Comparison between external locking plate fixation and conventional external fixation for extraarticular proximal tibial fractures: a finite element analysis

Abstract Background Good clinical outcomes for locking plates as an external fixator to treat tibial fractures have been reported. However, external locking plate fixation is still generally rarely performed. This study aimed to compare the stability of an external locking plate fixator with that of...

全面介绍

书目详细资料
Main Authors: Dejan Blažević, Janoš Kodvanj, Petra Adamović, Dinko Vidović, Zlatko Trobonjača, Srećko Sabalić
格式: 文件
语言:English
出版: BMC 2022-01-01
丛编:Journal of Orthopaedic Surgery and Research
主题:
在线阅读:https://doi.org/10.1186/s13018-021-02907-3
_version_ 1828152863394627584
author Dejan Blažević
Janoš Kodvanj
Petra Adamović
Dinko Vidović
Zlatko Trobonjača
Srećko Sabalić
author_facet Dejan Blažević
Janoš Kodvanj
Petra Adamović
Dinko Vidović
Zlatko Trobonjača
Srećko Sabalić
author_sort Dejan Blažević
collection DOAJ
description Abstract Background Good clinical outcomes for locking plates as an external fixator to treat tibial fractures have been reported. However, external locking plate fixation is still generally rarely performed. This study aimed to compare the stability of an external locking plate fixator with that of a conventional external fixator for extraarticular proximal tibial fractures using finite element analysis. Methods Three models were constructed: (1) external locking plate fixation of proximal tibial fracture with lateral proximal tibial locking plate and 5-mm screws (ELP), (2) conventional external fixation of proximal tibial fracture with an 11-mm rod and 5-mm Schanz screws (EF-11), and (3) conventional external fixation of a proximal tibial fracture with a 7-mm rod and 5-mm Schanz screws (EF-7). The stress distribution, displacement at the fracture gap, and stiffness of the three finite element models at 30-, 40-, 50-, and 60-mm plate–rod offsets from the lateral surface of the lateral condyle of the tibia were determined. Results The conventional external fixator showed higher stiffness than the external locking plate fixator. In all models, the stiffness decreased as the distance of the plate–rod from the bone surface increased. The maximum stiffness was 121.06 N/mm in the EF-11 model with 30-mm tibia–rod offset. In the EF-7 model group, the maximum stiffness was 40.00 N/mm in the model with 30-mm tibia–rod offset. In the ELP model group, the maximum stiffness was 35.79 N/mm in the model with 30-mm tibia–plate offset. Conclusions Finite element analysis indicated that external locking plate fixation is more flexible than conventional external fixation and can influence secondary bone healing. External locking plate fixation requires the placement of the plate as close as possible to the skin, which allows for a low-profile design because the increased distance from the plate to the bone can be too flexible for bone healing. Further experimental mechanical model tests are necessary to validate these finite element models, and further biological analysis is necessary to evaluate the effect of external locking plate fixation on fracture healing.
first_indexed 2024-04-11T22:17:32Z
format Article
id doaj.art-9d34d07fb8014df7b20b3bff5349c3a4
institution Directory Open Access Journal
issn 1749-799X
language English
last_indexed 2024-04-11T22:17:32Z
publishDate 2022-01-01
publisher BMC
record_format Article
series Journal of Orthopaedic Surgery and Research
spelling doaj.art-9d34d07fb8014df7b20b3bff5349c3a42022-12-22T04:00:19ZengBMCJournal of Orthopaedic Surgery and Research1749-799X2022-01-011711910.1186/s13018-021-02907-3Comparison between external locking plate fixation and conventional external fixation for extraarticular proximal tibial fractures: a finite element analysisDejan Blažević0Janoš Kodvanj1Petra Adamović2Dinko Vidović3Zlatko Trobonjača4Srećko Sabalić5Department of Traumatology, Sestre milosrdnice University Hospital CenterFaculty of Mechanical Engineering and Naval Architecture, University of ZagrebFaculty of Mechanical Engineering and Naval Architecture, University of ZagrebDepartment of Traumatology, Sestre milosrdnice University Hospital CenterSchool of Medicine, University of RijekaDepartment of Traumatology, Sestre milosrdnice University Hospital CenterAbstract Background Good clinical outcomes for locking plates as an external fixator to treat tibial fractures have been reported. However, external locking plate fixation is still generally rarely performed. This study aimed to compare the stability of an external locking plate fixator with that of a conventional external fixator for extraarticular proximal tibial fractures using finite element analysis. Methods Three models were constructed: (1) external locking plate fixation of proximal tibial fracture with lateral proximal tibial locking plate and 5-mm screws (ELP), (2) conventional external fixation of proximal tibial fracture with an 11-mm rod and 5-mm Schanz screws (EF-11), and (3) conventional external fixation of a proximal tibial fracture with a 7-mm rod and 5-mm Schanz screws (EF-7). The stress distribution, displacement at the fracture gap, and stiffness of the three finite element models at 30-, 40-, 50-, and 60-mm plate–rod offsets from the lateral surface of the lateral condyle of the tibia were determined. Results The conventional external fixator showed higher stiffness than the external locking plate fixator. In all models, the stiffness decreased as the distance of the plate–rod from the bone surface increased. The maximum stiffness was 121.06 N/mm in the EF-11 model with 30-mm tibia–rod offset. In the EF-7 model group, the maximum stiffness was 40.00 N/mm in the model with 30-mm tibia–rod offset. In the ELP model group, the maximum stiffness was 35.79 N/mm in the model with 30-mm tibia–plate offset. Conclusions Finite element analysis indicated that external locking plate fixation is more flexible than conventional external fixation and can influence secondary bone healing. External locking plate fixation requires the placement of the plate as close as possible to the skin, which allows for a low-profile design because the increased distance from the plate to the bone can be too flexible for bone healing. Further experimental mechanical model tests are necessary to validate these finite element models, and further biological analysis is necessary to evaluate the effect of external locking plate fixation on fracture healing.https://doi.org/10.1186/s13018-021-02907-3Finite element analysisLocking plateExternal fixatorTibiaFracture
spellingShingle Dejan Blažević
Janoš Kodvanj
Petra Adamović
Dinko Vidović
Zlatko Trobonjača
Srećko Sabalić
Comparison between external locking plate fixation and conventional external fixation for extraarticular proximal tibial fractures: a finite element analysis
Journal of Orthopaedic Surgery and Research
Finite element analysis
Locking plate
External fixator
Tibia
Fracture
title Comparison between external locking plate fixation and conventional external fixation for extraarticular proximal tibial fractures: a finite element analysis
title_full Comparison between external locking plate fixation and conventional external fixation for extraarticular proximal tibial fractures: a finite element analysis
title_fullStr Comparison between external locking plate fixation and conventional external fixation for extraarticular proximal tibial fractures: a finite element analysis
title_full_unstemmed Comparison between external locking plate fixation and conventional external fixation for extraarticular proximal tibial fractures: a finite element analysis
title_short Comparison between external locking plate fixation and conventional external fixation for extraarticular proximal tibial fractures: a finite element analysis
title_sort comparison between external locking plate fixation and conventional external fixation for extraarticular proximal tibial fractures a finite element analysis
topic Finite element analysis
Locking plate
External fixator
Tibia
Fracture
url https://doi.org/10.1186/s13018-021-02907-3
work_keys_str_mv AT dejanblazevic comparisonbetweenexternallockingplatefixationandconventionalexternalfixationforextraarticularproximaltibialfracturesafiniteelementanalysis
AT janoskodvanj comparisonbetweenexternallockingplatefixationandconventionalexternalfixationforextraarticularproximaltibialfracturesafiniteelementanalysis
AT petraadamovic comparisonbetweenexternallockingplatefixationandconventionalexternalfixationforextraarticularproximaltibialfracturesafiniteelementanalysis
AT dinkovidovic comparisonbetweenexternallockingplatefixationandconventionalexternalfixationforextraarticularproximaltibialfracturesafiniteelementanalysis
AT zlatkotrobonjaca comparisonbetweenexternallockingplatefixationandconventionalexternalfixationforextraarticularproximaltibialfracturesafiniteelementanalysis
AT sreckosabalic comparisonbetweenexternallockingplatefixationandconventionalexternalfixationforextraarticularproximaltibialfracturesafiniteelementanalysis