A Design Scheme for 1H/31P Dual-Nuclear Parallel MRI Coil

Based on the principle of birdcage coil and the decoupling theory of array coil, a design scheme for dual-nuclear magnetic resonance imaging (MRI) coil is proposed. A LC parallel trap is used to improve the adaptability of the inductive decoupling scheme. According to the design scheme and simulatio...

Full description

Bibliographic Details
Main Authors: LIAO Zhi-wen, CHEN Jun-fei, YANG Chun-sheng, ZHANG Zhi, CHEN Li, XIAO Li-zhi, CHEN Fang, LIU Chao-yang
Format: Article
Language:zho
Published: Science Press 2020-09-01
Series:Chinese Journal of Magnetic Resonance
Subjects:
Online Access:http://121.43.60.238/bpxzz/EN/10.11938/cjmr20192737
Description
Summary:Based on the principle of birdcage coil and the decoupling theory of array coil, a design scheme for dual-nuclear magnetic resonance imaging (MRI) coil is proposed. A LC parallel trap is used to improve the adaptability of the inductive decoupling scheme. According to the design scheme and simulation, a 1H/31P dual-nuclear parallel MRI coil is fabricated for a 4.7 T system. Dual-nuclear parallel imaging experiments are carried out on a custom-built MRI system to verify the feasibility of the coil design.
ISSN:1000-4556
1000-4556