Accuracy Assessment of the GlucoMen<sup>®</sup> Day CGM System in Individuals with Type 1 Diabetes: A Pilot Study

The aim of this study was to evaluate the accuracy and usability of a novel continuous glucose monitoring (CGM) system designed for needle-free insertion and reduced environmental impact. We assessed the sensor performance of two GlucoMen<sup>®</sup> Day CGM systems worn simultaneously b...

Full description

Bibliographic Details
Main Authors: Daniel A. Hochfellner, Amra Simic, Marlene T. Taucher, Lea S. Sailer, Julia Kopanz, Tina Pöttler, Julia K. Mader
Format: Article
Language:English
Published: MDPI AG 2022-02-01
Series:Biosensors
Subjects:
Online Access:https://www.mdpi.com/2079-6374/12/2/106
Description
Summary:The aim of this study was to evaluate the accuracy and usability of a novel continuous glucose monitoring (CGM) system designed for needle-free insertion and reduced environmental impact. We assessed the sensor performance of two GlucoMen<sup>®</sup> Day CGM systems worn simultaneously by eight participants with type 1 diabetes. Self-monitoring of blood glucose (SMBG) was performed regularly over 14 days at home. Participants underwent two standardized, 5-h meal challenges at the research center with frequent plasma glucose (PG) measurements using a laboratory reference (YSI) instrument. When comparing CGM to PG, the overall mean absolute relative difference (MARD) was 9.7 [2.6–14.6]%. The overall MARD for CGM vs. SMBG was 13.1 [3.5–18.6]%. The consensus error grid (CEG) analysis showed 98% of both CGM/PG and CGM/SMBG pairs in the clinically acceptable zones A and B. The analysis confirmed that GlucoMen<sup>®</sup> Day CGM meets the clinical requirements for state-of-the-art CGM. In addition, the needle-free insertion technology is well tolerated by users and reduces medical waste compared to conventional CGM systems.
ISSN:2079-6374