Aspheric versus Spherical Posterior Chamber Intraocular Lenses
Purpose: To compare spherical aberration and contrast sensitivity function following implantation of four different foldable posterior chamber intraocular lenses (IOLs), namely Sensar, Akreos AO, Tecnis, and AcrySof IQ. Methods: In this randomized clinical trial, 68 eyes of 68 patients with senile...
Main Authors: | , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Knowledge E
2010-01-01
|
Series: | Journal of Ophthalmic & Vision Research |
Subjects: | |
Online Access: | http://www.jovr.org/article.asp?issn=2008-322X;year=2010;volume=5;issue=4;spage=217;epage=222;aulast=Jafarinasab |
_version_ | 1828159153391009792 |
---|---|
author | Mohammad-Reza Jafarinasab Sepehr Feizi Ahmad-Reza Baghi Hossein Ziaie Mehdi Yaseri |
author_facet | Mohammad-Reza Jafarinasab Sepehr Feizi Ahmad-Reza Baghi Hossein Ziaie Mehdi Yaseri |
author_sort | Mohammad-Reza Jafarinasab |
collection | DOAJ |
description | Purpose: To compare spherical aberration and contrast sensitivity function following implantation of four different foldable posterior chamber intraocular lenses (IOLs), namely Sensar, Akreos AO, Tecnis, and AcrySof IQ.
Methods: In this randomized clinical trial, 68 eyes of 68 patients with senile cataracts underwent phacoemulsification and IOL implantation with Sensar (n=17), Akreos AO (n=17), Tecnis (n=17), or AcrySof IQ (n=17). Uncorrected visual acuity (UCVA) and best spectacle-corrected visual acuity (BSCVA), spherical aberration and contrast sensitivity function (CSF) were compared among the study groups, 3 months after surgery.
Results: There was no significant difference between the study groups in terms of age (P = 0.21). Mean postoperative BSCVA with Sensar, Akreos AO, Tecnis, and AcrySof IQ was 0.15±0.10, 0.12±0.9, 0.08±0.08, and 0.08±0.07 logMAR, respectively (P=0.08). Spherical aberration measured over a 4 mm pupil was significantly higher with Sensar and Akreos AO than the two other IOLs. The difference between Tecnis and AcrySof IQ was significantly in favor of the former IOL. Over a 6 mm pupil, spherical aberrations were comparable with Sensar and Akreos AO, furthermore spherical aberration was also comparable among eyes implanted with Akreos AO, AcrySof IQ, and Tecnis. Sensar yielded significantly inferior results as compared to Acrysof IQ and Tecnis. CSF with Sensar was inferior to the three aspheric IOLs at the majority of spatial frequencies. Tecnis yielded significantly better mesopic CSF at 1.5 and 3 cycles per degree spatial frequencies.
Conclusion: Tecnis and AcrySof IQ provided significantly better visual function as compared to Sensar and Akreos AO, especially with smaller pupil size. However, this difference diminished with increasing pupil size. |
first_indexed | 2024-04-11T23:56:41Z |
format | Article |
id | doaj.art-b512a40debd944c3b17e97b93dc7ea05 |
institution | Directory Open Access Journal |
issn | 2008-322X |
language | English |
last_indexed | 2024-04-11T23:56:41Z |
publishDate | 2010-01-01 |
publisher | Knowledge E |
record_format | Article |
series | Journal of Ophthalmic & Vision Research |
spelling | doaj.art-b512a40debd944c3b17e97b93dc7ea052022-12-22T03:56:21ZengKnowledge EJournal of Ophthalmic & Vision Research2008-322X2010-01-0154217222Aspheric versus Spherical Posterior Chamber Intraocular LensesMohammad-Reza JafarinasabSepehr FeiziAhmad-Reza BaghiHossein ZiaieMehdi YaseriPurpose: To compare spherical aberration and contrast sensitivity function following implantation of four different foldable posterior chamber intraocular lenses (IOLs), namely Sensar, Akreos AO, Tecnis, and AcrySof IQ. Methods: In this randomized clinical trial, 68 eyes of 68 patients with senile cataracts underwent phacoemulsification and IOL implantation with Sensar (n=17), Akreos AO (n=17), Tecnis (n=17), or AcrySof IQ (n=17). Uncorrected visual acuity (UCVA) and best spectacle-corrected visual acuity (BSCVA), spherical aberration and contrast sensitivity function (CSF) were compared among the study groups, 3 months after surgery. Results: There was no significant difference between the study groups in terms of age (P = 0.21). Mean postoperative BSCVA with Sensar, Akreos AO, Tecnis, and AcrySof IQ was 0.15±0.10, 0.12±0.9, 0.08±0.08, and 0.08±0.07 logMAR, respectively (P=0.08). Spherical aberration measured over a 4 mm pupil was significantly higher with Sensar and Akreos AO than the two other IOLs. The difference between Tecnis and AcrySof IQ was significantly in favor of the former IOL. Over a 6 mm pupil, spherical aberrations were comparable with Sensar and Akreos AO, furthermore spherical aberration was also comparable among eyes implanted with Akreos AO, AcrySof IQ, and Tecnis. Sensar yielded significantly inferior results as compared to Acrysof IQ and Tecnis. CSF with Sensar was inferior to the three aspheric IOLs at the majority of spatial frequencies. Tecnis yielded significantly better mesopic CSF at 1.5 and 3 cycles per degree spatial frequencies. Conclusion: Tecnis and AcrySof IQ provided significantly better visual function as compared to Sensar and Akreos AO, especially with smaller pupil size. However, this difference diminished with increasing pupil size.http://www.jovr.org/article.asp?issn=2008-322X;year=2010;volume=5;issue=4;spage=217;epage=222;aulast=JafarinasabIntraocular Lens; Sensar; Akreos AO; Tecnis; AcrySof IQ; Spherical Aberration; Contrast Sensitivity Function; Phacoemulsification |
spellingShingle | Mohammad-Reza Jafarinasab Sepehr Feizi Ahmad-Reza Baghi Hossein Ziaie Mehdi Yaseri Aspheric versus Spherical Posterior Chamber Intraocular Lenses Journal of Ophthalmic & Vision Research Intraocular Lens; Sensar; Akreos AO; Tecnis; AcrySof IQ; Spherical Aberration; Contrast Sensitivity Function; Phacoemulsification |
title | Aspheric versus Spherical Posterior Chamber Intraocular Lenses |
title_full | Aspheric versus Spherical Posterior Chamber Intraocular Lenses |
title_fullStr | Aspheric versus Spherical Posterior Chamber Intraocular Lenses |
title_full_unstemmed | Aspheric versus Spherical Posterior Chamber Intraocular Lenses |
title_short | Aspheric versus Spherical Posterior Chamber Intraocular Lenses |
title_sort | aspheric versus spherical posterior chamber intraocular lenses |
topic | Intraocular Lens; Sensar; Akreos AO; Tecnis; AcrySof IQ; Spherical Aberration; Contrast Sensitivity Function; Phacoemulsification |
url | http://www.jovr.org/article.asp?issn=2008-322X;year=2010;volume=5;issue=4;spage=217;epage=222;aulast=Jafarinasab |
work_keys_str_mv | AT mohammadrezajafarinasab asphericversussphericalposteriorchamberintraocularlenses AT sepehrfeizi asphericversussphericalposteriorchamberintraocularlenses AT ahmadrezabaghi asphericversussphericalposteriorchamberintraocularlenses AT hosseinziaie asphericversussphericalposteriorchamberintraocularlenses AT mehdiyaseri asphericversussphericalposteriorchamberintraocularlenses |