Evaluation of the dose reduction capabilities in digital radiography of the chest using contrast-detail phantom

Assessment of the quality of the images obtained using optimized (low-dose) protocols is the inherent part of the optimization in X-ray diagnostics. To perform the objective quantitative image quality assessment one can use dedicated test-objects, including several components for the simultaneous me...

Full description

Bibliographic Details
Main Author: A. V. Vodovatov
Format: Article
Language:English
Published: Saint-Petersburg Research Institute of Radiation Hygiene after Professor P.V. Ramzaev 2019-03-01
Series:Радиационная гигиена
Subjects:
Online Access:https://www.radhyg.ru/jour/article/view/596
_version_ 1797873815763550208
author A. V. Vodovatov
author_facet A. V. Vodovatov
author_sort A. V. Vodovatov
collection DOAJ
description Assessment of the quality of the images obtained using optimized (low-dose) protocols is the inherent part of the optimization in X-ray diagnostics. To perform the objective quantitative image quality assessment one can use dedicated test-objects, including several components for the simultaneous measurement of the different physical image characteristics (contrast and spatial resolution). The use of such test objects allows estimating and assessing the relations between the patient dose, parameter of the X-ray examination and image quality. That is especially important for the optimization of the digital radiographic examinations performed with automated exposure control. The aim of the current study was to evaluate the possibilities of the patient dose reduction using “contrast-detail” test-object for the digital radiography of the chest in posterior-anterior projection performed with automated exposure control. The study was performed in St-Petersburg Mariinsky hospital on a digital X-ray unit “ARC-Electron” with a flat-panel detector. The combination of a test-object and a tissue-equivalent phantom were imaged on a range of chest X-ray protocols: on a 60–150 kV tube voltage range with automated exposure control; and using fixed 90 kV tube voltage on a range of 2–100 mAs tube current-exposure time product. Dose-area product (cGy×cm2) was measured for each exposure; effective dose (mSv) was estimated for each exposure based on dose-area product. A dedicated software was developed for the automated image quality assessment. The results of the study indicate that the use of a high tube voltage (140–150 kV) with current automated exposure control settings would lead to 60% and 95% reduction of the dose-area product and effective dose, respectively, compared to the standard protocol. The adjustment of the current automated exposure control settings with the reduction of the tube current-exposure time product from 11,2 mAs to the 4,2 mAs for the tube voltage of 90 kV would lead to the reduction of both the dose-area product and effective dose up to a factor of three, compared to the standard protocol. For both scenarios image quality characteristics decreased by less than 15%. The proposed low-dose protocols are under the clinical approbation at Mariinsky hospital. The proposed method of image quality assessment and development of low-dose protocols is recommended for inclusion in the quality assurance program for the radiography examinations.
first_indexed 2024-04-10T01:22:13Z
format Article
id doaj.art-ce7bebefd3974c1099cd39e5d674f29c
institution Directory Open Access Journal
issn 1998-426X
language English
last_indexed 2024-04-10T01:22:13Z
publishDate 2019-03-01
publisher Saint-Petersburg Research Institute of Radiation Hygiene after Professor P.V. Ramzaev
record_format Article
series Радиационная гигиена
spelling doaj.art-ce7bebefd3974c1099cd39e5d674f29c2023-03-13T09:47:09ZengSaint-Petersburg Research Institute of Radiation Hygiene after Professor P.V. RamzaevРадиационная гигиена1998-426X2019-03-01121627310.21514/1998-426X-2019-12-1-62-73569Evaluation of the dose reduction capabilities in digital radiography of the chest using contrast-detail phantomA. V. Vodovatov0Санкт-Петербургский научно-исследовательский институт радиационной гигиены имени профессора П.В. Рамзаева, Федеральная служба по надзору в сфере защиты прав потребителей и благополучия человекаAssessment of the quality of the images obtained using optimized (low-dose) protocols is the inherent part of the optimization in X-ray diagnostics. To perform the objective quantitative image quality assessment one can use dedicated test-objects, including several components for the simultaneous measurement of the different physical image characteristics (contrast and spatial resolution). The use of such test objects allows estimating and assessing the relations between the patient dose, parameter of the X-ray examination and image quality. That is especially important for the optimization of the digital radiographic examinations performed with automated exposure control. The aim of the current study was to evaluate the possibilities of the patient dose reduction using “contrast-detail” test-object for the digital radiography of the chest in posterior-anterior projection performed with automated exposure control. The study was performed in St-Petersburg Mariinsky hospital on a digital X-ray unit “ARC-Electron” with a flat-panel detector. The combination of a test-object and a tissue-equivalent phantom were imaged on a range of chest X-ray protocols: on a 60–150 kV tube voltage range with automated exposure control; and using fixed 90 kV tube voltage on a range of 2–100 mAs tube current-exposure time product. Dose-area product (cGy×cm2) was measured for each exposure; effective dose (mSv) was estimated for each exposure based on dose-area product. A dedicated software was developed for the automated image quality assessment. The results of the study indicate that the use of a high tube voltage (140–150 kV) with current automated exposure control settings would lead to 60% and 95% reduction of the dose-area product and effective dose, respectively, compared to the standard protocol. The adjustment of the current automated exposure control settings with the reduction of the tube current-exposure time product from 11,2 mAs to the 4,2 mAs for the tube voltage of 90 kV would lead to the reduction of both the dose-area product and effective dose up to a factor of three, compared to the standard protocol. For both scenarios image quality characteristics decreased by less than 15%. The proposed low-dose protocols are under the clinical approbation at Mariinsky hospital. The proposed method of image quality assessment and development of low-dose protocols is recommended for inclusion in the quality assurance program for the radiography examinations.https://www.radhyg.ru/jour/article/view/596медицинское облучениеконтроль качества изображениятест-объект «контраст-деталь»пространственное разрешениеконтрастоптимизация
spellingShingle A. V. Vodovatov
Evaluation of the dose reduction capabilities in digital radiography of the chest using contrast-detail phantom
Радиационная гигиена
медицинское облучение
контроль качества изображения
тест-объект «контраст-деталь»
пространственное разрешение
контраст
оптимизация
title Evaluation of the dose reduction capabilities in digital radiography of the chest using contrast-detail phantom
title_full Evaluation of the dose reduction capabilities in digital radiography of the chest using contrast-detail phantom
title_fullStr Evaluation of the dose reduction capabilities in digital radiography of the chest using contrast-detail phantom
title_full_unstemmed Evaluation of the dose reduction capabilities in digital radiography of the chest using contrast-detail phantom
title_short Evaluation of the dose reduction capabilities in digital radiography of the chest using contrast-detail phantom
title_sort evaluation of the dose reduction capabilities in digital radiography of the chest using contrast detail phantom
topic медицинское облучение
контроль качества изображения
тест-объект «контраст-деталь»
пространственное разрешение
контраст
оптимизация
url https://www.radhyg.ru/jour/article/view/596
work_keys_str_mv AT avvodovatov evaluationofthedosereductioncapabilitiesindigitalradiographyofthechestusingcontrastdetailphantom