Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients.

<h4>Purpose</h4>To evaluate agreement of total corneal power (TCP) measured by swept-source anterior segment optical coherence tomography (CASIA 2), Scheimpflug tomography (Pentacam AXL), and swept-source optical biometer (IOLMaster 700) in normal and keratoconic patients.<h4>Metho...

Full description

Bibliographic Details
Main Authors: Rosepon Asawaworarit, Vannarut Satitpitakul, Parichart Taweekitikul, Krit Pongpirul
Format: Article
Language:English
Published: Public Library of Science (PLoS) 2022-01-01
Series:PLoS ONE
Online Access:https://doi.org/10.1371/journal.pone.0268856
_version_ 1818048568132370432
author Rosepon Asawaworarit
Vannarut Satitpitakul
Parichart Taweekitikul
Krit Pongpirul
author_facet Rosepon Asawaworarit
Vannarut Satitpitakul
Parichart Taweekitikul
Krit Pongpirul
author_sort Rosepon Asawaworarit
collection DOAJ
description <h4>Purpose</h4>To evaluate agreement of total corneal power (TCP) measured by swept-source anterior segment optical coherence tomography (CASIA 2), Scheimpflug tomography (Pentacam AXL), and swept-source optical biometer (IOLMaster 700) in normal and keratoconic patients.<h4>Methods</h4>This is a prospective observational study conducted at King Chulalongkorn Memorial Hospital, Bangkok, Thailand. Biometric values were measured by each device for three times by two operators to evaluate repeatability and reproducibility of TCP. The agreement of TCP and other parameters including total corneal astigmatism, anterior keratometry, anterior corneal astigmatism, posterior keratometry, posterior corneal astigmatism, anterior chamber depth, white-to-white corneal diameter (WTW), central corneal thickness, and intraocular power were also evaluated.<h4>Results</h4>This study enrolled 100 healthy participants and 34 patients with keratoconus. The repeatability and reproducibility of TCP were high in all devices (ICC > 0.9). The agreement of TCP was excellent among three devices in both groups (ICC > 0.9). However, the agreement of TCP between CASIA 2 and IOLMaster 700 was slightly lower in healthy participants (ICC = 0.867) and keratoconic patients (ICC = 0.852) with mean differences of more than 1.0 diopter is clinically significant. Wider 95% limit of agreement were found in keratoconic eyes. Most of other parameters showed good to excellent agreement except WTW which showed poor to moderate agreement in both groups. IOL power showed clinically significant differences in patients with keratoconus.<h4>Conclusions</h4>The agreement of TCP measured by three devices was excellent in normal and keratoconic patients. However, TCP cannot be used interchangeably between devices.
first_indexed 2024-12-10T10:23:45Z
format Article
id doaj.art-a31a15db3c3e4fc2bb9d591de529f5f5
institution Directory Open Access Journal
issn 1932-6203
language English
last_indexed 2024-12-10T10:23:45Z
publishDate 2022-01-01
publisher Public Library of Science (PLoS)
record_format Article
series PLoS ONE
spelling doaj.art-a31a15db3c3e4fc2bb9d591de529f5f52022-12-22T01:52:48ZengPublic Library of Science (PLoS)PLoS ONE1932-62032022-01-01175e026885610.1371/journal.pone.0268856Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients.Rosepon AsawaworaritVannarut SatitpitakulParichart TaweekitikulKrit Pongpirul<h4>Purpose</h4>To evaluate agreement of total corneal power (TCP) measured by swept-source anterior segment optical coherence tomography (CASIA 2), Scheimpflug tomography (Pentacam AXL), and swept-source optical biometer (IOLMaster 700) in normal and keratoconic patients.<h4>Methods</h4>This is a prospective observational study conducted at King Chulalongkorn Memorial Hospital, Bangkok, Thailand. Biometric values were measured by each device for three times by two operators to evaluate repeatability and reproducibility of TCP. The agreement of TCP and other parameters including total corneal astigmatism, anterior keratometry, anterior corneal astigmatism, posterior keratometry, posterior corneal astigmatism, anterior chamber depth, white-to-white corneal diameter (WTW), central corneal thickness, and intraocular power were also evaluated.<h4>Results</h4>This study enrolled 100 healthy participants and 34 patients with keratoconus. The repeatability and reproducibility of TCP were high in all devices (ICC > 0.9). The agreement of TCP was excellent among three devices in both groups (ICC > 0.9). However, the agreement of TCP between CASIA 2 and IOLMaster 700 was slightly lower in healthy participants (ICC = 0.867) and keratoconic patients (ICC = 0.852) with mean differences of more than 1.0 diopter is clinically significant. Wider 95% limit of agreement were found in keratoconic eyes. Most of other parameters showed good to excellent agreement except WTW which showed poor to moderate agreement in both groups. IOL power showed clinically significant differences in patients with keratoconus.<h4>Conclusions</h4>The agreement of TCP measured by three devices was excellent in normal and keratoconic patients. However, TCP cannot be used interchangeably between devices.https://doi.org/10.1371/journal.pone.0268856
spellingShingle Rosepon Asawaworarit
Vannarut Satitpitakul
Parichart Taweekitikul
Krit Pongpirul
Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients.
PLoS ONE
title Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients.
title_full Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients.
title_fullStr Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients.
title_full_unstemmed Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients.
title_short Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients.
title_sort agreement of total corneal power between 2 swept source optical coherence tomography and scheimpflug tomography in normal and keratoconic patients
url https://doi.org/10.1371/journal.pone.0268856
work_keys_str_mv AT roseponasawaworarit agreementoftotalcornealpowerbetween2sweptsourceopticalcoherencetomographyandscheimpflugtomographyinnormalandkeratoconicpatients
AT vannarutsatitpitakul agreementoftotalcornealpowerbetween2sweptsourceopticalcoherencetomographyandscheimpflugtomographyinnormalandkeratoconicpatients
AT paricharttaweekitikul agreementoftotalcornealpowerbetween2sweptsourceopticalcoherencetomographyandscheimpflugtomographyinnormalandkeratoconicpatients
AT kritpongpirul agreementoftotalcornealpowerbetween2sweptsourceopticalcoherencetomographyandscheimpflugtomographyinnormalandkeratoconicpatients