Electromagnetic Interference in a Private Swimming Pool

Although current lead design and filtering capabilities have greatly improved, Electromagnetic Interference (EMI) from environmental sources has been increasingly reported in patients with Cardiac Implantable Electronic Device (CIED) [1]. Few cases of inappropriate intracardiac Cardioverter Defibril...

Full description

Bibliographic Details
Main Authors: Sandia Iskandar, Madhav Lavu, Moustapha Atoui, Dhanunjaya Lakkireddy
Format: Article
Language:English
Published: Elsevier 2015-11-01
Series:Indian Pacing and Electrophysiology Journal
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S0972629216000097
Description
Summary:Although current lead design and filtering capabilities have greatly improved, Electromagnetic Interference (EMI) from environmental sources has been increasingly reported in patients with Cardiac Implantable Electronic Device (CIED) [1]. Few cases of inappropriate intracardiac Cardioverter Defibrillator (ICD) associated with swimming pool has been described [2]. Here we present a case of 64 year old male who presented with an interesting EMI signal that was subsequently identified to be related to AC current leak in his swimming pool.