Computer 3D modeling of radiofrequency ablation of atypical cartilaginous tumours in long bones using finite element methods and real patient anatomy

Abstract Background Radiofrequency ablation (RFA) is a minimally invasive technique used for the treatment of neoplasms, with a growing interest in the treatment of bone tumours. However, the lack of data concerning the size of the resulting ablation zones in RFA of bone tumours makes prospective pl...

Full description

Bibliographic Details
Main Authors: Ricardo Rivas Loya, Paul C. Jutte, Thomas C. Kwee, Peter M. A. van Ooijen
Format: Article
Language:English
Published: SpringerOpen 2022-04-01
Series:European Radiology Experimental
Subjects:
Online Access:https://doi.org/10.1186/s41747-022-00271-3
Description
Summary:Abstract Background Radiofrequency ablation (RFA) is a minimally invasive technique used for the treatment of neoplasms, with a growing interest in the treatment of bone tumours. However, the lack of data concerning the size of the resulting ablation zones in RFA of bone tumours makes prospective planning challenging, needed for safe and effective treatment. Methods Using retrospective computed tomography and magnetic resonance imaging data from patients treated with RFA of atypical cartilaginous tumours (ACTs), the bone, tumours, and final position of the RFA electrode were segmented from the medical images and used in finite element models to simulate RFA. Tissue parameters were optimised, and boundary conditions were defined to mimic the clinical scenario. The resulting ablation diameters from postoperative images were then measured and compared to the ones from the simulations, and the error between them was calculated. Results Seven cases had all the information required to create the finite element models. The resulting median error (in all three directions) was -1 mm, with interquartile ranges from -3 to 3 mm. The three-dimensional models showed that the thermal damage concentrates close to the cortical wall in the first minutes and then becomes more evenly distributed. Conclusions Computer simulations can predict the ablation diameters with acceptable accuracy and may thus be utilised for patient planning. This could allow interventional radiologists to accurately define the time, electrode length, and position required to treat ACTs with RFA and make adjustments as needed to guarantee total tumour destruction while sparing as much healthy tissue as possible.
ISSN:2509-9280