Reliability of Single-Use PEEP-Valves Attached to Self-Inflating Bags during Manual Ventilation of Neonates--An In Vitro Study.

International resuscitation guidelines suggest to use positive end-expiratory pressure (PEEP) during manual ventilation of neonates. Aim of our study was to test the reliability of self-inflating bags (SIB) with single-use PEEP valves regarding PEEP delivery and the effect of different peak inflatio...

Full description

Bibliographic Details
Main Authors: Julia C Hartung, Silke Wilitzki, Marta Thio-Lluch, Arjan B te Pas, Gerd Schmalisch, Charles C Roehr
Format: Article
Language:English
Published: Public Library of Science (PLoS) 2016-01-01
Series:PLoS ONE
Online Access:http://europepmc.org/articles/PMC4767411?pdf=render
_version_ 1828486651376041984
author Julia C Hartung
Silke Wilitzki
Marta Thio-Lluch
Arjan B te Pas
Gerd Schmalisch
Charles C Roehr
author_facet Julia C Hartung
Silke Wilitzki
Marta Thio-Lluch
Arjan B te Pas
Gerd Schmalisch
Charles C Roehr
author_sort Julia C Hartung
collection DOAJ
description International resuscitation guidelines suggest to use positive end-expiratory pressure (PEEP) during manual ventilation of neonates. Aim of our study was to test the reliability of self-inflating bags (SIB) with single-use PEEP valves regarding PEEP delivery and the effect of different peak inflation pressures (PIP) and ventilation rates (VR) on the delivered PEEP.Ten new single-use PEEP valves from 5 manufacturers were tested by ventilating an intubated 1 kg neonatal manikin containing a lung model with a SIB that was actuated by an electromechanical plunger device. Standard settings: PIP 20 cmH2O, VR 60/min, flow 8 L/min. PEEP settings of 5 and 10 cmH2O were studied. A second test was conducted with settings of PIP 40 cmH2O and VR 40/min. The delivered PEEP was measured by a respiratory function monitor (CO2SMO+).Valves from one manufacturer delivered no relevant PEEP and were excluded. The remaining valves showed a continuous decay of the delivered pressure during expiration. The median (25th and 75th percentile) delivered PEEP with standard settings was 3.4(2.7-3.8) cmH2O when set to 5 cmH2O and 6.1(4.9-7.1) cmH2O when set to 10 cmH2O. Increasing the PIP from 20 to 40 cmH2O led to a median (25th and 75th percentile) decrease in PEEP to 2.3(1.8-2.7) cmH2O and 4.3(3.2-4.8) cmH2O; changing VR from 60 to 40/min led to a PEEP decrease to 2.8(2.1-3.3) cmH2O and 5.0(3.5-6.2) cmH2O for both PEEP settings.Single-use PEEP valves do not reliably deliver the set PEEP. PIP and VR have an effect on the delivered PEEP. Operators should be aware of these limitations when manually ventilating neonates.
first_indexed 2024-12-11T09:36:40Z
format Article
id doaj.art-c623a9598ff3476c95a99c35bcee8750
institution Directory Open Access Journal
issn 1932-6203
language English
last_indexed 2024-12-11T09:36:40Z
publishDate 2016-01-01
publisher Public Library of Science (PLoS)
record_format Article
series PLoS ONE
spelling doaj.art-c623a9598ff3476c95a99c35bcee87502022-12-22T01:12:51ZengPublic Library of Science (PLoS)PLoS ONE1932-62032016-01-01112e015022410.1371/journal.pone.0150224Reliability of Single-Use PEEP-Valves Attached to Self-Inflating Bags during Manual Ventilation of Neonates--An In Vitro Study.Julia C HartungSilke WilitzkiMarta Thio-LluchArjan B te PasGerd SchmalischCharles C RoehrInternational resuscitation guidelines suggest to use positive end-expiratory pressure (PEEP) during manual ventilation of neonates. Aim of our study was to test the reliability of self-inflating bags (SIB) with single-use PEEP valves regarding PEEP delivery and the effect of different peak inflation pressures (PIP) and ventilation rates (VR) on the delivered PEEP.Ten new single-use PEEP valves from 5 manufacturers were tested by ventilating an intubated 1 kg neonatal manikin containing a lung model with a SIB that was actuated by an electromechanical plunger device. Standard settings: PIP 20 cmH2O, VR 60/min, flow 8 L/min. PEEP settings of 5 and 10 cmH2O were studied. A second test was conducted with settings of PIP 40 cmH2O and VR 40/min. The delivered PEEP was measured by a respiratory function monitor (CO2SMO+).Valves from one manufacturer delivered no relevant PEEP and were excluded. The remaining valves showed a continuous decay of the delivered pressure during expiration. The median (25th and 75th percentile) delivered PEEP with standard settings was 3.4(2.7-3.8) cmH2O when set to 5 cmH2O and 6.1(4.9-7.1) cmH2O when set to 10 cmH2O. Increasing the PIP from 20 to 40 cmH2O led to a median (25th and 75th percentile) decrease in PEEP to 2.3(1.8-2.7) cmH2O and 4.3(3.2-4.8) cmH2O; changing VR from 60 to 40/min led to a PEEP decrease to 2.8(2.1-3.3) cmH2O and 5.0(3.5-6.2) cmH2O for both PEEP settings.Single-use PEEP valves do not reliably deliver the set PEEP. PIP and VR have an effect on the delivered PEEP. Operators should be aware of these limitations when manually ventilating neonates.http://europepmc.org/articles/PMC4767411?pdf=render
spellingShingle Julia C Hartung
Silke Wilitzki
Marta Thio-Lluch
Arjan B te Pas
Gerd Schmalisch
Charles C Roehr
Reliability of Single-Use PEEP-Valves Attached to Self-Inflating Bags during Manual Ventilation of Neonates--An In Vitro Study.
PLoS ONE
title Reliability of Single-Use PEEP-Valves Attached to Self-Inflating Bags during Manual Ventilation of Neonates--An In Vitro Study.
title_full Reliability of Single-Use PEEP-Valves Attached to Self-Inflating Bags during Manual Ventilation of Neonates--An In Vitro Study.
title_fullStr Reliability of Single-Use PEEP-Valves Attached to Self-Inflating Bags during Manual Ventilation of Neonates--An In Vitro Study.
title_full_unstemmed Reliability of Single-Use PEEP-Valves Attached to Self-Inflating Bags during Manual Ventilation of Neonates--An In Vitro Study.
title_short Reliability of Single-Use PEEP-Valves Attached to Self-Inflating Bags during Manual Ventilation of Neonates--An In Vitro Study.
title_sort reliability of single use peep valves attached to self inflating bags during manual ventilation of neonates an in vitro study
url http://europepmc.org/articles/PMC4767411?pdf=render
work_keys_str_mv AT juliachartung reliabilityofsingleusepeepvalvesattachedtoselfinflatingbagsduringmanualventilationofneonatesaninvitrostudy
AT silkewilitzki reliabilityofsingleusepeepvalvesattachedtoselfinflatingbagsduringmanualventilationofneonatesaninvitrostudy
AT martathiolluch reliabilityofsingleusepeepvalvesattachedtoselfinflatingbagsduringmanualventilationofneonatesaninvitrostudy
AT arjanbtepas reliabilityofsingleusepeepvalvesattachedtoselfinflatingbagsduringmanualventilationofneonatesaninvitrostudy
AT gerdschmalisch reliabilityofsingleusepeepvalvesattachedtoselfinflatingbagsduringmanualventilationofneonatesaninvitrostudy
AT charlescroehr reliabilityofsingleusepeepvalvesattachedtoselfinflatingbagsduringmanualventilationofneonatesaninvitrostudy