The effect of aging on the accuracy of new friction-style mechanical torque limiting devices for dental implants.

High variability in delivering the target torque is reported for friction-style mechanical torque limiting devices (F-S MTLDs). The effect of aging (number of use) on the accuracy of these devices is not clear. The purpose of this study was to assess the effect of aging on the accuracy (±10% of the...

Full description

Bibliographic Details
Main Authors: Aboulfazl Saboury, Seyed Jalil Sadr, Ali Fayaz, Minoo Mahshid
Format: Article
Language:English
Published: Tehran University of Medical Sciences 2013-01-01
Series:Frontiers in Dentistry
Subjects:
Online Access:https://jdt.tums.ac.ir/index.php/jdt/article/view/707
Description
Summary:High variability in delivering the target torque is reported for friction-style mechanical torque limiting devices (F-S MTLDs). The effect of aging (number of use) on the accuracy of these devices is not clear. The purpose of this study was to assess the effect of aging on the accuracy (±10% of the target torque) of F-S MTLDs.Fifteen new F-S MTLDs and their appropriate drivers from three different implant manufacturers (Astra Tech, Biohorizon and Dr Idhe), five for each type, were selected. The procedure of peak torque measurement was performed in ten sequences before and after aging. In each sequence, ten repetitions of peak torque values were registered for the aging procedure. To measure the output of each device, a Tohnichi torque gauge was used.Before aging, peak torque measurements of all the devices tested in this study falled within 10% of their preset target values. After aging, a significant difference was seen between raw error values of three groups of MTLDs (P
ISSN:2676-296X