P122 Unrealistic Pulse Pressures from Calibrated Tonometric Waveforms

Abstract Background Calibration of carotid artery tonometry waveforms needs to account for pulse pressure augmentation [1]. Due to its location nearer to the heart, carotid pulse pressure (carPP) is about 10 mmHg lower than brachial (brachPP) [1]. In practice, perfect applanation is hard to achieve...

Full description

Bibliographic Details
Main Authors: Koen Reesink, Louise Marais, Pierre Boutouyrie
Format: Article
Language:English
Published: BMC 2020-02-01
Series:Artery Research
Online Access:https://doi.org/10.2991/artres.k-191224.148
_version_ 1797275289511788544
author Koen Reesink
Louise Marais
Pierre Boutouyrie
author_facet Koen Reesink
Louise Marais
Pierre Boutouyrie
author_sort Koen Reesink
collection DOAJ
description Abstract Background Calibration of carotid artery tonometry waveforms needs to account for pulse pressure augmentation [1]. Due to its location nearer to the heart, carotid pulse pressure (carPP) is about 10 mmHg lower than brachial (brachPP) [1]. In practice, perfect applanation is hard to achieve with tonometry and thus tracking-errors may occur, leading to an exaggerated fall in the diastolic limb of the tonometry waveform. Consequently, the mean-to-diastolic difference of the waveform is underestimated, leading to an overestimation of carPP after calibration. We assessed the relationship between brachPP and carPP in 100 subjects, aged 18–80 yrs. Methods and Results Brachial pressure (Omron) and carotid tonometry (Sphygmocor) measurements were performed by a single, highly skilled vascular technician, according to Artery Society methodological standards. CarPP was 5 ± 13 mmHg lower than brachPP. Bland-Altman analysis revealed a positive trend (r = 0.695, p < 0.0001) between the carPP–brachPP difference and their mean, with carPP being 20 mmHg lower than brachPP at a mean 40 mmHg which extended to carPP being 20 mmHg higher(!) than brachPP at a mean 80 mmHg. Accordingly, 30% of the study population showed unphysiological carPP, indicating an unrealistic i.e. reverse amplification. Even with a tolerance of +5 mmHg, still 20% of cases showed unrealistic values. Conclusion In practice, calibrated carotid tonometry waveforms may yield unphysiological overestimations of carotid pulse pressure. This finding implies (potential) estimation errors in regression model studies on central pressure and local stiffness, suggesting brachial pulse pressure would be a valid compromise.
first_indexed 2024-03-07T15:12:08Z
format Article
id doaj.art-830d75298c0a4e56a90b058e7a2e3eb6
institution Directory Open Access Journal
issn 1872-9312
1876-4401
language English
last_indexed 2024-03-07T15:12:08Z
publishDate 2020-02-01
publisher BMC
record_format Article
series Artery Research
spelling doaj.art-830d75298c0a4e56a90b058e7a2e3eb62024-03-05T18:14:42ZengBMCArtery Research1872-93121876-44012020-02-0125S1S162S16210.2991/artres.k-191224.148P122 Unrealistic Pulse Pressures from Calibrated Tonometric WaveformsKoen Reesink0Louise Marais1Pierre Boutouyrie2CARIM School of Cardiovascular Diseases, Maastricht University Medical CenterINSERMINSERMAbstract Background Calibration of carotid artery tonometry waveforms needs to account for pulse pressure augmentation [1]. Due to its location nearer to the heart, carotid pulse pressure (carPP) is about 10 mmHg lower than brachial (brachPP) [1]. In practice, perfect applanation is hard to achieve with tonometry and thus tracking-errors may occur, leading to an exaggerated fall in the diastolic limb of the tonometry waveform. Consequently, the mean-to-diastolic difference of the waveform is underestimated, leading to an overestimation of carPP after calibration. We assessed the relationship between brachPP and carPP in 100 subjects, aged 18–80 yrs. Methods and Results Brachial pressure (Omron) and carotid tonometry (Sphygmocor) measurements were performed by a single, highly skilled vascular technician, according to Artery Society methodological standards. CarPP was 5 ± 13 mmHg lower than brachPP. Bland-Altman analysis revealed a positive trend (r = 0.695, p < 0.0001) between the carPP–brachPP difference and their mean, with carPP being 20 mmHg lower than brachPP at a mean 40 mmHg which extended to carPP being 20 mmHg higher(!) than brachPP at a mean 80 mmHg. Accordingly, 30% of the study population showed unphysiological carPP, indicating an unrealistic i.e. reverse amplification. Even with a tolerance of +5 mmHg, still 20% of cases showed unrealistic values. Conclusion In practice, calibrated carotid tonometry waveforms may yield unphysiological overestimations of carotid pulse pressure. This finding implies (potential) estimation errors in regression model studies on central pressure and local stiffness, suggesting brachial pulse pressure would be a valid compromise.https://doi.org/10.2991/artres.k-191224.148
spellingShingle Koen Reesink
Louise Marais
Pierre Boutouyrie
P122 Unrealistic Pulse Pressures from Calibrated Tonometric Waveforms
Artery Research
title P122 Unrealistic Pulse Pressures from Calibrated Tonometric Waveforms
title_full P122 Unrealistic Pulse Pressures from Calibrated Tonometric Waveforms
title_fullStr P122 Unrealistic Pulse Pressures from Calibrated Tonometric Waveforms
title_full_unstemmed P122 Unrealistic Pulse Pressures from Calibrated Tonometric Waveforms
title_short P122 Unrealistic Pulse Pressures from Calibrated Tonometric Waveforms
title_sort p122 unrealistic pulse pressures from calibrated tonometric waveforms
url https://doi.org/10.2991/artres.k-191224.148
work_keys_str_mv AT koenreesink p122unrealisticpulsepressuresfromcalibratedtonometricwaveforms
AT louisemarais p122unrealisticpulsepressuresfromcalibratedtonometricwaveforms
AT pierreboutouyrie p122unrealisticpulsepressuresfromcalibratedtonometricwaveforms