İki farklı apeks bulucunun kök kanal boyu ölçümünde doğruluğunun ex vivo olarak değerlendirilmesi

Purpose: The aim of this ex vivo study was to evaluate the accuracy of two apex locators: Propex (Dentsply, Ballaigues Switzerland) and RomiAPEX™ (Romidan, Tel Aviv, Israel). Materials and methods: The accuracy of the measurement of two different electronic apex loctors were compared in root canal l...

Full description

Bibliographic Details
Main Authors: Hale Cimilli, Nevin Kartal, Seda Aydemir
Format: Article
Language:English
Published: Cumhuriyet University 2011-10-01
Series:Cumhuriyet Dental Journal
Subjects:
Online Access:https://dergipark.org.tr/tr/download/article-file/48268
_version_ 1797703800868306944
author Hale Cimilli
Nevin Kartal
Seda Aydemir
author_facet Hale Cimilli
Nevin Kartal
Seda Aydemir
author_sort Hale Cimilli
collection DOAJ
description Purpose: The aim of this ex vivo study was to evaluate the accuracy of two apex locators: Propex (Dentsply, Ballaigues Switzerland) and RomiAPEX™ (Romidan, Tel Aviv, Israel). Materials and methods: The accuracy of the measurement of two different electronic apex loctors were compared in root canal length determination by using forty extracted human teeth . The actual canal length was determined by introducing a size 15 K-file in the canal until the file tip became visible at the apical foramen under X5 magnification by using an operating microscope. A caliper to the nearest 0.5 mm were used during measurements. The teeth were embedded in acrylic as the apical one thirds and crowns were apparent. The measurements were taken as the apical one thirds embedded in alginate by using two electronic apex locators. All measurements were taken two times. Results: The accuracy of the electronical working length measurements were found for Propex as 97.5% ; for RomiAPEX™ D-30 as 95%. No significant differences were found between actual root canal lengths and the measurement values of apex locators statistically (p>0,05). Conclusion: High accuracy were confirmed for two apex locators during the measurements of root canal lengths. ÖZET Amaç: Bu ex vivo çalışmada Propex (Dentsply, Ballaigues, İsviçre) ve RomiAPEX™ D-30 (Romidan, Tel Aviv, İsrail) isimli iki farklı apeks bulucunun klinik doğruluğunun değerlendirilmesi amaçlanmıtır. Gereç ve Yöntem: Bu çalımada kırk adet üst kesici diş kullanıldı. İki farklı apeks bulucu (Propex ve RomiAPEX™ D-30) kullanıldı. Dişlerin gerçek kanal boyları operasyon mikroskobunda X5 büyütmede #15 K tipi eğe ucu apikal foramenden görülene kadar ilerletilerek ölçüldü. Ölçümler sırasında 0.5 mm aralıklı kaliper kullanıldı. Dişler kronları ve apikal 1/3’leri açıkta kalacak şekilde akriliğe gömüldü. Ardından dilerin apikal 1/3’lük kısımları aljinata gömüldü ve iki apeks bulucuyla ölçümler yapıldı. Tüm ölçümler ikişer defa tekrarlandı. Bulgular: Dişlerin elektronik kök kanal boyu ölçüm değerleri ortalaması gerçek kanal boyuyla karşılatırıldığı zaman ± 0.5 mm aralığında Propex için ölçüm değeri doğruluk oranı %97.5; RomiAPEX™ D-30 için %95 olarak hesaplandı. Gerçek kök kanal boyu ile elektronik apeks bulucuların ölçüm değerleri arasında istatistiksel olarak anlamlı bir farklılık bulunamadı (p>0,05). Sonuç: Her iki apeks bulucunun da kök kanal boyu ölçümünde yüksek oranda doğruluk gösterdiği tespit edildi.
first_indexed 2024-03-12T05:10:48Z
format Article
id doaj.art-8f8c44aacc504a6ebe40684cdce96655
institution Directory Open Access Journal
issn 1302-5805
2146-2852
language English
last_indexed 2024-03-12T05:10:48Z
publishDate 2011-10-01
publisher Cumhuriyet University
record_format Article
series Cumhuriyet Dental Journal
spelling doaj.art-8f8c44aacc504a6ebe40684cdce966552023-09-03T08:41:31ZengCumhuriyet UniversityCumhuriyet Dental Journal1302-58052146-28522011-10-011221251282057İki farklı apeks bulucunun kök kanal boyu ölçümünde doğruluğunun ex vivo olarak değerlendirilmesiHale CimilliNevin KartalSeda AydemirPurpose: The aim of this ex vivo study was to evaluate the accuracy of two apex locators: Propex (Dentsply, Ballaigues Switzerland) and RomiAPEX™ (Romidan, Tel Aviv, Israel). Materials and methods: The accuracy of the measurement of two different electronic apex loctors were compared in root canal length determination by using forty extracted human teeth . The actual canal length was determined by introducing a size 15 K-file in the canal until the file tip became visible at the apical foramen under X5 magnification by using an operating microscope. A caliper to the nearest 0.5 mm were used during measurements. The teeth were embedded in acrylic as the apical one thirds and crowns were apparent. The measurements were taken as the apical one thirds embedded in alginate by using two electronic apex locators. All measurements were taken two times. Results: The accuracy of the electronical working length measurements were found for Propex as 97.5% ; for RomiAPEX™ D-30 as 95%. No significant differences were found between actual root canal lengths and the measurement values of apex locators statistically (p>0,05). Conclusion: High accuracy were confirmed for two apex locators during the measurements of root canal lengths. ÖZET Amaç: Bu ex vivo çalışmada Propex (Dentsply, Ballaigues, İsviçre) ve RomiAPEX™ D-30 (Romidan, Tel Aviv, İsrail) isimli iki farklı apeks bulucunun klinik doğruluğunun değerlendirilmesi amaçlanmıtır. Gereç ve Yöntem: Bu çalımada kırk adet üst kesici diş kullanıldı. İki farklı apeks bulucu (Propex ve RomiAPEX™ D-30) kullanıldı. Dişlerin gerçek kanal boyları operasyon mikroskobunda X5 büyütmede #15 K tipi eğe ucu apikal foramenden görülene kadar ilerletilerek ölçüldü. Ölçümler sırasında 0.5 mm aralıklı kaliper kullanıldı. Dişler kronları ve apikal 1/3’leri açıkta kalacak şekilde akriliğe gömüldü. Ardından dilerin apikal 1/3’lük kısımları aljinata gömüldü ve iki apeks bulucuyla ölçümler yapıldı. Tüm ölçümler ikişer defa tekrarlandı. Bulgular: Dişlerin elektronik kök kanal boyu ölçüm değerleri ortalaması gerçek kanal boyuyla karşılatırıldığı zaman ± 0.5 mm aralığında Propex için ölçüm değeri doğruluk oranı %97.5; RomiAPEX™ D-30 için %95 olarak hesaplandı. Gerçek kök kanal boyu ile elektronik apeks bulucuların ölçüm değerleri arasında istatistiksel olarak anlamlı bir farklılık bulunamadı (p>0,05). Sonuç: Her iki apeks bulucunun da kök kanal boyu ölçümünde yüksek oranda doğruluk gösterdiği tespit edildi.https://dergipark.org.tr/tr/download/article-file/48268apex locators; working length; root canalapeks bulucular çal ma boyu kök kanal
spellingShingle Hale Cimilli
Nevin Kartal
Seda Aydemir
İki farklı apeks bulucunun kök kanal boyu ölçümünde doğruluğunun ex vivo olarak değerlendirilmesi
Cumhuriyet Dental Journal
apex locators; working length; root canal
apeks bulucular
çal ma boyu
kök kanal
title İki farklı apeks bulucunun kök kanal boyu ölçümünde doğruluğunun ex vivo olarak değerlendirilmesi
title_full İki farklı apeks bulucunun kök kanal boyu ölçümünde doğruluğunun ex vivo olarak değerlendirilmesi
title_fullStr İki farklı apeks bulucunun kök kanal boyu ölçümünde doğruluğunun ex vivo olarak değerlendirilmesi
title_full_unstemmed İki farklı apeks bulucunun kök kanal boyu ölçümünde doğruluğunun ex vivo olarak değerlendirilmesi
title_short İki farklı apeks bulucunun kök kanal boyu ölçümünde doğruluğunun ex vivo olarak değerlendirilmesi
title_sort iki farkli apeks bulucunun kok kanal boyu olcumunde dogrulugunun ex vivo olarak degerlendirilmesi
topic apex locators; working length; root canal
apeks bulucular
çal ma boyu
kök kanal
url https://dergipark.org.tr/tr/download/article-file/48268
work_keys_str_mv AT halecimilli ikifarklıapeksbulucununkokkanalboyuolcumundedogrulugununexvivoolarakdegerlendirilmesi
AT nevinkartal ikifarklıapeksbulucununkokkanalboyuolcumundedogrulugununexvivoolarakdegerlendirilmesi
AT sedaaydemir ikifarklıapeksbulucununkokkanalboyuolcumundedogrulugununexvivoolarakdegerlendirilmesi