Dosimetric Evaluation of Multislice CT Using AnthropomorphicHead Phantom
Purpose: The most general form to report dose in Multislice CT is the CTDI and DLP which are computed for several slices. The goal of the current study was to estimate actual doses and dose distribution during CT examinations in a head and neck anthropomorphic phantom. Methods: After construction of...
Main Authors: | , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Tehran University of Medical Sciences
2015-03-01
|
Series: | Frontiers in Biomedical Technologies |
Subjects: | |
Online Access: | https://fbt.tums.ac.ir/index.php/fbt/article/view/40 |
_version_ | 1811265822165827584 |
---|---|
author | Seyedeh Fatemeh Seyedatashi Mitra Athari Ahmad Bitarafan-Rajabi Hadi Hasanzadeh Mehravar Rafati Hamid Reza Pouraliakbar Ali Mohammadzadeh |
author_facet | Seyedeh Fatemeh Seyedatashi Mitra Athari Ahmad Bitarafan-Rajabi Hadi Hasanzadeh Mehravar Rafati Hamid Reza Pouraliakbar Ali Mohammadzadeh |
author_sort | Seyedeh Fatemeh Seyedatashi |
collection | DOAJ |
description | Purpose: The most general form to report dose in Multislice CT is the CTDI and DLP which are computed for several slices. The goal of the current study was to estimate actual doses and dose distribution during CT examinations in a head and neck anthropomorphic phantom.
Methods: After construction of the head and neck phantom using natural bone and paraffin wax with NaCl as impurity, several places were considered in different sites to fill with badges of Gafchromic film. These places include brain, Parotid, Thyroid and Lens of eye. Phantom was scanned at CT Angio and Spiral protocols with 10 and 256 slice scanners.
Results: Our findings showed that in 10 slice scanner, selected organ doses were in the range of 0.09-23.1 mSv while in the 256 slice scanner, it was in the range of 0.14-18.01 mSv. The CT Angio protocol has a higher organ dose at all.
Conclusion: In CT Angio protocol, organ dose (except for the lens of eye) is lower in 10 slice compared to 256 slice CT; the brain dose in both protocols has no difference statistically. In the spiral protocol, the dose in 256 slice scanner is lower than the 10 slice scanner which might be due to higher number of detector arrays in 256 slice scanner.Thyroid dose is mainly due to scattered radiation and because of strict beam collimation; it has a small value in all protocols. |
first_indexed | 2024-04-12T20:30:51Z |
format | Article |
id | doaj.art-ad5051eb07f34d52904848d107afd0d8 |
institution | Directory Open Access Journal |
issn | 2345-5837 |
language | English |
last_indexed | 2024-04-12T20:30:51Z |
publishDate | 2015-03-01 |
publisher | Tehran University of Medical Sciences |
record_format | Article |
series | Frontiers in Biomedical Technologies |
spelling | doaj.art-ad5051eb07f34d52904848d107afd0d82022-12-22T03:17:45ZengTehran University of Medical SciencesFrontiers in Biomedical Technologies2345-58372015-03-0121Dosimetric Evaluation of Multislice CT Using AnthropomorphicHead PhantomSeyedeh Fatemeh Seyedatashi0Mitra Athari1Ahmad Bitarafan-Rajabi2Hadi Hasanzadeh3Mehravar Rafati4Hamid Reza Pouraliakbar5Ali Mohammadzadeh6Islamic Azad University, Science and Research Branch, Tehran, Iran.Islamic Azad University, Science and Research Branch, Tehran, Iran.Cardiovascular interventional research center, Department of nuclear medicine, Rajaei Cardiovascular, Medical & Research Center, Iran University of medical sciences, Tehran, Iran.Cancer Research Center and Department of Medical Physics, Faculty of Medicine, Semnan University of Medical Sciences, Semnan, IranDepartment of Medical Physics, Faculty of Paramedical Sciences, Kashan University of Medical Science, Kashan, Iran.Rajaei Cardiovascular, Medical & Research Center, Radiology Department, Iran University of Medical sciences, Tehran, Iran.Rajaei Cardiovascular, Medical & Research Center, Radiology Department, Iran University of Medical sciences, Tehran, Iran.Purpose: The most general form to report dose in Multislice CT is the CTDI and DLP which are computed for several slices. The goal of the current study was to estimate actual doses and dose distribution during CT examinations in a head and neck anthropomorphic phantom. Methods: After construction of the head and neck phantom using natural bone and paraffin wax with NaCl as impurity, several places were considered in different sites to fill with badges of Gafchromic film. These places include brain, Parotid, Thyroid and Lens of eye. Phantom was scanned at CT Angio and Spiral protocols with 10 and 256 slice scanners. Results: Our findings showed that in 10 slice scanner, selected organ doses were in the range of 0.09-23.1 mSv while in the 256 slice scanner, it was in the range of 0.14-18.01 mSv. The CT Angio protocol has a higher organ dose at all. Conclusion: In CT Angio protocol, organ dose (except for the lens of eye) is lower in 10 slice compared to 256 slice CT; the brain dose in both protocols has no difference statistically. In the spiral protocol, the dose in 256 slice scanner is lower than the 10 slice scanner which might be due to higher number of detector arrays in 256 slice scanner.Thyroid dose is mainly due to scattered radiation and because of strict beam collimation; it has a small value in all protocols.https://fbt.tums.ac.ir/index.php/fbt/article/view/40Multidetector computed tomographyFilm dosimetryPhantomsImaging |
spellingShingle | Seyedeh Fatemeh Seyedatashi Mitra Athari Ahmad Bitarafan-Rajabi Hadi Hasanzadeh Mehravar Rafati Hamid Reza Pouraliakbar Ali Mohammadzadeh Dosimetric Evaluation of Multislice CT Using AnthropomorphicHead Phantom Frontiers in Biomedical Technologies Multidetector computed tomography Film dosimetry Phantoms Imaging |
title | Dosimetric Evaluation of Multislice CT Using AnthropomorphicHead Phantom |
title_full | Dosimetric Evaluation of Multislice CT Using AnthropomorphicHead Phantom |
title_fullStr | Dosimetric Evaluation of Multislice CT Using AnthropomorphicHead Phantom |
title_full_unstemmed | Dosimetric Evaluation of Multislice CT Using AnthropomorphicHead Phantom |
title_short | Dosimetric Evaluation of Multislice CT Using AnthropomorphicHead Phantom |
title_sort | dosimetric evaluation of multislice ct using anthropomorphichead phantom |
topic | Multidetector computed tomography Film dosimetry Phantoms Imaging |
url | https://fbt.tums.ac.ir/index.php/fbt/article/view/40 |
work_keys_str_mv | AT seyedehfatemehseyedatashi dosimetricevaluationofmultislicectusinganthropomorphicheadphantom AT mitraathari dosimetricevaluationofmultislicectusinganthropomorphicheadphantom AT ahmadbitarafanrajabi dosimetricevaluationofmultislicectusinganthropomorphicheadphantom AT hadihasanzadeh dosimetricevaluationofmultislicectusinganthropomorphicheadphantom AT mehravarrafati dosimetricevaluationofmultislicectusinganthropomorphicheadphantom AT hamidrezapouraliakbar dosimetricevaluationofmultislicectusinganthropomorphicheadphantom AT alimohammadzadeh dosimetricevaluationofmultislicectusinganthropomorphicheadphantom |