Evaluation of the measurement uncertainty of the stiffness modulus: test case of indirect tensile on cylindrical specimens

This study treats the measurement uncertainties that we can find in the stiffness modulus of the bituminous test. We present all the sensors installed on rigidity modulus measurement chains and also their uncertainty ranges. Several parameters influence the rigidity module's value, such as the...

Full description

Bibliographic Details
Main Authors: Mezouara Hicham, Dlimi Latifa, Salih Abdelouahhab, Afechcar Mohamed, Zniker Houcine
Format: Article
Language:English
Published: EDP Sciences 2021-01-01
Series:International Journal of Metrology and Quality Engineering
Subjects:
Online Access:https://www.metrology-journal.org/articles/ijmqe/full_html/2021/01/ijmqe180010/ijmqe180010.html
Description
Summary:This study treats the measurement uncertainties that we can find in the stiffness modulus of the bituminous test. We present all the sensors installed on rigidity modulus measurement chains and also their uncertainty ranges. Several parameters influence the rigidity module's value, such as the parameters related to experimental conditions, and others are rather connected to the equipment's specification, which are the speed, the loading level, the temperature, the tested sample dimension, and the data acquisition, etc. All these factors have a great influence on the value of the modulus of rigidity. To qualify the uncertainty factors, we used two approaches: the first one is made by following the method described by the GUM (Guide to the expression of uncertainty in measurement), the second approach based on the numerical simulation of the Monte Carlo. The two results are then compared for an interval of confidence of 95%. The paper also shows the employment of the basic methods of statistical analysis, such as the Comparing of two variances. Essential concepts in measurement uncertainty have been compiled and the determination of the stiffness module parameters are discussed. It has been demonstrated that the biggest source of error in the stiffness modulus measuring process is the repeatability has a contribution of around 45.23%.
ISSN:2107-6847