Comparison and Evaluation of the Effects of Rib and Lung Inhomogeneities on Lung Dose in Breast Brachytherapy using a Treatment Planning System and the MCNPX Code

Introduction: This study investigates to what extent the computed dose received by lung tissue in a commercially available treatment planning system (TPS) for 192Ir high-dose-rate breast brachytherapy is accurate in view of tissue inhomogeneities and presence of ribs. Materials and Methods: A CT sca...

Full description

Bibliographic Details
Main Authors: Hossein Salehi Yazdi, Mojtaba Shamsaei, Hamid Reza Shabani, Ramin Jaberi, Saeed Setayeshi
Format: Article
Language:English
Published: Mashhad University of Medical Sciences 2010-09-01
Series:Iranian Journal of Medical Physics
Subjects:
Online Access:http://ijmp.mums.ac.ir/article_7250_d033726042bbfec6c1812f2cdb332c1a.pdf
_version_ 1818305095783153664
author Hossein Salehi Yazdi
Mojtaba Shamsaei
Hamid Reza Shabani
Ramin Jaberi
Saeed Setayeshi
author_facet Hossein Salehi Yazdi
Mojtaba Shamsaei
Hamid Reza Shabani
Ramin Jaberi
Saeed Setayeshi
author_sort Hossein Salehi Yazdi
collection DOAJ
description Introduction: This study investigates to what extent the computed dose received by lung tissue in a commercially available treatment planning system (TPS) for 192Ir high-dose-rate breast brachytherapy is accurate in view of tissue inhomogeneities and presence of ribs. Materials and Methods: A CT scan of the breast was used to construct a patient-equivalent phantom in the clinical treatment planning system. An implant involving 13 plastic catheters and 383 programmed source dwell positions were simulated using the MCNPX code. Results: The results were compared with the corresponding commercial TPS in the form of isodoses and cumulative dose–volume histogram in breast, lung and ribs. The comparison of Monte Carlo results and TPS calculation showed that the isodoses greater than 62% in the breast that were located rather close to the implant or away from the breast curvature surface and lung boundary were in good agreement. TPS calculations, however, overestimated dose in the lung for lower isodose contours and points that were lying near the breast-air boundary and relatively away from the implant. Discussion and Conclusions: Taking into account the ribs and entering the actual data for breast, rib and lung, revealed an average overestimation of dose in lung in the TPS calculation.
first_indexed 2024-12-13T06:21:09Z
format Article
id doaj.art-ad5480d7949c4229916b63eca24cf1b6
institution Directory Open Access Journal
issn 2345-3672
2345-3672
language English
last_indexed 2024-12-13T06:21:09Z
publishDate 2010-09-01
publisher Mashhad University of Medical Sciences
record_format Article
series Iranian Journal of Medical Physics
spelling doaj.art-ad5480d7949c4229916b63eca24cf1b62022-12-21T23:56:50ZengMashhad University of Medical SciencesIranian Journal of Medical Physics2345-36722345-36722010-09-0173212710.22038/ijmp.2010.72507250Comparison and Evaluation of the Effects of Rib and Lung Inhomogeneities on Lung Dose in Breast Brachytherapy using a Treatment Planning System and the MCNPX CodeHossein Salehi Yazdi0Mojtaba Shamsaei1Hamid Reza Shabani2Ramin Jaberi3Saeed Setayeshi4M.Sc. Student of Medical Radiation Engineering , Nuclear Engineering and Physics Dept., Amirkabir University of Technology, Tehran, IranAssistant Professor, Nuclear Engineering and Physics Dept., Amirkabir University of Technology, Tehran, IranM.Sc. Student of Medical Physics, Medical Physics Dept., Tehran University of Medical Sciences, Tehran, IranRadiotherapy Physicist, Cancer Institute, Imam Khomeini Hospital, Tehran, IranAssociate Professor, Nuclear Engineering and Physics Dept., Amirkabir University of Technology, Tehran, IranIntroduction: This study investigates to what extent the computed dose received by lung tissue in a commercially available treatment planning system (TPS) for 192Ir high-dose-rate breast brachytherapy is accurate in view of tissue inhomogeneities and presence of ribs. Materials and Methods: A CT scan of the breast was used to construct a patient-equivalent phantom in the clinical treatment planning system. An implant involving 13 plastic catheters and 383 programmed source dwell positions were simulated using the MCNPX code. Results: The results were compared with the corresponding commercial TPS in the form of isodoses and cumulative dose–volume histogram in breast, lung and ribs. The comparison of Monte Carlo results and TPS calculation showed that the isodoses greater than 62% in the breast that were located rather close to the implant or away from the breast curvature surface and lung boundary were in good agreement. TPS calculations, however, overestimated dose in the lung for lower isodose contours and points that were lying near the breast-air boundary and relatively away from the implant. Discussion and Conclusions: Taking into account the ribs and entering the actual data for breast, rib and lung, revealed an average overestimation of dose in lung in the TPS calculation.http://ijmp.mums.ac.ir/article_7250_d033726042bbfec6c1812f2cdb332c1a.pdfBreast cancerBrachytherapyHigh Dose RateInhomogeneityMCNPXTreatment Planning System
spellingShingle Hossein Salehi Yazdi
Mojtaba Shamsaei
Hamid Reza Shabani
Ramin Jaberi
Saeed Setayeshi
Comparison and Evaluation of the Effects of Rib and Lung Inhomogeneities on Lung Dose in Breast Brachytherapy using a Treatment Planning System and the MCNPX Code
Iranian Journal of Medical Physics
Breast cancer
Brachytherapy
High Dose Rate
Inhomogeneity
MCNPX
Treatment Planning System
title Comparison and Evaluation of the Effects of Rib and Lung Inhomogeneities on Lung Dose in Breast Brachytherapy using a Treatment Planning System and the MCNPX Code
title_full Comparison and Evaluation of the Effects of Rib and Lung Inhomogeneities on Lung Dose in Breast Brachytherapy using a Treatment Planning System and the MCNPX Code
title_fullStr Comparison and Evaluation of the Effects of Rib and Lung Inhomogeneities on Lung Dose in Breast Brachytherapy using a Treatment Planning System and the MCNPX Code
title_full_unstemmed Comparison and Evaluation of the Effects of Rib and Lung Inhomogeneities on Lung Dose in Breast Brachytherapy using a Treatment Planning System and the MCNPX Code
title_short Comparison and Evaluation of the Effects of Rib and Lung Inhomogeneities on Lung Dose in Breast Brachytherapy using a Treatment Planning System and the MCNPX Code
title_sort comparison and evaluation of the effects of rib and lung inhomogeneities on lung dose in breast brachytherapy using a treatment planning system and the mcnpx code
topic Breast cancer
Brachytherapy
High Dose Rate
Inhomogeneity
MCNPX
Treatment Planning System
url http://ijmp.mums.ac.ir/article_7250_d033726042bbfec6c1812f2cdb332c1a.pdf
work_keys_str_mv AT hosseinsalehiyazdi comparisonandevaluationoftheeffectsofribandlunginhomogeneitiesonlungdoseinbreastbrachytherapyusingatreatmentplanningsystemandthemcnpxcode
AT mojtabashamsaei comparisonandevaluationoftheeffectsofribandlunginhomogeneitiesonlungdoseinbreastbrachytherapyusingatreatmentplanningsystemandthemcnpxcode
AT hamidrezashabani comparisonandevaluationoftheeffectsofribandlunginhomogeneitiesonlungdoseinbreastbrachytherapyusingatreatmentplanningsystemandthemcnpxcode
AT raminjaberi comparisonandevaluationoftheeffectsofribandlunginhomogeneitiesonlungdoseinbreastbrachytherapyusingatreatmentplanningsystemandthemcnpxcode
AT saeedsetayeshi comparisonandevaluationoftheeffectsofribandlunginhomogeneitiesonlungdoseinbreastbrachytherapyusingatreatmentplanningsystemandthemcnpxcode