False positive results of Bowie and Dick type test used for hospital steam sterilizer with slower come-up ramps: A case study.
OBJECTIVE:To determine if the standardized B&D type test for hospital steam sterilizer is correctly indicating cycle failures in slower come-up ramps cycles. METHODS:Two commercially available B&D type test were challenged in a quality control sterilizer. A common failure was simulated in tr...
Main Authors: | Paulo Roberto Laranjeira, Jeane Aparecida Gonzalez Bronzatti, Camila Quartim de Moraes Bruna, Rafael Queiroz de Souza, Kazuko Uchikawa Graziano, Viktoriya Lusignan |
---|---|
Format: | Article |
Language: | English |
Published: |
Public Library of Science (PLoS)
2020-01-01
|
Series: | PLoS ONE |
Online Access: | https://doi.org/10.1371/journal.pone.0227943 |
Similar Items
-
Pacotes molhados: o aumento do tempo de secagem aumenta o consumo de água (recurso natural escasso)?
by: Paulo Roberto Laranjeira, et al.
Published: (2019-02-01) -
Microbiological evaluation of the steam sterilization of assembled laparoscopic instruments
by: Tamara Carolina de Camargo, et al. -
Sterility Maintenance Assessment of Moist/Wet Material After Steam Sterilization and 30-day Storage Evaluación de la manutención de la esterilidad de materiales húmedos/mojados después de esterilizados con vapor y almacenados por 30 días Avaliação da manutenção da esterilidade de materiais úmidos/molhados após a esterilização por vapor e armazenamento por 30 dias
by: Giovana Abrahão de Araújo Moriya, et al.
Published: (2010-08-01) -
Sterility maintenance assessment of moist/wet material after steam sterilization and 30-day storage
by: Moriya GADA, et al.
Published: (2011-06-01) -
Confronting Inauthenticity – Utopias of David Bowie
by: Pavao Parunov
Published: (2015-06-01)