Evaluation of Dose Calculation Algorithms Accuracy for Eclipse, PCRT3D, and Monaco Treatment Planning Systems Using IAEA TPS commissioning tests in a Heterogeneous Phantom
Introduction: The accuracy of dose calculation algorithm (DCA) is highly considered in the radiotherapy sequences. This study aims at assessing the accuracy of five dose calculation algorithms in tissue inhomogeneity corrections, based on the International Atomic Energy Agency TEC-DOC 1583. Material...
Main Authors: | Nasibeh Kavousi, Hassan Ali Nedaie, Somayeh Gholami, mahbod Esfahani, Ghazale Geraily |
---|---|
Format: | Article |
Language: | English |
Published: |
Mashhad University of Medical Sciences
2019-07-01
|
Series: | Iranian Journal of Medical Physics |
Subjects: | |
Online Access: | http://ijmp.mums.ac.ir/article_11821_7eca03a32cd0ffccefffcbfe2c34b358.pdf |
Similar Items
-
Dosimetric Accuracy Comparison between ACUROSE XB, AAA and PBC Dose Calculation Algorithms in EclipseTM TPS Using a Heterogeneous Phantom
by: Raheleh Rooshenass, et al.
Published: (2018-12-01) -
Evaluation of dose uniformity in total body irradiation (TBI) technique in parallel-opposed anterior/posterior geometry using Varian Clinac 2100C/D linear accelerator
by: Elham Hoseinnezhad Zarghani, et al.
Published: (2020-10-01) -
Experimental Evaluation of Depth Dose by Exit Surface Diode Dosimeters for Off-Axis Wedged Fields in Radiation Therapy
by: Alireza Mohammadkarim, et al.
Published: (2016-04-01) -
An Investigation into the Surface Dose Using Eclipse Treatment Planning System and Film Dosimetry for Treatment of Breast Cancer
by: Nooshin Sheykholeslami, et al.
Published: (2023-01-01) -
The Minderoo-Monaco Commission on Plastics and Human Health
by: Philip J. Landrigan, et al.
Published: (2023-03-01)