Analytical Performance Evaluation of the New GEM<sup>®</sup> Premier™ 5000 in Comparison to the Epoc<sup>®</sup> Blood Gas Analyzer in Horses

Different blood gas analyzers are used in equine practice. Every machine needs to be validated, as they have not been designed for use in horses. The aim of this study was to compare the newly marketed GEM5000 machine to the formerly validated epoc machine for blood gas analysis in horses. In this p...

Full description

Bibliographic Details
Main Authors: Charlotte Sandersen, Petra Dmitrovic, Julien Dupont, Carla Cesarini, Hugues Guyot, Didier Serteyn, Katharina Kirsch
Format: Article
Language:English
Published: MDPI AG 2023-02-01
Series:Veterinary Sciences
Subjects:
Online Access:https://www.mdpi.com/2306-7381/10/2/114
_version_ 1797617854227415040
author Charlotte Sandersen
Petra Dmitrovic
Julien Dupont
Carla Cesarini
Hugues Guyot
Didier Serteyn
Katharina Kirsch
author_facet Charlotte Sandersen
Petra Dmitrovic
Julien Dupont
Carla Cesarini
Hugues Guyot
Didier Serteyn
Katharina Kirsch
author_sort Charlotte Sandersen
collection DOAJ
description Different blood gas analyzers are used in equine practice. Every machine needs to be validated, as they have not been designed for use in horses. The aim of this study was to compare the newly marketed GEM5000 machine to the formerly validated epoc machine for blood gas analysis in horses. In this prospective, non-blinded, comparative laboratory analyzer study, 43 equine blood samples were analyzed on both analyzers and values were compared between the two machines via Lin’s concordance analysis, Passing–Bablok regression analysis and Bland–Altman plots. Duplicate measurements were conducted on the GEM5000 machine to evaluate precision. The GEM5000 failed to achieve the required precision for tHb, Hct and iCa<sup>2+</sup>, but achieved acceptable precision for all other parameters. Concordance correlation analysis revealed poor correlation for Na<sup>+</sup>, Cl<sup>−</sup>, iCa<sup>2+</sup>, K<sup>+</sup>, Hct and tHb, while there was an at least moderate agreement for all other parameters. Passing–Bablok regression revealed significant constant bias for pCO<sub>2</sub>, pO<sub>2</sub>, Cl<sup>−</sup>, and iCa<sup>2+</sup> and significant proportional bias for pCO<sub>2</sub>, iCa<sup>2+</sup> and SO<sub>2</sub>. Bland–Altman analysis revealed significant systematic bias for Na<sup>+</sup>, Cl<sup>−</sup>, iCa<sup>2+</sup>, K<sup>+</sup>, Hct, tHb and SO<sub>2</sub>. This study shows that while precision of the GEM5000 is good, values should not be used interchangeably with data obtained from other blood gas analyzers.
first_indexed 2024-03-11T08:01:49Z
format Article
id doaj.art-7862a7fb9092485795cd15f2b9b2620c
institution Directory Open Access Journal
issn 2306-7381
language English
last_indexed 2024-03-11T08:01:49Z
publishDate 2023-02-01
publisher MDPI AG
record_format Article
series Veterinary Sciences
spelling doaj.art-7862a7fb9092485795cd15f2b9b2620c2023-11-16T23:45:49ZengMDPI AGVeterinary Sciences2306-73812023-02-0110211410.3390/vetsci10020114Analytical Performance Evaluation of the New GEM<sup>®</sup> Premier™ 5000 in Comparison to the Epoc<sup>®</sup> Blood Gas Analyzer in HorsesCharlotte Sandersen0Petra Dmitrovic1Julien Dupont2Carla Cesarini3Hugues Guyot4Didier Serteyn5Katharina Kirsch6Department of Clinical Sciences, Faculty of Veterinary Medicine, University of Liege, 4000 Liege, BelgiumDepartment of Clinical Sciences, Faculty of Veterinary Medicine, University of Liege, 4000 Liege, BelgiumDepartment of Clinical Sciences, Faculty of Veterinary Medicine, University of Liege, 4000 Liege, BelgiumDepartment of Clinical Sciences, Faculty of Veterinary Medicine, University of Liege, 4000 Liege, BelgiumDepartment of Clinical Sciences, Faculty of Veterinary Medicine, University of Liege, 4000 Liege, BelgiumDepartment of Clinical Sciences, Faculty of Veterinary Medicine, University of Liege, 4000 Liege, BelgiumDepartment of Clinical Sciences, Faculty of Veterinary Medicine, University of Liege, 4000 Liege, BelgiumDifferent blood gas analyzers are used in equine practice. Every machine needs to be validated, as they have not been designed for use in horses. The aim of this study was to compare the newly marketed GEM5000 machine to the formerly validated epoc machine for blood gas analysis in horses. In this prospective, non-blinded, comparative laboratory analyzer study, 43 equine blood samples were analyzed on both analyzers and values were compared between the two machines via Lin’s concordance analysis, Passing–Bablok regression analysis and Bland–Altman plots. Duplicate measurements were conducted on the GEM5000 machine to evaluate precision. The GEM5000 failed to achieve the required precision for tHb, Hct and iCa<sup>2+</sup>, but achieved acceptable precision for all other parameters. Concordance correlation analysis revealed poor correlation for Na<sup>+</sup>, Cl<sup>−</sup>, iCa<sup>2+</sup>, K<sup>+</sup>, Hct and tHb, while there was an at least moderate agreement for all other parameters. Passing–Bablok regression revealed significant constant bias for pCO<sub>2</sub>, pO<sub>2</sub>, Cl<sup>−</sup>, and iCa<sup>2+</sup> and significant proportional bias for pCO<sub>2</sub>, iCa<sup>2+</sup> and SO<sub>2</sub>. Bland–Altman analysis revealed significant systematic bias for Na<sup>+</sup>, Cl<sup>−</sup>, iCa<sup>2+</sup>, K<sup>+</sup>, Hct, tHb and SO<sub>2</sub>. This study shows that while precision of the GEM5000 is good, values should not be used interchangeably with data obtained from other blood gas analyzers.https://www.mdpi.com/2306-7381/10/2/114blood gas analysishorsePassing–Bablok analysisBland–Altman analysismethod comparison
spellingShingle Charlotte Sandersen
Petra Dmitrovic
Julien Dupont
Carla Cesarini
Hugues Guyot
Didier Serteyn
Katharina Kirsch
Analytical Performance Evaluation of the New GEM<sup>®</sup> Premier™ 5000 in Comparison to the Epoc<sup>®</sup> Blood Gas Analyzer in Horses
Veterinary Sciences
blood gas analysis
horse
Passing–Bablok analysis
Bland–Altman analysis
method comparison
title Analytical Performance Evaluation of the New GEM<sup>®</sup> Premier™ 5000 in Comparison to the Epoc<sup>®</sup> Blood Gas Analyzer in Horses
title_full Analytical Performance Evaluation of the New GEM<sup>®</sup> Premier™ 5000 in Comparison to the Epoc<sup>®</sup> Blood Gas Analyzer in Horses
title_fullStr Analytical Performance Evaluation of the New GEM<sup>®</sup> Premier™ 5000 in Comparison to the Epoc<sup>®</sup> Blood Gas Analyzer in Horses
title_full_unstemmed Analytical Performance Evaluation of the New GEM<sup>®</sup> Premier™ 5000 in Comparison to the Epoc<sup>®</sup> Blood Gas Analyzer in Horses
title_short Analytical Performance Evaluation of the New GEM<sup>®</sup> Premier™ 5000 in Comparison to the Epoc<sup>®</sup> Blood Gas Analyzer in Horses
title_sort analytical performance evaluation of the new gem sup r sup premier™ 5000 in comparison to the epoc sup r sup blood gas analyzer in horses
topic blood gas analysis
horse
Passing–Bablok analysis
Bland–Altman analysis
method comparison
url https://www.mdpi.com/2306-7381/10/2/114
work_keys_str_mv AT charlottesandersen analyticalperformanceevaluationofthenewgemsupsuppremier5000incomparisontotheepocsupsupbloodgasanalyzerinhorses
AT petradmitrovic analyticalperformanceevaluationofthenewgemsupsuppremier5000incomparisontotheepocsupsupbloodgasanalyzerinhorses
AT juliendupont analyticalperformanceevaluationofthenewgemsupsuppremier5000incomparisontotheepocsupsupbloodgasanalyzerinhorses
AT carlacesarini analyticalperformanceevaluationofthenewgemsupsuppremier5000incomparisontotheepocsupsupbloodgasanalyzerinhorses
AT huguesguyot analyticalperformanceevaluationofthenewgemsupsuppremier5000incomparisontotheepocsupsupbloodgasanalyzerinhorses
AT didierserteyn analyticalperformanceevaluationofthenewgemsupsuppremier5000incomparisontotheepocsupsupbloodgasanalyzerinhorses
AT katharinakirsch analyticalperformanceevaluationofthenewgemsupsuppremier5000incomparisontotheepocsupsupbloodgasanalyzerinhorses