods and results: A total of 20 types of pacemakers and 7 types of ICDs from 9 manufacturers are used in the experiments. The tested RFID reader/writers are 4 types of 125–135 kHz antennas and 26 types of HF antennas. The test system is newly constructed based on the one previously proposed for the estimation of EMI due to mobile phones. The measured reactions are missing of pacing pulses and generation of asynchronous pulses for pacemaker functions and inappropriate tachyarrhythmia detection and delivery of therapy for ICDs. In addition, measured EMI characteristics strongly depend on intermittent period of RFID signals. However, all AIMDs tested return to normal operation once removed from RFID exposure. Conclusion: The detailed EMI characteristics of AIMDs and low-band RFID reader/ writers are obtained to prevent adverse interactions.
ISSN:1880-4276