Indoor calibration method for UV index meters with a solar simulator and a reference spectroradiometer
We present a calibration method of UV index meters under a solar simulator based on a spectroradiometer as reference. The reference spectroradiometer is a double-grating-based type that can measure the erythemal effective irradiance and the global UV index in a waveleng...
Main Authors: | , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
EDP Sciences
2016-01-01
|
Series: | International Journal of Metrology and Quality Engineering |
Subjects: | |
Online Access: | https://doi.org/10.1051/ijmqe/2015031 |
_version_ | 1818723112548564992 |
---|---|
author | Zaini Helmi Yoo Jae-Keun Park Seongchong Lee Dong-Hoon |
author_facet | Zaini Helmi Yoo Jae-Keun Park Seongchong Lee Dong-Hoon |
author_sort | Zaini Helmi |
collection | DOAJ |
description | We present a calibration method of UV index meters under a solar simulator based on a
spectroradiometer as reference. The reference spectroradiometer is a double-grating-based
type that can measure the erythemal effective irradiance and the global UV index in a
wavelength range from 250 nm to 450 nm. After calibration of its wavelength scale and
spectral irradiance scale by using the KRISS standard lamp, we applied the
spectroradiometer to measure the reference value for calibrating broadband UV index meters
under a solar simulator. The uncertainty of the spectroradiometer for measuring the
reference value for the global UV index is evaluated to be 5% as a relative expanded
uncertainty (k =
2). |
first_indexed | 2024-12-17T21:05:21Z |
format | Article |
id | doaj.art-9a00dc42ce9043ac9e946a4e4694a8dd |
institution | Directory Open Access Journal |
issn | 2107-6839 2107-6847 |
language | English |
last_indexed | 2024-12-17T21:05:21Z |
publishDate | 2016-01-01 |
publisher | EDP Sciences |
record_format | Article |
series | International Journal of Metrology and Quality Engineering |
spelling | doaj.art-9a00dc42ce9043ac9e946a4e4694a8dd2022-12-21T21:32:36ZengEDP SciencesInternational Journal of Metrology and Quality Engineering2107-68392107-68472016-01-017110110.1051/ijmqe/2015031ijmqe151037Indoor calibration method for UV index meters with a solar simulator and a reference spectroradiometerZaini HelmiYoo Jae-KeunPark Seongchong0Lee Dong-HoonKorea Research Institute of Standards and ScienceWe present a calibration method of UV index meters under a solar simulator based on a spectroradiometer as reference. The reference spectroradiometer is a double-grating-based type that can measure the erythemal effective irradiance and the global UV index in a wavelength range from 250 nm to 450 nm. After calibration of its wavelength scale and spectral irradiance scale by using the KRISS standard lamp, we applied the spectroradiometer to measure the reference value for calibrating broadband UV index meters under a solar simulator. The uncertainty of the spectroradiometer for measuring the reference value for the global UV index is evaluated to be 5% as a relative expanded uncertainty (k = 2).https://doi.org/10.1051/ijmqe/2015031Global UV indexspectroradiometererythemal effective irradiance |
spellingShingle | Zaini Helmi Yoo Jae-Keun Park Seongchong Lee Dong-Hoon Indoor calibration method for UV index meters with a solar simulator and a reference spectroradiometer International Journal of Metrology and Quality Engineering Global UV index spectroradiometer erythemal effective irradiance |
title | Indoor calibration method for UV index meters with a solar
simulator and a reference spectroradiometer |
title_full | Indoor calibration method for UV index meters with a solar
simulator and a reference spectroradiometer |
title_fullStr | Indoor calibration method for UV index meters with a solar
simulator and a reference spectroradiometer |
title_full_unstemmed | Indoor calibration method for UV index meters with a solar
simulator and a reference spectroradiometer |
title_short | Indoor calibration method for UV index meters with a solar
simulator and a reference spectroradiometer |
title_sort | indoor calibration method for uv index meters with a solar simulator and a reference spectroradiometer |
topic | Global UV index spectroradiometer erythemal effective irradiance |
url | https://doi.org/10.1051/ijmqe/2015031 |
work_keys_str_mv | AT zainihelmi indoorcalibrationmethodforuvindexmeterswithasolarsimulatorandareferencespectroradiometer AT yoojaekeun indoorcalibrationmethodforuvindexmeterswithasolarsimulatorandareferencespectroradiometer AT parkseongchong indoorcalibrationmethodforuvindexmeterswithasolarsimulatorandareferencespectroradiometer AT leedonghoon indoorcalibrationmethodforuvindexmeterswithasolarsimulatorandareferencespectroradiometer |