Investigation of the Performance of Various Low-Cost Radon Monitors under Variable Environmental Conditions

A comparison of low-cost radon monitors was conducted at the Laboratory of Natural Radiation (LNR). The monitors we evaluated were EcoQube, RadonEye, RadonEye Plus2, Spirit, ViewPlus, ViewRadon and WavePlus. An AlphaGUARD monitor calibrated at the Laboratory of Environmental Radioactivity of the Uni...

Full description

Bibliographic Details
Main Authors: Daniel Rábago, Enrique Fernández, Santiago Celaya, Ismael Fuente, Alicia Fernández, Jorge Quindós, Raúl Rodriguez, Luis Quindós, Carlos Sainz
Format: Article
Language:English
Published: MDPI AG 2024-03-01
Series:Sensors
Subjects:
Online Access:https://www.mdpi.com/1424-8220/24/6/1836
_version_ 1797239397202001920
author Daniel Rábago
Enrique Fernández
Santiago Celaya
Ismael Fuente
Alicia Fernández
Jorge Quindós
Raúl Rodriguez
Luis Quindós
Carlos Sainz
author_facet Daniel Rábago
Enrique Fernández
Santiago Celaya
Ismael Fuente
Alicia Fernández
Jorge Quindós
Raúl Rodriguez
Luis Quindós
Carlos Sainz
author_sort Daniel Rábago
collection DOAJ
description A comparison of low-cost radon monitors was conducted at the Laboratory of Natural Radiation (LNR). The monitors we evaluated were EcoQube, RadonEye, RadonEye Plus2, Spirit, ViewPlus, ViewRadon and WavePlus. An AlphaGUARD monitor calibrated at the Laboratory of Environmental Radioactivity of the University of Cantabria (LaRUC), accredited for testing and calibration according to ISO/IEC 17025, provided the reference value of radon concentration. The temporal stability of the monitors was studied, obtaining a percentage of missing records ranged from 1% to 19% of the data. The main technical characteristics studied were temporal stability, measurement ranges, accuracy, correlation and response time. The main results show that the measurement ranges align with those specified by their manufacturers, with percentage differences with respect to the reference monitor of between 5% and 16%. The diversity found for response time is remarkable, with values ranging from 1 to 15 h, with Pearson correlation factors between 0.63 and 0.90.
first_indexed 2024-04-24T17:50:53Z
format Article
id doaj.art-b117323e6f0b4f05bc2942cdc5ad0525
institution Directory Open Access Journal
issn 1424-8220
language English
last_indexed 2024-04-24T17:50:53Z
publishDate 2024-03-01
publisher MDPI AG
record_format Article
series Sensors
spelling doaj.art-b117323e6f0b4f05bc2942cdc5ad05252024-03-27T14:03:55ZengMDPI AGSensors1424-82202024-03-01246183610.3390/s24061836Investigation of the Performance of Various Low-Cost Radon Monitors under Variable Environmental ConditionsDaniel Rábago0Enrique Fernández1Santiago Celaya2Ismael Fuente3Alicia Fernández4Jorge Quindós5Raúl Rodriguez6Luis Quindós7Carlos Sainz8Laboratory of Environmental Radioactivity (LaRUC), University of Cantabria, 39011 Santander, SpainLaboratory of Environmental Radioactivity (LaRUC), University of Cantabria, 39011 Santander, SpainLaboratory of Environmental Radioactivity (LaRUC), University of Cantabria, 39011 Santander, SpainLaboratory of Environmental Radioactivity (LaRUC), University of Cantabria, 39011 Santander, SpainLaboratory of Environmental Radioactivity (LaRUC), University of Cantabria, 39011 Santander, SpainLaboratory of Environmental Radioactivity (LaRUC), University of Cantabria, 39011 Santander, SpainSiglo 21 Consultores S.L., Av. Rosalía Castro, 94, Perillo, 15172 A Coruña, SpainLaboratory of Environmental Radioactivity (LaRUC), University of Cantabria, 39011 Santander, SpainLaboratory of Environmental Radioactivity (LaRUC), University of Cantabria, 39011 Santander, SpainA comparison of low-cost radon monitors was conducted at the Laboratory of Natural Radiation (LNR). The monitors we evaluated were EcoQube, RadonEye, RadonEye Plus2, Spirit, ViewPlus, ViewRadon and WavePlus. An AlphaGUARD monitor calibrated at the Laboratory of Environmental Radioactivity of the University of Cantabria (LaRUC), accredited for testing and calibration according to ISO/IEC 17025, provided the reference value of radon concentration. The temporal stability of the monitors was studied, obtaining a percentage of missing records ranged from 1% to 19% of the data. The main technical characteristics studied were temporal stability, measurement ranges, accuracy, correlation and response time. The main results show that the measurement ranges align with those specified by their manufacturers, with percentage differences with respect to the reference monitor of between 5% and 16%. The diversity found for response time is remarkable, with values ranging from 1 to 15 h, with Pearson correlation factors between 0.63 and 0.90.https://www.mdpi.com/1424-8220/24/6/1836response timeintercomparisonlow-cost radon monitor
spellingShingle Daniel Rábago
Enrique Fernández
Santiago Celaya
Ismael Fuente
Alicia Fernández
Jorge Quindós
Raúl Rodriguez
Luis Quindós
Carlos Sainz
Investigation of the Performance of Various Low-Cost Radon Monitors under Variable Environmental Conditions
Sensors
response time
intercomparison
low-cost radon monitor
title Investigation of the Performance of Various Low-Cost Radon Monitors under Variable Environmental Conditions
title_full Investigation of the Performance of Various Low-Cost Radon Monitors under Variable Environmental Conditions
title_fullStr Investigation of the Performance of Various Low-Cost Radon Monitors under Variable Environmental Conditions
title_full_unstemmed Investigation of the Performance of Various Low-Cost Radon Monitors under Variable Environmental Conditions
title_short Investigation of the Performance of Various Low-Cost Radon Monitors under Variable Environmental Conditions
title_sort investigation of the performance of various low cost radon monitors under variable environmental conditions
topic response time
intercomparison
low-cost radon monitor
url https://www.mdpi.com/1424-8220/24/6/1836
work_keys_str_mv AT danielrabago investigationoftheperformanceofvariouslowcostradonmonitorsundervariableenvironmentalconditions
AT enriquefernandez investigationoftheperformanceofvariouslowcostradonmonitorsundervariableenvironmentalconditions
AT santiagocelaya investigationoftheperformanceofvariouslowcostradonmonitorsundervariableenvironmentalconditions
AT ismaelfuente investigationoftheperformanceofvariouslowcostradonmonitorsundervariableenvironmentalconditions
AT aliciafernandez investigationoftheperformanceofvariouslowcostradonmonitorsundervariableenvironmentalconditions
AT jorgequindos investigationoftheperformanceofvariouslowcostradonmonitorsundervariableenvironmentalconditions
AT raulrodriguez investigationoftheperformanceofvariouslowcostradonmonitorsundervariableenvironmentalconditions
AT luisquindos investigationoftheperformanceofvariouslowcostradonmonitorsundervariableenvironmentalconditions
AT carlossainz investigationoftheperformanceofvariouslowcostradonmonitorsundervariableenvironmentalconditions