International Multicenter, Myopic and Myopic Astigmatism Femto LASIK, Customized by Automated Ray-Tracing Ablation Profile Calculation: A Postmarket Study
Anastasios John Kanellopoulos,1,2,* Matthias Maus,3,* Chandra Bala,4,* Cody Hamilton,5,* Sissimos Lemonis,6,* Maria Elena Jockovich,5,* Ramin Khoramnia7,* 1LaserVision Ambulatory Surgical Unit, Athens, Greece; 2Department of Ophthalmology, NYU...
Main Authors: | , , , , , , |
---|---|
格式: | 文件 |
语言: | English |
出版: |
Dove Medical Press
2024-02-01
|
丛编: | Clinical Ophthalmology |
主题: | |
在线阅读: | https://www.dovepress.com/international-multicenter-myopic-and-myopic-astigmatism-femto-lasik-cu-peer-reviewed-fulltext-article-OPTH |
_version_ | 1827346419103563776 |
---|---|
author | Kanellopoulos AJ Maus M Bala C Hamilton C Lemonis S Jockovich ME Khoramnia R |
author_facet | Kanellopoulos AJ Maus M Bala C Hamilton C Lemonis S Jockovich ME Khoramnia R |
author_sort | Kanellopoulos AJ |
collection | DOAJ |
description | Anastasios John Kanellopoulos,1,2,* Matthias Maus,3,* Chandra Bala,4,* Cody Hamilton,5,* Sissimos Lemonis,6,* Maria Elena Jockovich,5,* Ramin Khoramnia7,* 1LaserVision Ambulatory Surgical Unit, Athens, Greece; 2Department of Ophthalmology, NYU Grossman School of Medicine, New York City, NY, USA; 3Sehkraft Augenzentrum, Köln, Germany; 4PersonalEyes, Sydney, NSW, Australia; 5Alcon Vision, LLC, Fort Worth, TX, USA; 6Alcon Deutschland GmbH, Freiburg, Germany; 7Department of Ophthalmology, University of Heidelberg, Heidelberg, Germany*These authors contributed equally to this workCorrespondence: Anastasios John Kanellopoulos, LaserVision Ambulatory Surgical Unit, An Thocha 17, Athens, 115 21, Greece, Tel +30 (210) 7472777, Fax + 30 (210) 7472789, Email ajkmd@mac.comPurpose: To assess the safety and efficacy of a customized ablation treatment (InnovEyes) to correct myopia and myopic astigmatism with femtosecond laser-assisted in situ keratomileusis (Femto LASIK).Patients and Methods: In this prospective, nonrandomized, multicenter study, 113 patients (225 eyes) with preoperative myopia less than − 9.0 diopters (D) and astigmatism 0 to − 4.0 D (based on InnovEyes refraction) underwent wavefront, tomography, and biometry assessment using a single diagnostic device (InnovEyes sightmap). These data were imported and used unmodified by the InnovEyes algorithm to automatically calculate and optimize correction of lower- and higher‑order aberrations (HOAs) treated by the EX500 ablation profile. Visual acuity, refractive error, HOAs, and patient satisfaction were evaluated over 3 months.Results: A total of 106 patients (212 eyes) completed the study and were included in the analysis. Mean preoperative manifest refraction spherical equivalent (MRSE) was − 3.38± 1.76 D. At Month 3, uncorrected distance visual acuity was 20/20 or better in 208/212 (98.1%) eyes, and it was the same as, or better than, the preoperative best-corrected distance visual acuity (CDVA) in 162/212 (76.4%) eyes; 76/212 (35.8%) eyes gained ≥ 1 line of CDVA. MRSE was within ± 0.5 D in 195/212 (92.0%) eyes. Additionally, 201/209 (96.2%) eyes had no change (defined as a change between − 0.1 μm and 0.1 μm, inclusive) in HOAs, and 105/106 (99.1%) patients reported to be satisfied with the results.Conclusion: Customizing ray-tracing Femto LASIK with this platform appeared safe and effective in correcting myopic astigmatism and also achieved a significant percentage of eyes gaining lines of vision, potentially by addressing HOAs, along with a consistently high level of patient satisfaction.Keywords: myopic astigmatism, higher-order aberrations, manifest refraction spherical equivalent, corrected distance visual acuity |
first_indexed | 2024-03-07T23:29:53Z |
format | Article |
id | doaj.art-a8fc239e9d8d4f43a09da85bc3172786 |
institution | Directory Open Access Journal |
issn | 1177-5483 |
language | English |
last_indexed | 2024-03-07T23:29:53Z |
publishDate | 2024-02-01 |
publisher | Dove Medical Press |
record_format | Article |
series | Clinical Ophthalmology |
spelling | doaj.art-a8fc239e9d8d4f43a09da85bc31727862024-02-20T17:10:12ZengDove Medical PressClinical Ophthalmology1177-54832024-02-01Volume 1852553690581International Multicenter, Myopic and Myopic Astigmatism Femto LASIK, Customized by Automated Ray-Tracing Ablation Profile Calculation: A Postmarket StudyKanellopoulos AJMaus MBala CHamilton CLemonis SJockovich MEKhoramnia RAnastasios John Kanellopoulos,1,2,* Matthias Maus,3,* Chandra Bala,4,* Cody Hamilton,5,* Sissimos Lemonis,6,* Maria Elena Jockovich,5,* Ramin Khoramnia7,* 1LaserVision Ambulatory Surgical Unit, Athens, Greece; 2Department of Ophthalmology, NYU Grossman School of Medicine, New York City, NY, USA; 3Sehkraft Augenzentrum, Köln, Germany; 4PersonalEyes, Sydney, NSW, Australia; 5Alcon Vision, LLC, Fort Worth, TX, USA; 6Alcon Deutschland GmbH, Freiburg, Germany; 7Department of Ophthalmology, University of Heidelberg, Heidelberg, Germany*These authors contributed equally to this workCorrespondence: Anastasios John Kanellopoulos, LaserVision Ambulatory Surgical Unit, An Thocha 17, Athens, 115 21, Greece, Tel +30 (210) 7472777, Fax + 30 (210) 7472789, Email ajkmd@mac.comPurpose: To assess the safety and efficacy of a customized ablation treatment (InnovEyes) to correct myopia and myopic astigmatism with femtosecond laser-assisted in situ keratomileusis (Femto LASIK).Patients and Methods: In this prospective, nonrandomized, multicenter study, 113 patients (225 eyes) with preoperative myopia less than − 9.0 diopters (D) and astigmatism 0 to − 4.0 D (based on InnovEyes refraction) underwent wavefront, tomography, and biometry assessment using a single diagnostic device (InnovEyes sightmap). These data were imported and used unmodified by the InnovEyes algorithm to automatically calculate and optimize correction of lower- and higher‑order aberrations (HOAs) treated by the EX500 ablation profile. Visual acuity, refractive error, HOAs, and patient satisfaction were evaluated over 3 months.Results: A total of 106 patients (212 eyes) completed the study and were included in the analysis. Mean preoperative manifest refraction spherical equivalent (MRSE) was − 3.38± 1.76 D. At Month 3, uncorrected distance visual acuity was 20/20 or better in 208/212 (98.1%) eyes, and it was the same as, or better than, the preoperative best-corrected distance visual acuity (CDVA) in 162/212 (76.4%) eyes; 76/212 (35.8%) eyes gained ≥ 1 line of CDVA. MRSE was within ± 0.5 D in 195/212 (92.0%) eyes. Additionally, 201/209 (96.2%) eyes had no change (defined as a change between − 0.1 μm and 0.1 μm, inclusive) in HOAs, and 105/106 (99.1%) patients reported to be satisfied with the results.Conclusion: Customizing ray-tracing Femto LASIK with this platform appeared safe and effective in correcting myopic astigmatism and also achieved a significant percentage of eyes gaining lines of vision, potentially by addressing HOAs, along with a consistently high level of patient satisfaction.Keywords: myopic astigmatism, higher-order aberrations, manifest refraction spherical equivalent, corrected distance visual acuityhttps://www.dovepress.com/international-multicenter-myopic-and-myopic-astigmatism-femto-lasik-cu-peer-reviewed-fulltext-article-OPTHmyopic astigmatismhigher-order aberrationsmanifest refraction spherical equivalentcorrected distance visual acuity. |
spellingShingle | Kanellopoulos AJ Maus M Bala C Hamilton C Lemonis S Jockovich ME Khoramnia R International Multicenter, Myopic and Myopic Astigmatism Femto LASIK, Customized by Automated Ray-Tracing Ablation Profile Calculation: A Postmarket Study Clinical Ophthalmology myopic astigmatism higher-order aberrations manifest refraction spherical equivalent corrected distance visual acuity. |
title | International Multicenter, Myopic and Myopic Astigmatism Femto LASIK, Customized by Automated Ray-Tracing Ablation Profile Calculation: A Postmarket Study |
title_full | International Multicenter, Myopic and Myopic Astigmatism Femto LASIK, Customized by Automated Ray-Tracing Ablation Profile Calculation: A Postmarket Study |
title_fullStr | International Multicenter, Myopic and Myopic Astigmatism Femto LASIK, Customized by Automated Ray-Tracing Ablation Profile Calculation: A Postmarket Study |
title_full_unstemmed | International Multicenter, Myopic and Myopic Astigmatism Femto LASIK, Customized by Automated Ray-Tracing Ablation Profile Calculation: A Postmarket Study |
title_short | International Multicenter, Myopic and Myopic Astigmatism Femto LASIK, Customized by Automated Ray-Tracing Ablation Profile Calculation: A Postmarket Study |
title_sort | international multicenter myopic and myopic astigmatism femto lasik customized by automated ray tracing ablation profile calculation a postmarket study |
topic | myopic astigmatism higher-order aberrations manifest refraction spherical equivalent corrected distance visual acuity. |
url | https://www.dovepress.com/international-multicenter-myopic-and-myopic-astigmatism-femto-lasik-cu-peer-reviewed-fulltext-article-OPTH |
work_keys_str_mv | AT kanellopoulosaj internationalmulticentermyopicandmyopicastigmatismfemtolasikcustomizedbyautomatedraytracingablationprofilecalculationapostmarketstudy AT mausm internationalmulticentermyopicandmyopicastigmatismfemtolasikcustomizedbyautomatedraytracingablationprofilecalculationapostmarketstudy AT balac internationalmulticentermyopicandmyopicastigmatismfemtolasikcustomizedbyautomatedraytracingablationprofilecalculationapostmarketstudy AT hamiltonc internationalmulticentermyopicandmyopicastigmatismfemtolasikcustomizedbyautomatedraytracingablationprofilecalculationapostmarketstudy AT lemoniss internationalmulticentermyopicandmyopicastigmatismfemtolasikcustomizedbyautomatedraytracingablationprofilecalculationapostmarketstudy AT jockovichme internationalmulticentermyopicandmyopicastigmatismfemtolasikcustomizedbyautomatedraytracingablationprofilecalculationapostmarketstudy AT khoramniar internationalmulticentermyopicandmyopicastigmatismfemtolasikcustomizedbyautomatedraytracingablationprofilecalculationapostmarketstudy |