Laboratory and in-flight evaluation of measurement uncertainties from a commercial Cloud Droplet Probe (CDP)

Laboratory and in-flight evaluations of uncertainties of measurements from a Cloud Droplet Probe (CDP) are presented. A description of a water-droplet-generating device, similar to those used in previous studies, is provided along with validation of droplet sizing and positioning. Seven experime...

Full description

Bibliographic Details
Main Authors: S. Faber, J. R. French, R. Jackson
Format: Article
Language:English
Published: Copernicus Publications 2018-06-01
Series:Atmospheric Measurement Techniques
Online Access:https://www.atmos-meas-tech.net/11/3645/2018/amt-11-3645-2018.pdf
_version_ 1817984150239444992
author S. Faber
J. R. French
R. Jackson
R. Jackson
author_facet S. Faber
J. R. French
R. Jackson
R. Jackson
author_sort S. Faber
collection DOAJ
description Laboratory and in-flight evaluations of uncertainties of measurements from a Cloud Droplet Probe (CDP) are presented. A description of a water-droplet-generating device, similar to those used in previous studies, is provided along with validation of droplet sizing and positioning. Seven experiments with droplet diameters of 9, 17, 24, 29, 34, 38, and 46 µm tested sizing and counting performance across a 10 µm resolution grid throughout the sample area of a CDP. Results indicate errors in sizing that depend on both droplet diameter and position within the sample area through which a droplet transited. The CDP undersized 9 µm droplets by 1–4 µm. Droplets with diameters of 17 and 24 µm were sized to within 2 µm, which is the nominal CDP bin width for droplets of that size. The majority of droplets larger than 17 µm were oversized by 2–4 µm, while a small percentage were severely undersized, by as much as 30 µm. This combination led to an artificial broadening and skewing of the spectra such that mean diameters from a near-monodisperse distribution compared well (within a few percent), while the median diameters were oversized by 5–15 %. This has implications on how users should calibrate their probes. Errors in higher-order moments were generally less than 10 %. Comparisons of liquid water content (LWC) calculated from the CDP and that measured from a Nevzorov hot-wire probe were conducted for 17 917 1 Hz in-cloud points. Although some differences were noted based on volume-weighted mean diameter and total droplet concentration, the CDP-estimated LWC exceeded that measured by the Nevzorov by approximately 20 %, more than twice the expected difference based on results of the laboratory tests and considerations of Nevzorov collection efficiency.
first_indexed 2024-04-13T23:41:53Z
format Article
id doaj.art-ad08862d364e4e01bdb2e8c788313225
institution Directory Open Access Journal
issn 1867-1381
1867-8548
language English
last_indexed 2024-04-13T23:41:53Z
publishDate 2018-06-01
publisher Copernicus Publications
record_format Article
series Atmospheric Measurement Techniques
spelling doaj.art-ad08862d364e4e01bdb2e8c7883132252022-12-22T02:24:29ZengCopernicus PublicationsAtmospheric Measurement Techniques1867-13811867-85482018-06-01113645365910.5194/amt-11-3645-2018Laboratory and in-flight evaluation of measurement uncertainties from a commercial Cloud Droplet Probe (CDP)S. Faber0J. R. French1R. Jackson2R. Jackson3Department of Atmospheric Science, University of Wyoming, Laramie, WY 82071, USADepartment of Atmospheric Science, University of Wyoming, Laramie, WY 82071, USADepartment of Atmospheric Science, University of Wyoming, Laramie, WY 82071, USApresent address: Argonne National Laboratory, Environmental Science Division, Argonne, IL 60439, USALaboratory and in-flight evaluations of uncertainties of measurements from a Cloud Droplet Probe (CDP) are presented. A description of a water-droplet-generating device, similar to those used in previous studies, is provided along with validation of droplet sizing and positioning. Seven experiments with droplet diameters of 9, 17, 24, 29, 34, 38, and 46 µm tested sizing and counting performance across a 10 µm resolution grid throughout the sample area of a CDP. Results indicate errors in sizing that depend on both droplet diameter and position within the sample area through which a droplet transited. The CDP undersized 9 µm droplets by 1–4 µm. Droplets with diameters of 17 and 24 µm were sized to within 2 µm, which is the nominal CDP bin width for droplets of that size. The majority of droplets larger than 17 µm were oversized by 2–4 µm, while a small percentage were severely undersized, by as much as 30 µm. This combination led to an artificial broadening and skewing of the spectra such that mean diameters from a near-monodisperse distribution compared well (within a few percent), while the median diameters were oversized by 5–15 %. This has implications on how users should calibrate their probes. Errors in higher-order moments were generally less than 10 %. Comparisons of liquid water content (LWC) calculated from the CDP and that measured from a Nevzorov hot-wire probe were conducted for 17 917 1 Hz in-cloud points. Although some differences were noted based on volume-weighted mean diameter and total droplet concentration, the CDP-estimated LWC exceeded that measured by the Nevzorov by approximately 20 %, more than twice the expected difference based on results of the laboratory tests and considerations of Nevzorov collection efficiency.https://www.atmos-meas-tech.net/11/3645/2018/amt-11-3645-2018.pdf
spellingShingle S. Faber
J. R. French
R. Jackson
R. Jackson
Laboratory and in-flight evaluation of measurement uncertainties from a commercial Cloud Droplet Probe (CDP)
Atmospheric Measurement Techniques
title Laboratory and in-flight evaluation of measurement uncertainties from a commercial Cloud Droplet Probe (CDP)
title_full Laboratory and in-flight evaluation of measurement uncertainties from a commercial Cloud Droplet Probe (CDP)
title_fullStr Laboratory and in-flight evaluation of measurement uncertainties from a commercial Cloud Droplet Probe (CDP)
title_full_unstemmed Laboratory and in-flight evaluation of measurement uncertainties from a commercial Cloud Droplet Probe (CDP)
title_short Laboratory and in-flight evaluation of measurement uncertainties from a commercial Cloud Droplet Probe (CDP)
title_sort laboratory and in flight evaluation of measurement uncertainties from a commercial cloud droplet probe cdp
url https://www.atmos-meas-tech.net/11/3645/2018/amt-11-3645-2018.pdf
work_keys_str_mv AT sfaber laboratoryandinflightevaluationofmeasurementuncertaintiesfromacommercialclouddropletprobecdp
AT jrfrench laboratoryandinflightevaluationofmeasurementuncertaintiesfromacommercialclouddropletprobecdp
AT rjackson laboratoryandinflightevaluationofmeasurementuncertaintiesfromacommercialclouddropletprobecdp
AT rjackson laboratoryandinflightevaluationofmeasurementuncertaintiesfromacommercialclouddropletprobecdp