Aspheric versus Spherical Posterior Chamber Intraocular Lenses

Purpose: To compare spherical aberration and contrast sensitivity function following implantation of four different foldable posterior chamber intraocular lenses (IOLs), namely Sensar, Akreos AO, Tecnis, and AcrySof IQ. Methods: In this randomized clinical trial, 68 eyes of 68 patients with senile...

Full description

Bibliographic Details
Main Authors: Mohammad-Reza Jafarinasab, Sepehr Feizi, Ahmad-Reza Baghi, Hossein Ziaie, Mehdi Yaseri
Format: Article
Language:English
Published: Knowledge E 2010-01-01
Series:Journal of Ophthalmic & Vision Research
Subjects:
Online Access:http://www.jovr.org/article.asp?issn=2008-322X;year=2010;volume=5;issue=4;spage=217;epage=222;aulast=Jafarinasab
_version_ 1828159153391009792
author Mohammad-Reza Jafarinasab
Sepehr Feizi
Ahmad-Reza Baghi
Hossein Ziaie
Mehdi Yaseri
author_facet Mohammad-Reza Jafarinasab
Sepehr Feizi
Ahmad-Reza Baghi
Hossein Ziaie
Mehdi Yaseri
author_sort Mohammad-Reza Jafarinasab
collection DOAJ
description Purpose: To compare spherical aberration and contrast sensitivity function following implantation of four different foldable posterior chamber intraocular lenses (IOLs), namely Sensar, Akreos AO, Tecnis, and AcrySof IQ. Methods: In this randomized clinical trial, 68 eyes of 68 patients with senile cataracts underwent phacoemulsification and IOL implantation with Sensar (n=17), Akreos AO (n=17), Tecnis (n=17), or AcrySof IQ (n=17). Uncorrected visual acuity (UCVA) and best spectacle-corrected visual acuity (BSCVA), spherical aberration and contrast sensitivity function (CSF) were compared among the study groups, 3 months after surgery. Results: There was no significant difference between the study groups in terms of age (P = 0.21). Mean postoperative BSCVA with Sensar, Akreos AO, Tecnis, and AcrySof IQ was 0.15±0.10, 0.12±0.9, 0.08±0.08, and 0.08±0.07 logMAR, respectively (P=0.08). Spherical aberration measured over a 4 mm pupil was significantly higher with Sensar and Akreos AO than the two other IOLs. The difference between Tecnis and AcrySof IQ was significantly in favor of the former IOL. Over a 6 mm pupil, spherical aberrations were comparable with Sensar and Akreos AO, furthermore spherical aberration was also comparable among eyes implanted with Akreos AO, AcrySof IQ, and Tecnis. Sensar yielded significantly inferior results as compared to Acrysof IQ and Tecnis. CSF with Sensar was inferior to the three aspheric IOLs at the majority of spatial frequencies. Tecnis yielded significantly better mesopic CSF at 1.5 and 3 cycles per degree spatial frequencies. Conclusion: Tecnis and AcrySof IQ provided significantly better visual function as compared to Sensar and Akreos AO, especially with smaller pupil size. However, this difference diminished with increasing pupil size.
first_indexed 2024-04-11T23:56:41Z
format Article
id doaj.art-b512a40debd944c3b17e97b93dc7ea05
institution Directory Open Access Journal
issn 2008-322X
language English
last_indexed 2024-04-11T23:56:41Z
publishDate 2010-01-01
publisher Knowledge E
record_format Article
series Journal of Ophthalmic & Vision Research
spelling doaj.art-b512a40debd944c3b17e97b93dc7ea052022-12-22T03:56:21ZengKnowledge EJournal of Ophthalmic & Vision Research2008-322X2010-01-0154217222Aspheric versus Spherical Posterior Chamber Intraocular LensesMohammad-Reza JafarinasabSepehr FeiziAhmad-Reza BaghiHossein ZiaieMehdi YaseriPurpose: To compare spherical aberration and contrast sensitivity function following implantation of four different foldable posterior chamber intraocular lenses (IOLs), namely Sensar, Akreos AO, Tecnis, and AcrySof IQ. Methods: In this randomized clinical trial, 68 eyes of 68 patients with senile cataracts underwent phacoemulsification and IOL implantation with Sensar (n=17), Akreos AO (n=17), Tecnis (n=17), or AcrySof IQ (n=17). Uncorrected visual acuity (UCVA) and best spectacle-corrected visual acuity (BSCVA), spherical aberration and contrast sensitivity function (CSF) were compared among the study groups, 3 months after surgery. Results: There was no significant difference between the study groups in terms of age (P = 0.21). Mean postoperative BSCVA with Sensar, Akreos AO, Tecnis, and AcrySof IQ was 0.15±0.10, 0.12±0.9, 0.08±0.08, and 0.08±0.07 logMAR, respectively (P=0.08). Spherical aberration measured over a 4 mm pupil was significantly higher with Sensar and Akreos AO than the two other IOLs. The difference between Tecnis and AcrySof IQ was significantly in favor of the former IOL. Over a 6 mm pupil, spherical aberrations were comparable with Sensar and Akreos AO, furthermore spherical aberration was also comparable among eyes implanted with Akreos AO, AcrySof IQ, and Tecnis. Sensar yielded significantly inferior results as compared to Acrysof IQ and Tecnis. CSF with Sensar was inferior to the three aspheric IOLs at the majority of spatial frequencies. Tecnis yielded significantly better mesopic CSF at 1.5 and 3 cycles per degree spatial frequencies. Conclusion: Tecnis and AcrySof IQ provided significantly better visual function as compared to Sensar and Akreos AO, especially with smaller pupil size. However, this difference diminished with increasing pupil size.http://www.jovr.org/article.asp?issn=2008-322X;year=2010;volume=5;issue=4;spage=217;epage=222;aulast=JafarinasabIntraocular Lens; Sensar; Akreos AO; Tecnis; AcrySof IQ; Spherical Aberration; Contrast Sensitivity Function; Phacoemulsification
spellingShingle Mohammad-Reza Jafarinasab
Sepehr Feizi
Ahmad-Reza Baghi
Hossein Ziaie
Mehdi Yaseri
Aspheric versus Spherical Posterior Chamber Intraocular Lenses
Journal of Ophthalmic & Vision Research
Intraocular Lens; Sensar; Akreos AO; Tecnis; AcrySof IQ; Spherical Aberration; Contrast Sensitivity Function; Phacoemulsification
title Aspheric versus Spherical Posterior Chamber Intraocular Lenses
title_full Aspheric versus Spherical Posterior Chamber Intraocular Lenses
title_fullStr Aspheric versus Spherical Posterior Chamber Intraocular Lenses
title_full_unstemmed Aspheric versus Spherical Posterior Chamber Intraocular Lenses
title_short Aspheric versus Spherical Posterior Chamber Intraocular Lenses
title_sort aspheric versus spherical posterior chamber intraocular lenses
topic Intraocular Lens; Sensar; Akreos AO; Tecnis; AcrySof IQ; Spherical Aberration; Contrast Sensitivity Function; Phacoemulsification
url http://www.jovr.org/article.asp?issn=2008-322X;year=2010;volume=5;issue=4;spage=217;epage=222;aulast=Jafarinasab
work_keys_str_mv AT mohammadrezajafarinasab asphericversussphericalposteriorchamberintraocularlenses
AT sepehrfeizi asphericversussphericalposteriorchamberintraocularlenses
AT ahmadrezabaghi asphericversussphericalposteriorchamberintraocularlenses
AT hosseinziaie asphericversussphericalposteriorchamberintraocularlenses
AT mehdiyaseri asphericversussphericalposteriorchamberintraocularlenses