Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis

Abstract Background To determine the repeatability and agreement using corneal tomography of a swept-source optical coherence tomographer (SS-OCT) compared to a rotating Scheimpflug camera (RSC) in normal eyes and keratoconus (KC) eyes. Methods This prospective repeatability analysis was performed a...

Full description

Bibliographic Details
Main Authors: Robert Herber, Janine Lenk, Lutz E. Pillunat, Frederik Raiskup
Format: Article
Language:English
Published: BMC 2022-05-01
Series:Eye and Vision
Subjects:
Online Access:https://doi.org/10.1186/s40662-022-00290-6
_version_ 1817989766108413952
author Robert Herber
Janine Lenk
Lutz E. Pillunat
Frederik Raiskup
author_facet Robert Herber
Janine Lenk
Lutz E. Pillunat
Frederik Raiskup
author_sort Robert Herber
collection DOAJ
description Abstract Background To determine the repeatability and agreement using corneal tomography of a swept-source optical coherence tomographer (SS-OCT) compared to a rotating Scheimpflug camera (RSC) in normal eyes and keratoconus (KC) eyes. Methods This prospective repeatability analysis was performed at the Department of Ophthalmology of University Hospital Carl Gustav Carus, Dresden, Germany. Forty-three normal and 57 KC eyes were enrolled in the study. Three consecutive measurements were performed by the same operator on each device. Corneal parameters of anterior and posterior corneal surface, such as simulated keratometry (SimK), as well as central and thinnest corneal thickness were evaluated. Repeatability and agreement were assessed by using the coefficient of repeatability and Bland-Altman analysis. Results The repeatability of anterior corneal parameters was comparable between RSC and SS-OCT in normal eyes (repeatability < 0.5 D). Repeatability was increased in mild and moderate KC for all parameters using both devices. In moderate KC, repeatability of Kmax was 1.33 D and 0.78 D for RSC and SS-OCT, respectively. Repeatability of posterior corneal parameters was consistently better for SS-OCT. Significant offsets and wide ranges of limits of agreement were found between the devices for SimK and corneal thickness values. Conclusions SS-OCT showed highly repeatable measurements of anterior and posterior corneal parameters in normal and KC eyes. Compared to RSC, the SS-OCT had a better repeatability of anterior corneal parameters in mild and moderate KC as well as posterior corneal parameters in all groups. Both devices should not be used interchangeably in the diagnostic process of patients. Trial registration NCT04251143 at Clinicaltrials.gov, registered on 12 March 2018, https://clinicaltrials.gov/ct2/show/NCT04251143?cond=Keratoconus&cntry=DE&city=Dresden&draw=2&rank=1
first_indexed 2024-04-14T00:50:05Z
format Article
id doaj.art-ce84d8b03b17497c89746640af4ca845
institution Directory Open Access Journal
issn 2326-0254
language English
last_indexed 2024-04-14T00:50:05Z
publishDate 2022-05-01
publisher BMC
record_format Article
series Eye and Vision
spelling doaj.art-ce84d8b03b17497c89746640af4ca8452022-12-22T02:21:49ZengBMCEye and Vision2326-02542022-05-019111210.1186/s40662-022-00290-6Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysisRobert Herber0Janine Lenk1Lutz E. Pillunat2Frederik Raiskup3Department of Ophthalmology, University Hospital Carl Gustav Carus, TU DresdenDepartment of Ophthalmology, University Hospital Carl Gustav Carus, TU DresdenDepartment of Ophthalmology, University Hospital Carl Gustav Carus, TU DresdenDepartment of Ophthalmology, University Hospital Carl Gustav Carus, TU DresdenAbstract Background To determine the repeatability and agreement using corneal tomography of a swept-source optical coherence tomographer (SS-OCT) compared to a rotating Scheimpflug camera (RSC) in normal eyes and keratoconus (KC) eyes. Methods This prospective repeatability analysis was performed at the Department of Ophthalmology of University Hospital Carl Gustav Carus, Dresden, Germany. Forty-three normal and 57 KC eyes were enrolled in the study. Three consecutive measurements were performed by the same operator on each device. Corneal parameters of anterior and posterior corneal surface, such as simulated keratometry (SimK), as well as central and thinnest corneal thickness were evaluated. Repeatability and agreement were assessed by using the coefficient of repeatability and Bland-Altman analysis. Results The repeatability of anterior corneal parameters was comparable between RSC and SS-OCT in normal eyes (repeatability < 0.5 D). Repeatability was increased in mild and moderate KC for all parameters using both devices. In moderate KC, repeatability of Kmax was 1.33 D and 0.78 D for RSC and SS-OCT, respectively. Repeatability of posterior corneal parameters was consistently better for SS-OCT. Significant offsets and wide ranges of limits of agreement were found between the devices for SimK and corneal thickness values. Conclusions SS-OCT showed highly repeatable measurements of anterior and posterior corneal parameters in normal and KC eyes. Compared to RSC, the SS-OCT had a better repeatability of anterior corneal parameters in mild and moderate KC as well as posterior corneal parameters in all groups. Both devices should not be used interchangeably in the diagnostic process of patients. Trial registration NCT04251143 at Clinicaltrials.gov, registered on 12 March 2018, https://clinicaltrials.gov/ct2/show/NCT04251143?cond=Keratoconus&cntry=DE&city=Dresden&draw=2&rank=1https://doi.org/10.1186/s40662-022-00290-6ANTERIONCorneaKeratoconusOCTPentacamScheimpflug
spellingShingle Robert Herber
Janine Lenk
Lutz E. Pillunat
Frederik Raiskup
Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis
Eye and Vision
ANTERION
Cornea
Keratoconus
OCT
Pentacam
Scheimpflug
title Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis
title_full Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis
title_fullStr Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis
title_full_unstemmed Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis
title_short Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis
title_sort comparison of corneal tomography using a novel swept source optical coherence tomographer and rotating scheimpflug system in normal and keratoconus eyes repeatability and agreement analysis
topic ANTERION
Cornea
Keratoconus
OCT
Pentacam
Scheimpflug
url https://doi.org/10.1186/s40662-022-00290-6
work_keys_str_mv AT robertherber comparisonofcornealtomographyusinganovelsweptsourceopticalcoherencetomographerandrotatingscheimpflugsysteminnormalandkeratoconuseyesrepeatabilityandagreementanalysis
AT janinelenk comparisonofcornealtomographyusinganovelsweptsourceopticalcoherencetomographerandrotatingscheimpflugsysteminnormalandkeratoconuseyesrepeatabilityandagreementanalysis
AT lutzepillunat comparisonofcornealtomographyusinganovelsweptsourceopticalcoherencetomographerandrotatingscheimpflugsysteminnormalandkeratoconuseyesrepeatabilityandagreementanalysis
AT frederikraiskup comparisonofcornealtomographyusinganovelsweptsourceopticalcoherencetomographerandrotatingscheimpflugsysteminnormalandkeratoconuseyesrepeatabilityandagreementanalysis