Individualized parameterization of multiparametric monitors alarms in infarcted patients

ABSTRACT Objective: To measure the magnitude of the effect of an individualized parameterization protocol for hemodynamic alarms in patients with acute myocardial infarction. Method: Pragmatic clinical trial, open label and single arm, whose intervention was performed through a protocol validated...

Full description

Bibliographic Details
Main Authors: Allan Peixoto de Assis, Francimar Tinoco de Oliveira, Flávia Giron Camerini, Roberto Carlos Lyra da Silva, Camila Mendonça de Moraes
Format: Article
Language:English
Published: Associação Brasileira de Enfermagem
Series:Revista Brasileira de Enfermagem
Subjects:
Online Access:http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0034-71672019000300609&lng=en&tlng=en
_version_ 1819032065640759296
author Allan Peixoto de Assis
Francimar Tinoco de Oliveira
Flávia Giron Camerini
Roberto Carlos Lyra da Silva
Camila Mendonça de Moraes
author_facet Allan Peixoto de Assis
Francimar Tinoco de Oliveira
Flávia Giron Camerini
Roberto Carlos Lyra da Silva
Camila Mendonça de Moraes
author_sort Allan Peixoto de Assis
collection DOAJ
description ABSTRACT Objective: To measure the magnitude of the effect of an individualized parameterization protocol for hemodynamic alarms in patients with acute myocardial infarction. Method: Pragmatic clinical trial, open label and single arm, whose intervention was performed through a protocol validated and tested in 32 patients using multiparametric monitors. The heart rate, blood pressure, respiratory rate, oxygen saturation and ST segment-monitoring were measured and classified for clinical consistency one hour before and after the intervention, for 64 hours. Results: The protocol obtained Content Validity Index of 0.92. Of the 460 registered alarms, 261 were considered inconsistent before the intervention and 47 after it. The Relative Risk of inconsistent alarms after the protocol was 0.32 (95% CI 0.23-0.43, p <0.0001). Conclusion: The protocol proved to be a protective factor to the appearance of inconsistent clinical alarms of multiparametric monitors.
first_indexed 2024-12-21T06:56:01Z
format Article
id doaj.art-ac1b8fd2ea2b4370b550475524b6e6c6
institution Directory Open Access Journal
issn 1984-0446
language English
last_indexed 2024-12-21T06:56:01Z
publisher Associação Brasileira de Enfermagem
record_format Article
series Revista Brasileira de Enfermagem
spelling doaj.art-ac1b8fd2ea2b4370b550475524b6e6c62022-12-21T19:12:20ZengAssociação Brasileira de EnfermagemRevista Brasileira de Enfermagem1984-044672360961610.1590/0034-7167-2018-0485S0034-71672019000300609Individualized parameterization of multiparametric monitors alarms in infarcted patientsAllan Peixoto de AssisFrancimar Tinoco de OliveiraFlávia Giron CameriniRoberto Carlos Lyra da SilvaCamila Mendonça de MoraesABSTRACT Objective: To measure the magnitude of the effect of an individualized parameterization protocol for hemodynamic alarms in patients with acute myocardial infarction. Method: Pragmatic clinical trial, open label and single arm, whose intervention was performed through a protocol validated and tested in 32 patients using multiparametric monitors. The heart rate, blood pressure, respiratory rate, oxygen saturation and ST segment-monitoring were measured and classified for clinical consistency one hour before and after the intervention, for 64 hours. Results: The protocol obtained Content Validity Index of 0.92. Of the 460 registered alarms, 261 were considered inconsistent before the intervention and 47 after it. The Relative Risk of inconsistent alarms after the protocol was 0.32 (95% CI 0.23-0.43, p <0.0001). Conclusion: The protocol proved to be a protective factor to the appearance of inconsistent clinical alarms of multiparametric monitors.http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0034-71672019000300609&lng=en&tlng=enAlarmas ClínicasMonitor FisiológicoParámetrosInfarto de MiocardioCuidados de Enfermería
spellingShingle Allan Peixoto de Assis
Francimar Tinoco de Oliveira
Flávia Giron Camerini
Roberto Carlos Lyra da Silva
Camila Mendonça de Moraes
Individualized parameterization of multiparametric monitors alarms in infarcted patients
Revista Brasileira de Enfermagem
Alarmas Clínicas
Monitor Fisiológico
Parámetros
Infarto de Miocardio
Cuidados de Enfermería
title Individualized parameterization of multiparametric monitors alarms in infarcted patients
title_full Individualized parameterization of multiparametric monitors alarms in infarcted patients
title_fullStr Individualized parameterization of multiparametric monitors alarms in infarcted patients
title_full_unstemmed Individualized parameterization of multiparametric monitors alarms in infarcted patients
title_short Individualized parameterization of multiparametric monitors alarms in infarcted patients
title_sort individualized parameterization of multiparametric monitors alarms in infarcted patients
topic Alarmas Clínicas
Monitor Fisiológico
Parámetros
Infarto de Miocardio
Cuidados de Enfermería
url http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0034-71672019000300609&lng=en&tlng=en
work_keys_str_mv AT allanpeixotodeassis individualizedparameterizationofmultiparametricmonitorsalarmsininfarctedpatients
AT francimartinocodeoliveira individualizedparameterizationofmultiparametricmonitorsalarmsininfarctedpatients
AT flaviagironcamerini individualizedparameterizationofmultiparametricmonitorsalarmsininfarctedpatients
AT robertocarloslyradasilva individualizedparameterizationofmultiparametricmonitorsalarmsininfarctedpatients
AT camilamendoncademoraes individualizedparameterizationofmultiparametricmonitorsalarmsininfarctedpatients