Does endo-tracheal tube clamping prevent air leaks and maintain positive end-expiratory pressure during the switching of a ventilator in a patient in an intensive care unit? A bench study.

<h4>Objectives</h4>When patients with acute respiratory distress syndrome are moved out of an intensive care unit, the ventilator often requires changing. This procedure suppresses positive end expiratory pressure and promotes lung derecruitment. Clamping the endotracheal tube may preven...

Full description

Bibliographic Details
Main Authors: Emanuele Turbil, Nicolas Terzi, Carole Schwebel, Martin Cour, Laurent Argaud, Claude Guérin
Format: Article
Language:English
Published: Public Library of Science (PLoS) 2020-01-01
Series:PLoS ONE
Online Access:https://doi.org/10.1371/journal.pone.0230147
_version_ 1828919166146445312
author Emanuele Turbil
Nicolas Terzi
Carole Schwebel
Martin Cour
Laurent Argaud
Claude Guérin
author_facet Emanuele Turbil
Nicolas Terzi
Carole Schwebel
Martin Cour
Laurent Argaud
Claude Guérin
author_sort Emanuele Turbil
collection DOAJ
description <h4>Objectives</h4>When patients with acute respiratory distress syndrome are moved out of an intensive care unit, the ventilator often requires changing. This procedure suppresses positive end expiratory pressure and promotes lung derecruitment. Clamping the endotracheal tube may prevent this from occurring. Whether or not such clamping maintains positive end-expiratory pressure has never been investigated. We designed a bench study to explore this further.<h4>How the study was done</h4>We used the Elysee 350 ventilator in 'volume controlled' mode with a positive end-expiratory pressure of 15 cmH2O, connected to an endotracheal tube with an 8 mm internal diameter inserted into a lung model with 40 ml/cmH2O compliance and 10 cmH2O/L/s resistance. We measured airway pressure and flow between the distal end of the endotracheal tube and the lung model. We tested a plastic, a metal, and an Extra Corporeal Membrane Oxygenation clamp, each with an oral/nasal, a nasal, and a reinforced endotracheal tube. We performed an end-expiratory hold then clamped the endotracheal tube and disconnected the ventilator. We measured the change in airway pressure and volume for 30 s following the disconnection of the ventilator.<h4>Results</h4>Airway pressure decreased thirty seconds after disconnection with all combinations of clamp and endotracheal tube. The largest fall in airway pressure (-17.486 cmH2O/s at 5 s and -18.834 cmH2O/s at 30 s) was observed with the plastic clamp combined with the reinforced endotracheal tube. The smallest decrease in airway pressure (0 cmH2O/s at 5 s and -0.163 cmH2O/s at 30 s) was observed using the Extra Corporeal Membrane Oxygenation clamp with the nasal endotracheal tube.<h4>Conclusions</h4>Only the Extra Corporeal Membrane Oxygenation clamp was efficient. Even with an Extra Corporeal Membrane Oxygenation clamp, it is important to limit the duration the ventilator is disconnected to a few seconds (ideally 5 s).
first_indexed 2024-12-13T21:22:22Z
format Article
id doaj.art-90cbe8cdeea24169b33863be637f5757
institution Directory Open Access Journal
issn 1932-6203
language English
last_indexed 2024-12-13T21:22:22Z
publishDate 2020-01-01
publisher Public Library of Science (PLoS)
record_format Article
series PLoS ONE
spelling doaj.art-90cbe8cdeea24169b33863be637f57572022-12-21T23:31:04ZengPublic Library of Science (PLoS)PLoS ONE1932-62032020-01-01153e023014710.1371/journal.pone.0230147Does endo-tracheal tube clamping prevent air leaks and maintain positive end-expiratory pressure during the switching of a ventilator in a patient in an intensive care unit? A bench study.Emanuele TurbilNicolas TerziCarole SchwebelMartin CourLaurent ArgaudClaude Guérin<h4>Objectives</h4>When patients with acute respiratory distress syndrome are moved out of an intensive care unit, the ventilator often requires changing. This procedure suppresses positive end expiratory pressure and promotes lung derecruitment. Clamping the endotracheal tube may prevent this from occurring. Whether or not such clamping maintains positive end-expiratory pressure has never been investigated. We designed a bench study to explore this further.<h4>How the study was done</h4>We used the Elysee 350 ventilator in 'volume controlled' mode with a positive end-expiratory pressure of 15 cmH2O, connected to an endotracheal tube with an 8 mm internal diameter inserted into a lung model with 40 ml/cmH2O compliance and 10 cmH2O/L/s resistance. We measured airway pressure and flow between the distal end of the endotracheal tube and the lung model. We tested a plastic, a metal, and an Extra Corporeal Membrane Oxygenation clamp, each with an oral/nasal, a nasal, and a reinforced endotracheal tube. We performed an end-expiratory hold then clamped the endotracheal tube and disconnected the ventilator. We measured the change in airway pressure and volume for 30 s following the disconnection of the ventilator.<h4>Results</h4>Airway pressure decreased thirty seconds after disconnection with all combinations of clamp and endotracheal tube. The largest fall in airway pressure (-17.486 cmH2O/s at 5 s and -18.834 cmH2O/s at 30 s) was observed with the plastic clamp combined with the reinforced endotracheal tube. The smallest decrease in airway pressure (0 cmH2O/s at 5 s and -0.163 cmH2O/s at 30 s) was observed using the Extra Corporeal Membrane Oxygenation clamp with the nasal endotracheal tube.<h4>Conclusions</h4>Only the Extra Corporeal Membrane Oxygenation clamp was efficient. Even with an Extra Corporeal Membrane Oxygenation clamp, it is important to limit the duration the ventilator is disconnected to a few seconds (ideally 5 s).https://doi.org/10.1371/journal.pone.0230147
spellingShingle Emanuele Turbil
Nicolas Terzi
Carole Schwebel
Martin Cour
Laurent Argaud
Claude Guérin
Does endo-tracheal tube clamping prevent air leaks and maintain positive end-expiratory pressure during the switching of a ventilator in a patient in an intensive care unit? A bench study.
PLoS ONE
title Does endo-tracheal tube clamping prevent air leaks and maintain positive end-expiratory pressure during the switching of a ventilator in a patient in an intensive care unit? A bench study.
title_full Does endo-tracheal tube clamping prevent air leaks and maintain positive end-expiratory pressure during the switching of a ventilator in a patient in an intensive care unit? A bench study.
title_fullStr Does endo-tracheal tube clamping prevent air leaks and maintain positive end-expiratory pressure during the switching of a ventilator in a patient in an intensive care unit? A bench study.
title_full_unstemmed Does endo-tracheal tube clamping prevent air leaks and maintain positive end-expiratory pressure during the switching of a ventilator in a patient in an intensive care unit? A bench study.
title_short Does endo-tracheal tube clamping prevent air leaks and maintain positive end-expiratory pressure during the switching of a ventilator in a patient in an intensive care unit? A bench study.
title_sort does endo tracheal tube clamping prevent air leaks and maintain positive end expiratory pressure during the switching of a ventilator in a patient in an intensive care unit a bench study
url https://doi.org/10.1371/journal.pone.0230147
work_keys_str_mv AT emanueleturbil doesendotrachealtubeclampingpreventairleaksandmaintainpositiveendexpiratorypressureduringtheswitchingofaventilatorinapatientinanintensivecareunitabenchstudy
AT nicolasterzi doesendotrachealtubeclampingpreventairleaksandmaintainpositiveendexpiratorypressureduringtheswitchingofaventilatorinapatientinanintensivecareunitabenchstudy
AT caroleschwebel doesendotrachealtubeclampingpreventairleaksandmaintainpositiveendexpiratorypressureduringtheswitchingofaventilatorinapatientinanintensivecareunitabenchstudy
AT martincour doesendotrachealtubeclampingpreventairleaksandmaintainpositiveendexpiratorypressureduringtheswitchingofaventilatorinapatientinanintensivecareunitabenchstudy
AT laurentargaud doesendotrachealtubeclampingpreventairleaksandmaintainpositiveendexpiratorypressureduringtheswitchingofaventilatorinapatientinanintensivecareunitabenchstudy
AT claudeguerin doesendotrachealtubeclampingpreventairleaksandmaintainpositiveendexpiratorypressureduringtheswitchingofaventilatorinapatientinanintensivecareunitabenchstudy