Comparison of Central Corneal Thickness Measurements with Four Different New Devices and Ultrasound Pachymetry
Objectives:To assess the central corneal thickness (CCT) with 5 different devices, evaluate the repeatability of the devices, and determine the possible relationship between thickness values and sex.Materials and Methods:The study included 308 eyes of 154 patients (76 women, 78 men) between the ages...
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Galenos Yayinevi
2022-10-01
|
Series: | Türk Oftalmoloji Dergisi |
Subjects: | |
Online Access: |
http://www.oftalmoloji.org/archives/archive-detail/article-preview/comparison-of-central-corneal-thickness-measuremen/53341
|
_version_ | 1797907878105841664 |
---|---|
author | Cem Şimşek Cansu Kaya Aylin Karalezli |
author_facet | Cem Şimşek Cansu Kaya Aylin Karalezli |
author_sort | Cem Şimşek |
collection | DOAJ |
description | Objectives:To assess the central corneal thickness (CCT) with 5 different devices, evaluate the repeatability of the devices, and determine the possible relationship between thickness values and sex.Materials and Methods:The study included 308 eyes of 154 patients (76 women, 78 men) between the ages of 18-30 who presented to the Ophthalmology Clinic of Muğla Sıtkı Koçman University Training and Research Hospital. Autorefractor (Topcon, Japan), ultrasound pachymetry (UP) (Ceniscan, USA), high-resolution Pentacam (Oculus, USA), anterior segment-optical coherence tomography (AS-OCT) (Optovue, USA), and Spectralis AS-OCT (Heidelberg, Germany) measurements were assessed.Results:The mean age of the study participants was 23.2±0.2 years and the mean CCT was 540±14.1 μm, with no statistically significant difference in CCT between sexes (p>0.05). Mean CCT values were 557.0±26.7 μm with the autorefractor, 543.6±32.9 μm with UP, 533.8±30.2 μm with the Oculus Pentacam, 519.8±30.1 μm with Optovue AS-OCT, and 547.5±31.6 μm with Heidelberg AS-OCT. Pairwise comparisons between devices showed that the Optovue AS-OCT gave significantly lower CCT measurements than the autorefractor and Heidelberg AS-OCT device (p=0.027 and p=0.033, respectively). The coefficient of repeatability for autorefractor, UP, high-resolution Pentacam, Optovue AS-OCT, and Heidelberg AS-OCT CCT measurements were 1.51%, 2.46%, 3.72%, 2.57%, and 3.34%, respectively.Conclusion:Measurements made with five different devices showed that CCT was comparable and clinically usable. However, it was determined that the Optovue AS-OCT showed lower CCT values compare to other devices. When compared in terms of repeatability, it was found to be lower in the Pentacam than other devices. |
first_indexed | 2024-04-10T10:43:35Z |
format | Article |
id | doaj.art-be3d0573c66846b68e0c47d140839382 |
institution | Directory Open Access Journal |
issn | 1300-0659 2147-2661 |
language | English |
last_indexed | 2024-04-10T10:43:35Z |
publishDate | 2022-10-01 |
publisher | Galenos Yayinevi |
record_format | Article |
series | Türk Oftalmoloji Dergisi |
spelling | doaj.art-be3d0573c66846b68e0c47d1408393822023-02-15T16:20:24ZengGalenos YayineviTürk Oftalmoloji Dergisi1300-06592147-26612022-10-0152531832310.4274/tjo.galenos.2021.2246513049054Comparison of Central Corneal Thickness Measurements with Four Different New Devices and Ultrasound PachymetryCem Şimşek0Cansu Kaya1Aylin Karalezli2 Muğla Sıtkı Koçman Üniversitesi Tıp Fakültesi, Göz Hastalıkları Anabilim Dalı, Muğla, Türkiye Muğla Sıtkı Koçman Üniversitesi Tıp Fakültesi, Göz Hastalıkları Anabilim Dalı, Muğla, Türkiye Muğla Sıtkı Koçman Üniversitesi Tıp Fakültesi, Göz Hastalıkları Anabilim Dalı, Muğla, Türkiye Objectives:To assess the central corneal thickness (CCT) with 5 different devices, evaluate the repeatability of the devices, and determine the possible relationship between thickness values and sex.Materials and Methods:The study included 308 eyes of 154 patients (76 women, 78 men) between the ages of 18-30 who presented to the Ophthalmology Clinic of Muğla Sıtkı Koçman University Training and Research Hospital. Autorefractor (Topcon, Japan), ultrasound pachymetry (UP) (Ceniscan, USA), high-resolution Pentacam (Oculus, USA), anterior segment-optical coherence tomography (AS-OCT) (Optovue, USA), and Spectralis AS-OCT (Heidelberg, Germany) measurements were assessed.Results:The mean age of the study participants was 23.2±0.2 years and the mean CCT was 540±14.1 μm, with no statistically significant difference in CCT between sexes (p>0.05). Mean CCT values were 557.0±26.7 μm with the autorefractor, 543.6±32.9 μm with UP, 533.8±30.2 μm with the Oculus Pentacam, 519.8±30.1 μm with Optovue AS-OCT, and 547.5±31.6 μm with Heidelberg AS-OCT. Pairwise comparisons between devices showed that the Optovue AS-OCT gave significantly lower CCT measurements than the autorefractor and Heidelberg AS-OCT device (p=0.027 and p=0.033, respectively). The coefficient of repeatability for autorefractor, UP, high-resolution Pentacam, Optovue AS-OCT, and Heidelberg AS-OCT CCT measurements were 1.51%, 2.46%, 3.72%, 2.57%, and 3.34%, respectively.Conclusion:Measurements made with five different devices showed that CCT was comparable and clinically usable. However, it was determined that the Optovue AS-OCT showed lower CCT values compare to other devices. When compared in terms of repeatability, it was found to be lower in the Pentacam than other devices. http://www.oftalmoloji.org/archives/archive-detail/article-preview/comparison-of-central-corneal-thickness-measuremen/53341 central corneal thicknesscorneapachymetryoptical coherence tomography |
spellingShingle | Cem Şimşek Cansu Kaya Aylin Karalezli Comparison of Central Corneal Thickness Measurements with Four Different New Devices and Ultrasound Pachymetry Türk Oftalmoloji Dergisi central corneal thickness cornea pachymetry optical coherence tomography |
title | Comparison of Central Corneal Thickness Measurements with Four Different New Devices and Ultrasound Pachymetry |
title_full | Comparison of Central Corneal Thickness Measurements with Four Different New Devices and Ultrasound Pachymetry |
title_fullStr | Comparison of Central Corneal Thickness Measurements with Four Different New Devices and Ultrasound Pachymetry |
title_full_unstemmed | Comparison of Central Corneal Thickness Measurements with Four Different New Devices and Ultrasound Pachymetry |
title_short | Comparison of Central Corneal Thickness Measurements with Four Different New Devices and Ultrasound Pachymetry |
title_sort | comparison of central corneal thickness measurements with four different new devices and ultrasound pachymetry |
topic | central corneal thickness cornea pachymetry optical coherence tomography |
url |
http://www.oftalmoloji.org/archives/archive-detail/article-preview/comparison-of-central-corneal-thickness-measuremen/53341
|
work_keys_str_mv | AT cemsimsek comparisonofcentralcornealthicknessmeasurementswithfourdifferentnewdevicesandultrasoundpachymetry AT cansukaya comparisonofcentralcornealthicknessmeasurementswithfourdifferentnewdevicesandultrasoundpachymetry AT aylinkaralezli comparisonofcentralcornealthicknessmeasurementswithfourdifferentnewdevicesandultrasoundpachymetry |