Agreement of Tear Break-Up Time and Meniscus Height between Medmont E300 and Visionix VX120+

The goal of this study was to analyze the agreement between the Medmont E300 and the Visionix VX120+ systems in terms of non-invasive tear break-up time (NIBUT) and tear meniscus height (TMH) measurements. A total of 60 eyes (30 healthy subjects) were enrolled. NIBUT and TMH were evaluated with Medm...

Full description

Bibliographic Details
Main Authors: Elena Martínez-Plaza, Ainhoa Molina-Martín, David P. Piñero
Format: Article
Language:English
Published: MDPI AG 2022-04-01
Series:Applied Sciences
Subjects:
Online Access:https://www.mdpi.com/2076-3417/12/9/4589
_version_ 1797505560126423040
author Elena Martínez-Plaza
Ainhoa Molina-Martín
David P. Piñero
author_facet Elena Martínez-Plaza
Ainhoa Molina-Martín
David P. Piñero
author_sort Elena Martínez-Plaza
collection DOAJ
description The goal of this study was to analyze the agreement between the Medmont E300 and the Visionix VX120+ systems in terms of non-invasive tear break-up time (NIBUT) and tear meniscus height (TMH) measurements. A total of 60 eyes (30 healthy subjects) were enrolled. NIBUT and TMH were evaluated with Medmont E300; first NIBUT, NIBUT50%, and TMH were evaluated with Visionix VX120+. Both evaluations were performed in a random order by the same clinician for right, left, and both eyes. The Medmont E300 provided significantly higher NIBUT than Visionix VX120+ for first NIBUT in right, left, and both eyes (<i>p</i> ≤ 0.003) and NIBUT50% in left and both eyes (<i>p</i> ≤ 0.042). The TMH measured with VX120+ was significantly higher than with Medmont E300 considering both eyes (<i>p</i> = 0.037). No significant correlations were found between both devices for either NIBUT (<i>p</i> ≥ 0.11) or TMH (<i>p</i> ≥ 0.09). Passing–Bablok regression analyses revealed poor agreement between devices for NIBUT and TMH outcomes. VX120+ is expected to provide substantial lower first NIBUT values than the NIBUT measured by Medmont E300. Clinicians should consider not using both instruments as interchangeable for dry eye diagnosis.
first_indexed 2024-03-10T04:20:20Z
format Article
id doaj.art-c9692c501d1a4177b99af2e521312fb3
institution Directory Open Access Journal
issn 2076-3417
language English
last_indexed 2024-03-10T04:20:20Z
publishDate 2022-04-01
publisher MDPI AG
record_format Article
series Applied Sciences
spelling doaj.art-c9692c501d1a4177b99af2e521312fb32023-11-23T07:51:10ZengMDPI AGApplied Sciences2076-34172022-04-01129458910.3390/app12094589Agreement of Tear Break-Up Time and Meniscus Height between Medmont E300 and Visionix VX120+Elena Martínez-Plaza0Ainhoa Molina-Martín1David P. Piñero2Group of Optics and Visual Perception, Department of Optics, Pharmacology and Anatomy, University of Alicante, 03690 Alicante, SpainGroup of Optics and Visual Perception, Department of Optics, Pharmacology and Anatomy, University of Alicante, 03690 Alicante, SpainGroup of Optics and Visual Perception, Department of Optics, Pharmacology and Anatomy, University of Alicante, 03690 Alicante, SpainThe goal of this study was to analyze the agreement between the Medmont E300 and the Visionix VX120+ systems in terms of non-invasive tear break-up time (NIBUT) and tear meniscus height (TMH) measurements. A total of 60 eyes (30 healthy subjects) were enrolled. NIBUT and TMH were evaluated with Medmont E300; first NIBUT, NIBUT50%, and TMH were evaluated with Visionix VX120+. Both evaluations were performed in a random order by the same clinician for right, left, and both eyes. The Medmont E300 provided significantly higher NIBUT than Visionix VX120+ for first NIBUT in right, left, and both eyes (<i>p</i> ≤ 0.003) and NIBUT50% in left and both eyes (<i>p</i> ≤ 0.042). The TMH measured with VX120+ was significantly higher than with Medmont E300 considering both eyes (<i>p</i> = 0.037). No significant correlations were found between both devices for either NIBUT (<i>p</i> ≥ 0.11) or TMH (<i>p</i> ≥ 0.09). Passing–Bablok regression analyses revealed poor agreement between devices for NIBUT and TMH outcomes. VX120+ is expected to provide substantial lower first NIBUT values than the NIBUT measured by Medmont E300. Clinicians should consider not using both instruments as interchangeable for dry eye diagnosis.https://www.mdpi.com/2076-3417/12/9/4589non-invasive tear break-up timeNIBUTtear meniscus heightTMHagreementinterchangeability
spellingShingle Elena Martínez-Plaza
Ainhoa Molina-Martín
David P. Piñero
Agreement of Tear Break-Up Time and Meniscus Height between Medmont E300 and Visionix VX120+
Applied Sciences
non-invasive tear break-up time
NIBUT
tear meniscus height
TMH
agreement
interchangeability
title Agreement of Tear Break-Up Time and Meniscus Height between Medmont E300 and Visionix VX120+
title_full Agreement of Tear Break-Up Time and Meniscus Height between Medmont E300 and Visionix VX120+
title_fullStr Agreement of Tear Break-Up Time and Meniscus Height between Medmont E300 and Visionix VX120+
title_full_unstemmed Agreement of Tear Break-Up Time and Meniscus Height between Medmont E300 and Visionix VX120+
title_short Agreement of Tear Break-Up Time and Meniscus Height between Medmont E300 and Visionix VX120+
title_sort agreement of tear break up time and meniscus height between medmont e300 and visionix vx120
topic non-invasive tear break-up time
NIBUT
tear meniscus height
TMH
agreement
interchangeability
url https://www.mdpi.com/2076-3417/12/9/4589
work_keys_str_mv AT elenamartinezplaza agreementoftearbreakuptimeandmeniscusheightbetweenmedmonte300andvisionixvx120
AT ainhoamolinamartin agreementoftearbreakuptimeandmeniscusheightbetweenmedmonte300andvisionixvx120
AT davidppinero agreementoftearbreakuptimeandmeniscusheightbetweenmedmonte300andvisionixvx120