Evaluation of small-capacity force transducers using a developed 2 N dead-weight type force standard machine

A 2 N dead-weight type force standard machine (DWM) had been developed at the National Metrology Institute of Japan (NMIJ). Modifications of the force transducers and weight receivers of the DWM improved measurement reproducibility with rotational position change. Two modified small-capacity force t...

全面介绍

书目详细资料
Main Authors: Junfang Zhu, Toshiyuki Hayashi, Koji Ogushi
格式: 文件
语言:English
出版: Taylor & Francis Group 2021-01-01
丛编:SICE Journal of Control, Measurement, and System Integration
主题:
在线阅读:http://dx.doi.org/10.1080/18824889.2020.1863613
实物特征
总结:A 2 N dead-weight type force standard machine (DWM) had been developed at the National Metrology Institute of Japan (NMIJ). Modifications of the force transducers and weight receivers of the DWM improved measurement reproducibility with rotational position change. Two modified small-capacity force transducers with rated capacities of 1 and 2 N could be precisely calibrated according to International Organization for Standardization (ISO) 376. The results showed quite smaller uncertainties in the calibration ranges of 100 mN to 1 N and 200 mN to 2 N than the catalogue accuracies of commercial transducers. The validity of the calibration capability of the 2 N DWM was confirmed by comparing it with a 50 N DWM. The results obtained by two DWMs coincided within the uncertainties of calibration.