Comparison of Central Corneal Thickness Measurements with Four Different New Devices and Ultrasound Pachymetry

Objectives:To assess the central corneal thickness (CCT) with 5 different devices, evaluate the repeatability of the devices, and determine the possible relationship between thickness values and sex.Materials and Methods:The study included 308 eyes of 154 patients (76 women, 78 men) between the ages...

Full description

Bibliographic Details
Main Authors: Cem Şimşek, Cansu Kaya, Aylin Karalezli
Format: Article
Language:English
Published: Galenos Yayinevi 2022-10-01
Series:Türk Oftalmoloji Dergisi
Subjects:
Online Access: http://www.oftalmoloji.org/archives/archive-detail/article-preview/comparison-of-central-corneal-thickness-measuremen/53341
_version_ 1797907878105841664
author Cem Şimşek
Cansu Kaya
Aylin Karalezli
author_facet Cem Şimşek
Cansu Kaya
Aylin Karalezli
author_sort Cem Şimşek
collection DOAJ
description Objectives:To assess the central corneal thickness (CCT) with 5 different devices, evaluate the repeatability of the devices, and determine the possible relationship between thickness values and sex.Materials and Methods:The study included 308 eyes of 154 patients (76 women, 78 men) between the ages of 18-30 who presented to the Ophthalmology Clinic of Muğla Sıtkı Koçman University Training and Research Hospital. Autorefractor (Topcon, Japan), ultrasound pachymetry (UP) (Ceniscan, USA), high-resolution Pentacam (Oculus, USA), anterior segment-optical coherence tomography (AS-OCT) (Optovue, USA), and Spectralis AS-OCT (Heidelberg, Germany) measurements were assessed.Results:The mean age of the study participants was 23.2±0.2 years and the mean CCT was 540±14.1 μm, with no statistically significant difference in CCT between sexes (p>0.05). Mean CCT values were 557.0±26.7 μm with the autorefractor, 543.6±32.9 μm with UP, 533.8±30.2 μm with the Oculus Pentacam, 519.8±30.1 μm with Optovue AS-OCT, and 547.5±31.6 μm with Heidelberg AS-OCT. Pairwise comparisons between devices showed that the Optovue AS-OCT gave significantly lower CCT measurements than the autorefractor and Heidelberg AS-OCT device (p=0.027 and p=0.033, respectively). The coefficient of repeatability for autorefractor, UP, high-resolution Pentacam, Optovue AS-OCT, and Heidelberg AS-OCT CCT measurements were 1.51%, 2.46%, 3.72%, 2.57%, and 3.34%, respectively.Conclusion:Measurements made with five different devices showed that CCT was comparable and clinically usable. However, it was determined that the Optovue AS-OCT showed lower CCT values compare to other devices. When compared in terms of repeatability, it was found to be lower in the Pentacam than other devices.
first_indexed 2024-04-10T10:43:35Z
format Article
id doaj.art-be3d0573c66846b68e0c47d140839382
institution Directory Open Access Journal
issn 1300-0659
2147-2661
language English
last_indexed 2024-04-10T10:43:35Z
publishDate 2022-10-01
publisher Galenos Yayinevi
record_format Article
series Türk Oftalmoloji Dergisi
spelling doaj.art-be3d0573c66846b68e0c47d1408393822023-02-15T16:20:24ZengGalenos YayineviTürk Oftalmoloji Dergisi1300-06592147-26612022-10-0152531832310.4274/tjo.galenos.2021.2246513049054Comparison of Central Corneal Thickness Measurements with Four Different New Devices and Ultrasound PachymetryCem Şimşek0Cansu Kaya1Aylin Karalezli2 Muğla Sıtkı Koçman Üniversitesi Tıp Fakültesi, Göz Hastalıkları Anabilim Dalı, Muğla, Türkiye Muğla Sıtkı Koçman Üniversitesi Tıp Fakültesi, Göz Hastalıkları Anabilim Dalı, Muğla, Türkiye Muğla Sıtkı Koçman Üniversitesi Tıp Fakültesi, Göz Hastalıkları Anabilim Dalı, Muğla, Türkiye Objectives:To assess the central corneal thickness (CCT) with 5 different devices, evaluate the repeatability of the devices, and determine the possible relationship between thickness values and sex.Materials and Methods:The study included 308 eyes of 154 patients (76 women, 78 men) between the ages of 18-30 who presented to the Ophthalmology Clinic of Muğla Sıtkı Koçman University Training and Research Hospital. Autorefractor (Topcon, Japan), ultrasound pachymetry (UP) (Ceniscan, USA), high-resolution Pentacam (Oculus, USA), anterior segment-optical coherence tomography (AS-OCT) (Optovue, USA), and Spectralis AS-OCT (Heidelberg, Germany) measurements were assessed.Results:The mean age of the study participants was 23.2±0.2 years and the mean CCT was 540±14.1 μm, with no statistically significant difference in CCT between sexes (p>0.05). Mean CCT values were 557.0±26.7 μm with the autorefractor, 543.6±32.9 μm with UP, 533.8±30.2 μm with the Oculus Pentacam, 519.8±30.1 μm with Optovue AS-OCT, and 547.5±31.6 μm with Heidelberg AS-OCT. Pairwise comparisons between devices showed that the Optovue AS-OCT gave significantly lower CCT measurements than the autorefractor and Heidelberg AS-OCT device (p=0.027 and p=0.033, respectively). The coefficient of repeatability for autorefractor, UP, high-resolution Pentacam, Optovue AS-OCT, and Heidelberg AS-OCT CCT measurements were 1.51%, 2.46%, 3.72%, 2.57%, and 3.34%, respectively.Conclusion:Measurements made with five different devices showed that CCT was comparable and clinically usable. However, it was determined that the Optovue AS-OCT showed lower CCT values compare to other devices. When compared in terms of repeatability, it was found to be lower in the Pentacam than other devices. http://www.oftalmoloji.org/archives/archive-detail/article-preview/comparison-of-central-corneal-thickness-measuremen/53341 central corneal thicknesscorneapachymetryoptical coherence tomography
spellingShingle Cem Şimşek
Cansu Kaya
Aylin Karalezli
Comparison of Central Corneal Thickness Measurements with Four Different New Devices and Ultrasound Pachymetry
Türk Oftalmoloji Dergisi
central corneal thickness
cornea
pachymetry
optical coherence tomography
title Comparison of Central Corneal Thickness Measurements with Four Different New Devices and Ultrasound Pachymetry
title_full Comparison of Central Corneal Thickness Measurements with Four Different New Devices and Ultrasound Pachymetry
title_fullStr Comparison of Central Corneal Thickness Measurements with Four Different New Devices and Ultrasound Pachymetry
title_full_unstemmed Comparison of Central Corneal Thickness Measurements with Four Different New Devices and Ultrasound Pachymetry
title_short Comparison of Central Corneal Thickness Measurements with Four Different New Devices and Ultrasound Pachymetry
title_sort comparison of central corneal thickness measurements with four different new devices and ultrasound pachymetry
topic central corneal thickness
cornea
pachymetry
optical coherence tomography
url http://www.oftalmoloji.org/archives/archive-detail/article-preview/comparison-of-central-corneal-thickness-measuremen/53341
work_keys_str_mv AT cemsimsek comparisonofcentralcornealthicknessmeasurementswithfourdifferentnewdevicesandultrasoundpachymetry
AT cansukaya comparisonofcentralcornealthicknessmeasurementswithfourdifferentnewdevicesandultrasoundpachymetry
AT aylinkaralezli comparisonofcentralcornealthicknessmeasurementswithfourdifferentnewdevicesandultrasoundpachymetry