Optimizing the detection, ablation, and ion extraction efficiency of a single-particle laser ablation mass spectrometer for application in environments with low aerosol particle concentrations

<p>The aim of this study is to show how a newly developed aerodynamic lens system (ALS), a delayed ion extraction (DIE), and better electric shielding improve the efficiency of the Aircraft-based Laser ABlation Aerosol MAss spectrometer (ALABAMA). These improvements are applicable to single-pa...

Full description

Bibliographic Details
Main Authors: H.-C. Clemen, J. Schneider, T. Klimach, F. Helleis, F. Köllner, A. Hünig, F. Rubach, S. Mertes, H. Wex, F. Stratmann, A. Welti, R. Kohl, F. Frank, S. Borrmann
Format: Article
Language:English
Published: Copernicus Publications 2020-11-01
Series:Atmospheric Measurement Techniques
Online Access:https://amt.copernicus.org/articles/13/5923/2020/amt-13-5923-2020.pdf
_version_ 1811295308249825280
author H.-C. Clemen
J. Schneider
T. Klimach
F. Helleis
F. Köllner
A. Hünig
A. Hünig
F. Rubach
F. Rubach
S. Mertes
H. Wex
F. Stratmann
A. Welti
A. Welti
R. Kohl
F. Frank
S. Borrmann
S. Borrmann
author_facet H.-C. Clemen
J. Schneider
T. Klimach
F. Helleis
F. Köllner
A. Hünig
A. Hünig
F. Rubach
F. Rubach
S. Mertes
H. Wex
F. Stratmann
A. Welti
A. Welti
R. Kohl
F. Frank
S. Borrmann
S. Borrmann
author_sort H.-C. Clemen
collection DOAJ
description <p>The aim of this study is to show how a newly developed aerodynamic lens system (ALS), a delayed ion extraction (DIE), and better electric shielding improve the efficiency of the Aircraft-based Laser ABlation Aerosol MAss spectrometer (ALABAMA). These improvements are applicable to single-particle laser ablation mass spectrometers in general. To characterize the modifications, extensive size-resolved measurements with spherical polystyrene latex particles (PSL; 150–6000&thinsp;<span class="inline-formula">nm</span>) and cubic sodium chloride particles (NaCl; 400–1700&thinsp;<span class="inline-formula">nm</span>) were performed. Measurements at a fixed ALS position show an improved detectable particle size range of the new ALS compared to the previously used Liu-type ALS, especially for supermicron particles. At a lens pressure of 2.4&thinsp;<span class="inline-formula">hPa</span>, the new ALS achieves a PSL particle size range from 230 to 3240&thinsp;<span class="inline-formula">nm</span> with 50&thinsp;<span class="inline-formula">%</span> detection efficiency and between 350 and 2000&thinsp;<span class="inline-formula">nm</span> with 95&thinsp;<span class="inline-formula">%</span> detection efficiency. The particle beam divergence was determined by measuring the detection efficiency at variable ALS positions along the laser cross sections and found to be minimal for PSL at about 800&thinsp;<span class="inline-formula">nm</span>. Compared to measurements by single-particle mass spectrometry (SPMS) instruments using Liu-type ALSs, the minimum particle beam divergence is shifted towards larger particle sizes. However, there are no disadvantages compared to the Liu-type lenses for particle sizes down to 200&thinsp;<span class="inline-formula">nm</span>. Improvements achieved by using the DIE and an additional electric shielding could be evaluated by size-resolved measurements of the hit rate, which is the ratio of laser pulses yielding a detectable amount of ions to the total number of emitted laser pulses. In particular, the hit rate for multiply charged particles smaller than 500&thinsp;<span class="inline-formula">nm</span> is significantly improved by preventing an undesired deflection of these particles in the ion extraction field. Moreover, it was found that by using the DIE the ion yield of the ablation, ionization, and ion extraction process could be increased, resulting in up to 7 times higher signal intensities of the cation spectra. The enhanced ion yield results in a larger effective width of the ablation laser beam, which in turn leads to a hit rate of almost 100&thinsp;<span class="inline-formula">%</span> for PSL particles in the size range from 350 to 2000&thinsp;<span class="inline-formula">nm</span>. Regarding cubic NaCl particles the modifications of the ALABAMA result in an up to 2 times increased detection efficiency and an up to 5 times increased hit rate. The need for such instrument modifications arises in particular for measurements of particles that are present in low number concentrations such as ice-nucleating particles (INPs) in general, but also aerosol particles at high altitudes or in pristine environments. Especially for these low particle number concentrations, improved efficiencies help to overcome the statistical limitations of single-particle mass spectrometer measurements. As an example, laboratory INP measurements carried out in this study show that<span id="page5924"/> the application of the DIE alone increases the number of INP mass spectra per time unit by a factor of 2 to 3 for the sampled substances. Overall, the combination of instrument modifications presented here resulted in an increased measurement efficiency of the ALABAMA for different particle types and particles shape as well as for highly charged particles.</p>
first_indexed 2024-04-13T05:30:24Z
format Article
id doaj.art-973deed0c9ac4649b43725dcd9c841ce
institution Directory Open Access Journal
issn 1867-1381
1867-8548
language English
last_indexed 2024-04-13T05:30:24Z
publishDate 2020-11-01
publisher Copernicus Publications
record_format Article
series Atmospheric Measurement Techniques
spelling doaj.art-973deed0c9ac4649b43725dcd9c841ce2022-12-22T03:00:27ZengCopernicus PublicationsAtmospheric Measurement Techniques1867-13811867-85482020-11-01135923595310.5194/amt-13-5923-2020Optimizing the detection, ablation, and ion extraction efficiency of a single-particle laser ablation mass spectrometer for application in environments with low aerosol particle concentrationsH.-C. Clemen0J. Schneider1T. Klimach2F. Helleis3F. Köllner4A. Hünig5A. Hünig6F. Rubach7F. Rubach8S. Mertes9H. Wex10F. Stratmann11A. Welti12A. Welti13R. Kohl14F. Frank15S. Borrmann16S. Borrmann17Max Planck Institute for Chemistry, Mainz, GermanyMax Planck Institute for Chemistry, Mainz, GermanyMax Planck Institute for Chemistry, Mainz, GermanyMax Planck Institute for Chemistry, Mainz, GermanyMax Planck Institute for Chemistry, Mainz, GermanyMax Planck Institute for Chemistry, Mainz, GermanyInstitute for Atmospheric Physics, Johannes Gutenberg University, Mainz, GermanyMax Planck Institute for Chemistry, Mainz, GermanyLeibniz Institute for Tropospheric Research, Leipzig, GermanyLeibniz Institute for Tropospheric Research, Leipzig, GermanyLeibniz Institute for Tropospheric Research, Leipzig, GermanyLeibniz Institute for Tropospheric Research, Leipzig, GermanyLeibniz Institute for Tropospheric Research, Leipzig, Germanynow at: Finnish Meteorological Institute, Helsinki, FinlandInstitute for Atmospheric and Environmental Sciences, Goethe University, Frankfurt am Main, GermanyInstitute for Atmospheric and Environmental Sciences, Goethe University, Frankfurt am Main, GermanyMax Planck Institute for Chemistry, Mainz, GermanyInstitute for Atmospheric Physics, Johannes Gutenberg University, Mainz, Germany<p>The aim of this study is to show how a newly developed aerodynamic lens system (ALS), a delayed ion extraction (DIE), and better electric shielding improve the efficiency of the Aircraft-based Laser ABlation Aerosol MAss spectrometer (ALABAMA). These improvements are applicable to single-particle laser ablation mass spectrometers in general. To characterize the modifications, extensive size-resolved measurements with spherical polystyrene latex particles (PSL; 150–6000&thinsp;<span class="inline-formula">nm</span>) and cubic sodium chloride particles (NaCl; 400–1700&thinsp;<span class="inline-formula">nm</span>) were performed. Measurements at a fixed ALS position show an improved detectable particle size range of the new ALS compared to the previously used Liu-type ALS, especially for supermicron particles. At a lens pressure of 2.4&thinsp;<span class="inline-formula">hPa</span>, the new ALS achieves a PSL particle size range from 230 to 3240&thinsp;<span class="inline-formula">nm</span> with 50&thinsp;<span class="inline-formula">%</span> detection efficiency and between 350 and 2000&thinsp;<span class="inline-formula">nm</span> with 95&thinsp;<span class="inline-formula">%</span> detection efficiency. The particle beam divergence was determined by measuring the detection efficiency at variable ALS positions along the laser cross sections and found to be minimal for PSL at about 800&thinsp;<span class="inline-formula">nm</span>. Compared to measurements by single-particle mass spectrometry (SPMS) instruments using Liu-type ALSs, the minimum particle beam divergence is shifted towards larger particle sizes. However, there are no disadvantages compared to the Liu-type lenses for particle sizes down to 200&thinsp;<span class="inline-formula">nm</span>. Improvements achieved by using the DIE and an additional electric shielding could be evaluated by size-resolved measurements of the hit rate, which is the ratio of laser pulses yielding a detectable amount of ions to the total number of emitted laser pulses. In particular, the hit rate for multiply charged particles smaller than 500&thinsp;<span class="inline-formula">nm</span> is significantly improved by preventing an undesired deflection of these particles in the ion extraction field. Moreover, it was found that by using the DIE the ion yield of the ablation, ionization, and ion extraction process could be increased, resulting in up to 7 times higher signal intensities of the cation spectra. The enhanced ion yield results in a larger effective width of the ablation laser beam, which in turn leads to a hit rate of almost 100&thinsp;<span class="inline-formula">%</span> for PSL particles in the size range from 350 to 2000&thinsp;<span class="inline-formula">nm</span>. Regarding cubic NaCl particles the modifications of the ALABAMA result in an up to 2 times increased detection efficiency and an up to 5 times increased hit rate. The need for such instrument modifications arises in particular for measurements of particles that are present in low number concentrations such as ice-nucleating particles (INPs) in general, but also aerosol particles at high altitudes or in pristine environments. Especially for these low particle number concentrations, improved efficiencies help to overcome the statistical limitations of single-particle mass spectrometer measurements. As an example, laboratory INP measurements carried out in this study show that<span id="page5924"/> the application of the DIE alone increases the number of INP mass spectra per time unit by a factor of 2 to 3 for the sampled substances. Overall, the combination of instrument modifications presented here resulted in an increased measurement efficiency of the ALABAMA for different particle types and particles shape as well as for highly charged particles.</p>https://amt.copernicus.org/articles/13/5923/2020/amt-13-5923-2020.pdf
spellingShingle H.-C. Clemen
J. Schneider
T. Klimach
F. Helleis
F. Köllner
A. Hünig
A. Hünig
F. Rubach
F. Rubach
S. Mertes
H. Wex
F. Stratmann
A. Welti
A. Welti
R. Kohl
F. Frank
S. Borrmann
S. Borrmann
Optimizing the detection, ablation, and ion extraction efficiency of a single-particle laser ablation mass spectrometer for application in environments with low aerosol particle concentrations
Atmospheric Measurement Techniques
title Optimizing the detection, ablation, and ion extraction efficiency of a single-particle laser ablation mass spectrometer for application in environments with low aerosol particle concentrations
title_full Optimizing the detection, ablation, and ion extraction efficiency of a single-particle laser ablation mass spectrometer for application in environments with low aerosol particle concentrations
title_fullStr Optimizing the detection, ablation, and ion extraction efficiency of a single-particle laser ablation mass spectrometer for application in environments with low aerosol particle concentrations
title_full_unstemmed Optimizing the detection, ablation, and ion extraction efficiency of a single-particle laser ablation mass spectrometer for application in environments with low aerosol particle concentrations
title_short Optimizing the detection, ablation, and ion extraction efficiency of a single-particle laser ablation mass spectrometer for application in environments with low aerosol particle concentrations
title_sort optimizing the detection ablation and ion extraction efficiency of a single particle laser ablation mass spectrometer for application in environments with low aerosol particle concentrations
url https://amt.copernicus.org/articles/13/5923/2020/amt-13-5923-2020.pdf
work_keys_str_mv AT hcclemen optimizingthedetectionablationandionextractionefficiencyofasingleparticlelaserablationmassspectrometerforapplicationinenvironmentswithlowaerosolparticleconcentrations
AT jschneider optimizingthedetectionablationandionextractionefficiencyofasingleparticlelaserablationmassspectrometerforapplicationinenvironmentswithlowaerosolparticleconcentrations
AT tklimach optimizingthedetectionablationandionextractionefficiencyofasingleparticlelaserablationmassspectrometerforapplicationinenvironmentswithlowaerosolparticleconcentrations
AT fhelleis optimizingthedetectionablationandionextractionefficiencyofasingleparticlelaserablationmassspectrometerforapplicationinenvironmentswithlowaerosolparticleconcentrations
AT fkollner optimizingthedetectionablationandionextractionefficiencyofasingleparticlelaserablationmassspectrometerforapplicationinenvironmentswithlowaerosolparticleconcentrations
AT ahunig optimizingthedetectionablationandionextractionefficiencyofasingleparticlelaserablationmassspectrometerforapplicationinenvironmentswithlowaerosolparticleconcentrations
AT ahunig optimizingthedetectionablationandionextractionefficiencyofasingleparticlelaserablationmassspectrometerforapplicationinenvironmentswithlowaerosolparticleconcentrations
AT frubach optimizingthedetectionablationandionextractionefficiencyofasingleparticlelaserablationmassspectrometerforapplicationinenvironmentswithlowaerosolparticleconcentrations
AT frubach optimizingthedetectionablationandionextractionefficiencyofasingleparticlelaserablationmassspectrometerforapplicationinenvironmentswithlowaerosolparticleconcentrations
AT smertes optimizingthedetectionablationandionextractionefficiencyofasingleparticlelaserablationmassspectrometerforapplicationinenvironmentswithlowaerosolparticleconcentrations
AT hwex optimizingthedetectionablationandionextractionefficiencyofasingleparticlelaserablationmassspectrometerforapplicationinenvironmentswithlowaerosolparticleconcentrations
AT fstratmann optimizingthedetectionablationandionextractionefficiencyofasingleparticlelaserablationmassspectrometerforapplicationinenvironmentswithlowaerosolparticleconcentrations
AT awelti optimizingthedetectionablationandionextractionefficiencyofasingleparticlelaserablationmassspectrometerforapplicationinenvironmentswithlowaerosolparticleconcentrations
AT awelti optimizingthedetectionablationandionextractionefficiencyofasingleparticlelaserablationmassspectrometerforapplicationinenvironmentswithlowaerosolparticleconcentrations
AT rkohl optimizingthedetectionablationandionextractionefficiencyofasingleparticlelaserablationmassspectrometerforapplicationinenvironmentswithlowaerosolparticleconcentrations
AT ffrank optimizingthedetectionablationandionextractionefficiencyofasingleparticlelaserablationmassspectrometerforapplicationinenvironmentswithlowaerosolparticleconcentrations
AT sborrmann optimizingthedetectionablationandionextractionefficiencyofasingleparticlelaserablationmassspectrometerforapplicationinenvironmentswithlowaerosolparticleconcentrations
AT sborrmann optimizingthedetectionablationandionextractionefficiencyofasingleparticlelaserablationmassspectrometerforapplicationinenvironmentswithlowaerosolparticleconcentrations