Evaluation of effect of aneurysm model material on coil contact force and catheter movement

To ensure safe coil embolization for cerebral aneurysms, it is important to investigate the contact force between a coil and an aneurysm wall. Therefore, in our previous study, we developed an experimental system for measuring the contact force between a coil and an aneurysm biomodel. However, becau...

Full description

Bibliographic Details
Main Authors: Takuya OISHI, Kazuto TAKASHIMA, Kiyoshi YOSHINAKA, Kaihong YU, Makoto OHTA, Koji MORI, Naoki TOMA
Format: Article
Language:English
Published: The Japan Society of Mechanical Engineers 2021-09-01
Series:Journal of Biomechanical Science and Engineering
Subjects:
Online Access:https://www.jstage.jst.go.jp/article/jbse/17/1/17_21-00261/_pdf/-char/en
_version_ 1811225556757250048
author Takuya OISHI
Kazuto TAKASHIMA
Kiyoshi YOSHINAKA
Kaihong YU
Makoto OHTA
Koji MORI
Naoki TOMA
author_facet Takuya OISHI
Kazuto TAKASHIMA
Kiyoshi YOSHINAKA
Kaihong YU
Makoto OHTA
Koji MORI
Naoki TOMA
author_sort Takuya OISHI
collection DOAJ
description To ensure safe coil embolization for cerebral aneurysms, it is important to investigate the contact force between a coil and an aneurysm wall. Therefore, in our previous study, we developed an experimental system for measuring the contact force between a coil and an aneurysm biomodel. However, because the aneurysm model was made of silicone rubber, its physical properties, such as the friction coefficient, differed from those of aneurysms in vivo. Therefore, in this study, we made an aneurysm model using poly (vinyl alcohol) hydrogel (PVA-H) and evaluated the effect of the model material on the contact force. In addition, we acquired images during the experiment and evaluated the behavior of the coil and catheter and the relationship between the catheter movement and the contact force. A commercially available coil was inserted through a catheter into two types of aneurysm models (silicone rubber and PVA-H), and the effects of the model material and the catheter tip position (near dome, at the center, and near neck) were evaluated. The contact force and the total movement of the catheter tip for the PVA-H model were smaller than those for the silicone rubber model. The contact force (total catheter movement) was greater (smaller) when the catheter tip was inserted deeper into the silicone rubber model. These results suggest that the state of contact between the aneurysm and the coil affects the contact force and the catheter movement and that these two values are related.
first_indexed 2024-04-12T09:09:18Z
format Article
id doaj.art-79bc9214a604491f9c2577410c421584
institution Directory Open Access Journal
issn 1880-9863
language English
last_indexed 2024-04-12T09:09:18Z
publishDate 2021-09-01
publisher The Japan Society of Mechanical Engineers
record_format Article
series Journal of Biomechanical Science and Engineering
spelling doaj.art-79bc9214a604491f9c2577410c4215842022-12-22T03:39:00ZengThe Japan Society of Mechanical EngineersJournal of Biomechanical Science and Engineering1880-98632021-09-0117121-0026121-0026110.1299/jbse.21-00261jbseEvaluation of effect of aneurysm model material on coil contact force and catheter movementTakuya OISHI0Kazuto TAKASHIMA1Kiyoshi YOSHINAKA2Kaihong YU3Makoto OHTA4Koji MORI5Naoki TOMA6Graduate School of Life Science and Systems Engineering, Kyushu Institute of TechnologyGraduate School of Life Science and Systems Engineering, Kyushu Institute of TechnologyNational Institute of Advanced Industrial Science and TechnologyInstitute of Fluid Science, Tohoku UniversityInstitute of Fluid Science, Tohoku UniversityGraduate School of Sciences and Technology for Innovation, Yamaguchi UniversityGraduate School of Medicine, Mie UniversityTo ensure safe coil embolization for cerebral aneurysms, it is important to investigate the contact force between a coil and an aneurysm wall. Therefore, in our previous study, we developed an experimental system for measuring the contact force between a coil and an aneurysm biomodel. However, because the aneurysm model was made of silicone rubber, its physical properties, such as the friction coefficient, differed from those of aneurysms in vivo. Therefore, in this study, we made an aneurysm model using poly (vinyl alcohol) hydrogel (PVA-H) and evaluated the effect of the model material on the contact force. In addition, we acquired images during the experiment and evaluated the behavior of the coil and catheter and the relationship between the catheter movement and the contact force. A commercially available coil was inserted through a catheter into two types of aneurysm models (silicone rubber and PVA-H), and the effects of the model material and the catheter tip position (near dome, at the center, and near neck) were evaluated. The contact force and the total movement of the catheter tip for the PVA-H model were smaller than those for the silicone rubber model. The contact force (total catheter movement) was greater (smaller) when the catheter tip was inserted deeper into the silicone rubber model. These results suggest that the state of contact between the aneurysm and the coil affects the contact force and the catheter movement and that these two values are related.https://www.jstage.jst.go.jp/article/jbse/17/1/17_21-00261/_pdf/-char/enendovascular treatmentcerebral aneurysmcoil embolizationcontact forcecatheterphantomfirst coil
spellingShingle Takuya OISHI
Kazuto TAKASHIMA
Kiyoshi YOSHINAKA
Kaihong YU
Makoto OHTA
Koji MORI
Naoki TOMA
Evaluation of effect of aneurysm model material on coil contact force and catheter movement
Journal of Biomechanical Science and Engineering
endovascular treatment
cerebral aneurysm
coil embolization
contact force
catheter
phantom
first coil
title Evaluation of effect of aneurysm model material on coil contact force and catheter movement
title_full Evaluation of effect of aneurysm model material on coil contact force and catheter movement
title_fullStr Evaluation of effect of aneurysm model material on coil contact force and catheter movement
title_full_unstemmed Evaluation of effect of aneurysm model material on coil contact force and catheter movement
title_short Evaluation of effect of aneurysm model material on coil contact force and catheter movement
title_sort evaluation of effect of aneurysm model material on coil contact force and catheter movement
topic endovascular treatment
cerebral aneurysm
coil embolization
contact force
catheter
phantom
first coil
url https://www.jstage.jst.go.jp/article/jbse/17/1/17_21-00261/_pdf/-char/en
work_keys_str_mv AT takuyaoishi evaluationofeffectofaneurysmmodelmaterialoncoilcontactforceandcathetermovement
AT kazutotakashima evaluationofeffectofaneurysmmodelmaterialoncoilcontactforceandcathetermovement
AT kiyoshiyoshinaka evaluationofeffectofaneurysmmodelmaterialoncoilcontactforceandcathetermovement
AT kaihongyu evaluationofeffectofaneurysmmodelmaterialoncoilcontactforceandcathetermovement
AT makotoohta evaluationofeffectofaneurysmmodelmaterialoncoilcontactforceandcathetermovement
AT kojimori evaluationofeffectofaneurysmmodelmaterialoncoilcontactforceandcathetermovement
AT naokitoma evaluationofeffectofaneurysmmodelmaterialoncoilcontactforceandcathetermovement