Mathematical analysis of the effects of valvular regurgitation on the pumping efficacy of continuous and pulsatile left ventricular assist devices

Background: A left ventricular assist device (LVAD) is normally contraindicated in significant aortic regurgitation (AR) and requires intraoperative valve repair or exclusion. Nevertheless, AR can coexist with an LVAD, so a valid question when asked might still be of clinical significance. The purpo...

Full description

Bibliographic Details
Main Authors: Yoo Seok Kim, Eun-Hye Kim, Hyeong-Gyun Kim, Eun Bo Shim, Kwang-Soup Song, Ki Moo Lim
Format: Article
Language:English
Published: Elsevier 2016-03-01
Series:Integrative Medicine Research
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S2213422016000020
Description
Summary:Background: A left ventricular assist device (LVAD) is normally contraindicated in significant aortic regurgitation (AR) and requires intraoperative valve repair or exclusion. Nevertheless, AR can coexist with an LVAD, so a valid question when asked might still be of clinical significance. The purpose of this study is to analyze the effects of valve regurgitation on the pumping efficacy of continuous and pulsatile LVADs with a computational method. Methods: A cardiovascular model was developed based on the Windkessel model, which reflects the hemodynamic flow resistance and the blood wall elasticity. Using the Windkessel model, important cardiovascular components, such as the right atrium, right ventricle, pulmonary artery, pulmonary vein, left atrium (LA), left ventricle (LV), aorta, and branching blood vessels, were expressed. Results: In the case of AR, continuous and pulsatile LVADs improved cardiac output and reduced mechanical load slightly. In the case of mitral regurgitation, the LVADs improved cardiac output (cardiac outputs were about 5 L/min regardless of the severity of regurgitation) and reduced afterload significantly. Conclusion: AR reduced both continuous and pulsatile LVAD function significantly while mitral regurgitation did not affect their pumping efficacy.
ISSN:2213-4220