Introduction of a new parameter for evaluation of digital radiography system performance
Background: The aim of this study was to compare the image quality and radiation doses in various digital radiography systems using contrast-detail radiography (CDRAD) phantom. Methods: The image quality and radiation dose for seven different digital radiography systems were compared using the CDRAD...
Main Authors: | Mohammad Reza Choopani, Ali Chaparian |
---|---|
Format: | Article |
Language: | English |
Published: |
Wolters Kluwer Medknow Publications
2020-01-01
|
Series: | Journal of Medical Signals and Sensors |
Subjects: | |
Online Access: | http://www.jmssjournal.net/article.asp?issn=2228-7477;year=2020;volume=10;issue=3;spage=196;epage=200;aulast=Choopani |