Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter.

<h4>Objective</h4>To compare the radiation dose and the objective and subjective image quality of 80 kVp and 80/150 kVp with tin filter (80/Sn150 kVp) computed tomography (CT) in oncology patients.<h4>Methods</h4>One-hundred-and-forty-five consecutive oncology patients who un...

Full description

Bibliographic Details
Main Authors: Seung Joon Choi, Su Joa Ahn, So Hyun Park, Seong Ho Park, Seong Yong Pak, Jae Won Choi, Young Sup Shim, Yu Mi Jeong, Bohyun Kim
Format: Article
Language:English
Published: Public Library of Science (PLoS) 2020-01-01
Series:PLoS ONE
Online Access:https://doi.org/10.1371/journal.pone.0231431
_version_ 1818400971208785920
author Seung Joon Choi
Su Joa Ahn
So Hyun Park
Seong Ho Park
Seong Yong Pak
Jae Won Choi
Young Sup Shim
Yu Mi Jeong
Bohyun Kim
author_facet Seung Joon Choi
Su Joa Ahn
So Hyun Park
Seong Ho Park
Seong Yong Pak
Jae Won Choi
Young Sup Shim
Yu Mi Jeong
Bohyun Kim
author_sort Seung Joon Choi
collection DOAJ
description <h4>Objective</h4>To compare the radiation dose and the objective and subjective image quality of 80 kVp and 80/150 kVp with tin filter (80/Sn150 kVp) computed tomography (CT) in oncology patients.<h4>Methods</h4>One-hundred-and-forty-five consecutive oncology patients who underwent third-generation dual-source dual-energy CT of the abdomen for evaluation of malignant visceral, peritoneal, extraperitoneal, and bone tumor were retrospectively recruited. Two radiologists independently reviewed each observation in 80 kVp CT and 80/Sn150 kVp CT. Modified line-density profile of the tumor and contrast-to-noise ratio (CNR) were measured. Diagnostic confidence, lesion conspicuity, and subjective image quality were calculated and compared between image sets. The effective dose and size-specific dose estimate (SSDE) were calculated in the image sets.<h4>Results</h4>Modified line-density profile analysis revealed higher attenuation differences between the tumor and normal tissue in 80 kVp CT than in 80/Sn150 kVp CT (127 vs. 107, P = 0.05). The 80 kVp CT showed increased CNR in the liver (8.0 vs. 7.6) and the aorta (18.9 vs. 16.3) than the 80/Sn150 kVp CT. The 80 kVp CT yielded higher enhancement of organs (4.9 ± 0.2 vs. 4.7 ± 0.4, P<0.001) and lesion conspicuity (4.9 ± 0.3 vs. 4.8 ± 0.5, P = 0.035) than the 80/Sn150 kVp CT; overall image quality and confidence index were comparable. The effective dose was reduced by 45.2% with 80 kVp CT (2.3 mSv ± 0.9) compared to 80/Sn150 kVp CT (4.1 mSv ± 1.5). The SSDE was 7.4 ± 3.8 mGy on 80/Sn150 kVp CT and 4.1 ± 2.2 mGy on 80 kVp CT.<h4>Conclusions</h4>The 80 kVp CT reduced the radiation dose by 45.2% in oncology patients while showing comparable or superior image quality to that of 80/Sn150 kVp CT for abdominal tumor evaluation.
first_indexed 2024-12-14T07:45:03Z
format Article
id doaj.art-ce39795442d14040a00949912c666e7c
institution Directory Open Access Journal
issn 1932-6203
language English
last_indexed 2024-12-14T07:45:03Z
publishDate 2020-01-01
publisher Public Library of Science (PLoS)
record_format Article
series PLoS ONE
spelling doaj.art-ce39795442d14040a00949912c666e7c2022-12-21T23:10:54ZengPublic Library of Science (PLoS)PLoS ONE1932-62032020-01-01159e023143110.1371/journal.pone.0231431Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter.Seung Joon ChoiSu Joa AhnSo Hyun ParkSeong Ho ParkSeong Yong PakJae Won ChoiYoung Sup ShimYu Mi JeongBohyun Kim<h4>Objective</h4>To compare the radiation dose and the objective and subjective image quality of 80 kVp and 80/150 kVp with tin filter (80/Sn150 kVp) computed tomography (CT) in oncology patients.<h4>Methods</h4>One-hundred-and-forty-five consecutive oncology patients who underwent third-generation dual-source dual-energy CT of the abdomen for evaluation of malignant visceral, peritoneal, extraperitoneal, and bone tumor were retrospectively recruited. Two radiologists independently reviewed each observation in 80 kVp CT and 80/Sn150 kVp CT. Modified line-density profile of the tumor and contrast-to-noise ratio (CNR) were measured. Diagnostic confidence, lesion conspicuity, and subjective image quality were calculated and compared between image sets. The effective dose and size-specific dose estimate (SSDE) were calculated in the image sets.<h4>Results</h4>Modified line-density profile analysis revealed higher attenuation differences between the tumor and normal tissue in 80 kVp CT than in 80/Sn150 kVp CT (127 vs. 107, P = 0.05). The 80 kVp CT showed increased CNR in the liver (8.0 vs. 7.6) and the aorta (18.9 vs. 16.3) than the 80/Sn150 kVp CT. The 80 kVp CT yielded higher enhancement of organs (4.9 ± 0.2 vs. 4.7 ± 0.4, P<0.001) and lesion conspicuity (4.9 ± 0.3 vs. 4.8 ± 0.5, P = 0.035) than the 80/Sn150 kVp CT; overall image quality and confidence index were comparable. The effective dose was reduced by 45.2% with 80 kVp CT (2.3 mSv ± 0.9) compared to 80/Sn150 kVp CT (4.1 mSv ± 1.5). The SSDE was 7.4 ± 3.8 mGy on 80/Sn150 kVp CT and 4.1 ± 2.2 mGy on 80 kVp CT.<h4>Conclusions</h4>The 80 kVp CT reduced the radiation dose by 45.2% in oncology patients while showing comparable or superior image quality to that of 80/Sn150 kVp CT for abdominal tumor evaluation.https://doi.org/10.1371/journal.pone.0231431
spellingShingle Seung Joon Choi
Su Joa Ahn
So Hyun Park
Seong Ho Park
Seong Yong Pak
Jae Won Choi
Young Sup Shim
Yu Mi Jeong
Bohyun Kim
Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter.
PLoS ONE
title Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter.
title_full Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter.
title_fullStr Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter.
title_full_unstemmed Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter.
title_short Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter.
title_sort dual source abdominopelvic computed tomography comparison of image quality and radiation dose of 80 kvp and 80 150 kvp with tin filter
url https://doi.org/10.1371/journal.pone.0231431
work_keys_str_mv AT seungjoonchoi dualsourceabdominopelviccomputedtomographycomparisonofimagequalityandradiationdoseof80kvpand80150kvpwithtinfilter
AT sujoaahn dualsourceabdominopelviccomputedtomographycomparisonofimagequalityandradiationdoseof80kvpand80150kvpwithtinfilter
AT sohyunpark dualsourceabdominopelviccomputedtomographycomparisonofimagequalityandradiationdoseof80kvpand80150kvpwithtinfilter
AT seonghopark dualsourceabdominopelviccomputedtomographycomparisonofimagequalityandradiationdoseof80kvpand80150kvpwithtinfilter
AT seongyongpak dualsourceabdominopelviccomputedtomographycomparisonofimagequalityandradiationdoseof80kvpand80150kvpwithtinfilter
AT jaewonchoi dualsourceabdominopelviccomputedtomographycomparisonofimagequalityandradiationdoseof80kvpand80150kvpwithtinfilter
AT youngsupshim dualsourceabdominopelviccomputedtomographycomparisonofimagequalityandradiationdoseof80kvpand80150kvpwithtinfilter
AT yumijeong dualsourceabdominopelviccomputedtomographycomparisonofimagequalityandradiationdoseof80kvpand80150kvpwithtinfilter
AT bohyunkim dualsourceabdominopelviccomputedtomographycomparisonofimagequalityandradiationdoseof80kvpand80150kvpwithtinfilter