Optimization of High Uniformity Steady-State Magnetic Field Generator
[Introduction] To evaluate the anti magnetic interference ability of electrical and electronic devices, it is necessary to carry out the corresponding magnetic field immunity test, in which the high uniformity steady-state magnetic field generator is the core component of a steady-state test device....
Main Author: | |
---|---|
Format: | Article |
Language: | English |
Published: |
Energy Observer Magazine Co., Ltd.
2022-06-01
|
Series: | 南方能源建设 |
Subjects: | |
Online Access: | https://www.energychina.press/en/article/doi/10.16516/j.gedi.issn2095-8676.2022.02.011 |
_version_ | 1797857898013917184 |
---|---|
author | Yan HE |
author_facet | Yan HE |
author_sort | Yan HE |
collection | DOAJ |
description | [Introduction] To evaluate the anti magnetic interference ability of electrical and electronic devices, it is necessary to carry out the corresponding magnetic field immunity test, in which the high uniformity steady-state magnetic field generator is the core component of a steady-state test device. [Method] Taking the conventional Helmholtz and Maxwell square coil as the research object, the parameters were optimized through numerical calculation, and the coil structure that can produce high-performance magnetic field was obtained analytically. The finite element method was used to establish a model for calculation, and the correctness of the results was verified. [Result] The general calculation formula of coil parameters is obtained through numerical analysis, and the two coil types are compared. The optimized Maxwell square coil with small size and power consumption is selected as the final structure. [Conclusion] The research of this paper provides a detailed calculation process and design reference for the high uniformity steady-state magnetic field generator, and can provide a theoretical basis for the design of corresponding equipment in the electrical industry. |
first_indexed | 2024-04-09T21:04:18Z |
format | Article |
id | doaj.art-9ab12aa212514d1fb61738b90426b4f7 |
institution | Directory Open Access Journal |
issn | 2095-8676 |
language | English |
last_indexed | 2024-04-09T21:04:18Z |
publishDate | 2022-06-01 |
publisher | Energy Observer Magazine Co., Ltd. |
record_format | Article |
series | 南方能源建设 |
spelling | doaj.art-9ab12aa212514d1fb61738b90426b4f72023-03-29T02:02:03ZengEnergy Observer Magazine Co., Ltd.南方能源建设2095-86762022-06-0192828910.16516/j.gedi.issn2095-8676.2022.02.0112022-070Optimization of High Uniformity Steady-State Magnetic Field GeneratorYan HE0Hefei Xinhu Canned Motor Pump Co., Ltd., Hefei 230088, Anhui, China[Introduction] To evaluate the anti magnetic interference ability of electrical and electronic devices, it is necessary to carry out the corresponding magnetic field immunity test, in which the high uniformity steady-state magnetic field generator is the core component of a steady-state test device. [Method] Taking the conventional Helmholtz and Maxwell square coil as the research object, the parameters were optimized through numerical calculation, and the coil structure that can produce high-performance magnetic field was obtained analytically. The finite element method was used to establish a model for calculation, and the correctness of the results was verified. [Result] The general calculation formula of coil parameters is obtained through numerical analysis, and the two coil types are compared. The optimized Maxwell square coil with small size and power consumption is selected as the final structure. [Conclusion] The research of this paper provides a detailed calculation process and design reference for the high uniformity steady-state magnetic field generator, and can provide a theoretical basis for the design of corresponding equipment in the electrical industry.https://www.energychina.press/en/article/doi/10.16516/j.gedi.issn2095-8676.2022.02.011immunity testinduction coilsolenoidmagnetic field measurementfinite element methodmagnetic field uniformity |
spellingShingle | Yan HE Optimization of High Uniformity Steady-State Magnetic Field Generator 南方能源建设 immunity test induction coil solenoid magnetic field measurement finite element method magnetic field uniformity |
title | Optimization of High Uniformity Steady-State Magnetic Field Generator |
title_full | Optimization of High Uniformity Steady-State Magnetic Field Generator |
title_fullStr | Optimization of High Uniformity Steady-State Magnetic Field Generator |
title_full_unstemmed | Optimization of High Uniformity Steady-State Magnetic Field Generator |
title_short | Optimization of High Uniformity Steady-State Magnetic Field Generator |
title_sort | optimization of high uniformity steady state magnetic field generator |
topic | immunity test induction coil solenoid magnetic field measurement finite element method magnetic field uniformity |
url | https://www.energychina.press/en/article/doi/10.16516/j.gedi.issn2095-8676.2022.02.011 |
work_keys_str_mv | AT yanhe optimizationofhighuniformitysteadystatemagneticfieldgenerator |