Accuracy of the Cosmed K5 portable calorimeter.

PURPOSE:The purpose of this study was to assess the accuracy of the Cosmed K5 portable metabolic system dynamic mixing chamber (MC) and breath-by-breath (BxB) modes against the criterion Douglas bag (DB) method. METHODS:Fifteen participants (mean age±SD, 30.6±7.4 yrs) had their metabolic variables m...

Full description

Bibliographic Details
Main Authors: Scott E Crouter, Samuel R LaMunion, Paul R Hibbing, Andrew S Kaplan, David R Bassett
Format: Article
Language:English
Published: Public Library of Science (PLoS) 2019-01-01
Series:PLoS ONE
Online Access:https://doi.org/10.1371/journal.pone.0226290
_version_ 1818578044404629504
author Scott E Crouter
Samuel R LaMunion
Paul R Hibbing
Andrew S Kaplan
David R Bassett
author_facet Scott E Crouter
Samuel R LaMunion
Paul R Hibbing
Andrew S Kaplan
David R Bassett
author_sort Scott E Crouter
collection DOAJ
description PURPOSE:The purpose of this study was to assess the accuracy of the Cosmed K5 portable metabolic system dynamic mixing chamber (MC) and breath-by-breath (BxB) modes against the criterion Douglas bag (DB) method. METHODS:Fifteen participants (mean age±SD, 30.6±7.4 yrs) had their metabolic variables measured at rest and during cycling at 50, 100, 150, 200, and 250W. During each stage, participants were connected to the first respiratory gas collection method (randomized) for the first four minutes to reach steady state, followed by 3-min (or 5-min for DB) collection periods for the resting condition, and 2-min collection periods for all cycling intensities. Collection periods for the second and third methods were preceded by a washout of 1-3 min. Repeated measures ANOVAs were used to compare metabolic variables measured by each method, for seated rest and each cycling work rate. RESULTS:For ventilation (VE) and oxygen uptake (VO2), the K5 MC and BxB modes were within 2.1 l/min (VE) and 0.08 l/min (VO2) of the DB (p≥0.05). Compared to DB values, carbon dioxide production (VCO2) was significantly underestimated by the K5 BxB mode at work rates ≥150W by 0.12-0.31 l/min (p<0.05). K5 MC and BxB respiratory exchange ratio values were significantly lower than DB at cycling work rates ≥100W by 0.03-0.08 (p<0.05). CONCLUSION:Compared to the DB method, the K5 MC and BxB modes are acceptable for measuring VE and VO2 across a wide range of cycling intensities. Both K5 modes provided comparable values to each other.
first_indexed 2024-12-16T06:39:33Z
format Article
id doaj.art-8d0e28b23ce84daabd72730c0ed69e72
institution Directory Open Access Journal
issn 1932-6203
language English
last_indexed 2024-12-16T06:39:33Z
publishDate 2019-01-01
publisher Public Library of Science (PLoS)
record_format Article
series PLoS ONE
spelling doaj.art-8d0e28b23ce84daabd72730c0ed69e722022-12-21T22:40:44ZengPublic Library of Science (PLoS)PLoS ONE1932-62032019-01-011412e022629010.1371/journal.pone.0226290Accuracy of the Cosmed K5 portable calorimeter.Scott E CrouterSamuel R LaMunionPaul R HibbingAndrew S KaplanDavid R BassettPURPOSE:The purpose of this study was to assess the accuracy of the Cosmed K5 portable metabolic system dynamic mixing chamber (MC) and breath-by-breath (BxB) modes against the criterion Douglas bag (DB) method. METHODS:Fifteen participants (mean age±SD, 30.6±7.4 yrs) had their metabolic variables measured at rest and during cycling at 50, 100, 150, 200, and 250W. During each stage, participants were connected to the first respiratory gas collection method (randomized) for the first four minutes to reach steady state, followed by 3-min (or 5-min for DB) collection periods for the resting condition, and 2-min collection periods for all cycling intensities. Collection periods for the second and third methods were preceded by a washout of 1-3 min. Repeated measures ANOVAs were used to compare metabolic variables measured by each method, for seated rest and each cycling work rate. RESULTS:For ventilation (VE) and oxygen uptake (VO2), the K5 MC and BxB modes were within 2.1 l/min (VE) and 0.08 l/min (VO2) of the DB (p≥0.05). Compared to DB values, carbon dioxide production (VCO2) was significantly underestimated by the K5 BxB mode at work rates ≥150W by 0.12-0.31 l/min (p<0.05). K5 MC and BxB respiratory exchange ratio values were significantly lower than DB at cycling work rates ≥100W by 0.03-0.08 (p<0.05). CONCLUSION:Compared to the DB method, the K5 MC and BxB modes are acceptable for measuring VE and VO2 across a wide range of cycling intensities. Both K5 modes provided comparable values to each other.https://doi.org/10.1371/journal.pone.0226290
spellingShingle Scott E Crouter
Samuel R LaMunion
Paul R Hibbing
Andrew S Kaplan
David R Bassett
Accuracy of the Cosmed K5 portable calorimeter.
PLoS ONE
title Accuracy of the Cosmed K5 portable calorimeter.
title_full Accuracy of the Cosmed K5 portable calorimeter.
title_fullStr Accuracy of the Cosmed K5 portable calorimeter.
title_full_unstemmed Accuracy of the Cosmed K5 portable calorimeter.
title_short Accuracy of the Cosmed K5 portable calorimeter.
title_sort accuracy of the cosmed k5 portable calorimeter
url https://doi.org/10.1371/journal.pone.0226290
work_keys_str_mv AT scottecrouter accuracyofthecosmedk5portablecalorimeter
AT samuelrlamunion accuracyofthecosmedk5portablecalorimeter
AT paulrhibbing accuracyofthecosmedk5portablecalorimeter
AT andrewskaplan accuracyofthecosmedk5portablecalorimeter
AT davidrbassett accuracyofthecosmedk5portablecalorimeter