Evaluation of small-capacity force transducers using a developed 2 N dead-weight type force standard machine

A 2 N dead-weight type force standard machine (DWM) had been developed at the National Metrology Institute of Japan (NMIJ). Modifications of the force transducers and weight receivers of the DWM improved measurement reproducibility with rotational position change. Two modified small-capacity force t...

Cur síos iomlán

Sonraí bibleagrafaíochta
Príomhchruthaitheoirí: Junfang Zhu, Toshiyuki Hayashi, Koji Ogushi
Formáid: Alt
Teanga:English
Foilsithe / Cruthaithe: Taylor & Francis Group 2021-01-01
Sraith:SICE Journal of Control, Measurement, and System Integration
Ábhair:
Rochtain ar líne:http://dx.doi.org/10.1080/18824889.2020.1863613
Cur síos
Achoimre:A 2 N dead-weight type force standard machine (DWM) had been developed at the National Metrology Institute of Japan (NMIJ). Modifications of the force transducers and weight receivers of the DWM improved measurement reproducibility with rotational position change. Two modified small-capacity force transducers with rated capacities of 1 and 2 N could be precisely calibrated according to International Organization for Standardization (ISO) 376. The results showed quite smaller uncertainties in the calibration ranges of 100 mN to 1 N and 200 mN to 2 N than the catalogue accuracies of commercial transducers. The validity of the calibration capability of the 2 N DWM was confirmed by comparing it with a 50 N DWM. The results obtained by two DWMs coincided within the uncertainties of calibration.
ISSN:1884-9970