A Design Scheme for 1H/31P Dual-Nuclear Parallel MRI Coil
Based on the principle of birdcage coil and the decoupling theory of array coil, a design scheme for dual-nuclear magnetic resonance imaging (MRI) coil is proposed. A LC parallel trap is used to improve the adaptability of the inductive decoupling scheme. According to the design scheme and simulatio...
Main Authors: | , , , , , , , |
---|---|
Format: | Article |
Language: | zho |
Published: |
Science Press
2020-09-01
|
Series: | Chinese Journal of Magnetic Resonance |
Subjects: | |
Online Access: | http://121.43.60.238/bpxzz/EN/10.11938/cjmr20192737 |
Summary: | Based on the principle of birdcage coil and the decoupling theory of array coil, a design scheme for dual-nuclear magnetic resonance imaging (MRI) coil is proposed. A LC parallel trap is used to improve the adaptability of the inductive decoupling scheme. According to the design scheme and simulation, a 1H/31P dual-nuclear parallel MRI coil is fabricated for a 4.7 T system. Dual-nuclear parallel imaging experiments are carried out on a custom-built MRI system to verify the feasibility of the coil design. |
---|---|
ISSN: | 1000-4556 1000-4556 |