Simultaneous determination of colistin sulfate and tigecycline in human plasma by liquid chromatography-tandem mass spectrometry method

Abstract Objective To establish a high-performance liquid chromatography–tandem mass spectrometry method (HPLC–MS/MS) to simultaneously determine colistin sulfate and tigecycline in human plasma. Methods Polymyxin B1 internal standard (20 µL) was added into 200 µL of plasma sample. The samples were...

Full description

Bibliographic Details
Main Authors: Ying-Chao Ma, Xi-Kun Wu, Xiu-Ling Yang, Zhi-Qing Zhang
Format: Article
Language:English
Published: BMC 2024-01-01
Series:BMC Chemistry
Subjects:
Online Access:https://doi.org/10.1186/s13065-023-01109-8
Description
Summary:Abstract Objective To establish a high-performance liquid chromatography–tandem mass spectrometry method (HPLC–MS/MS) to simultaneously determine colistin sulfate and tigecycline in human plasma. Methods Polymyxin B1 internal standard (20 µL) was added into 200 µL of plasma sample. The samples were treated with methanol-5% trichloroacetic acid (50:50, V/V) solution, and the protein precipitation method was adopted for post-injection analysis. The chromatographic column was a Dikma C18 (4.6 mm × 150 mm, 5 μm). For the mobile phase, 0.1% formic acid in aqueous solution was used for phase A, 0.1% formic acid in acetonitrile solution for phase B, and gradient elution was also applied. The flow rate was 0.8 mL/min, the column temperature was 40 °C, and the injection volume was 10 µL; Electrospray ionization and multiple reaction ion monitoring were adopted and scanned by the HPLC–MS/MS positive ion mode. Results The endogenous impurities in the plasma had no interference in the determination of the analytes. There existed a good linear relationship of colistin sulfate within the range of 0.1–10 µg/mL (R2 = 0.9986), with the lower limit of quantification (LLOQ) of 0.1 µg/mL. There existed a good linear relationship of tigecycline within the range of 0.05–5 µg/ mL (R2 = 0.9987), with the LLOQ of 0.05 µg/mL. The intra- and inter-day relative standard deviations of colistin sulfate and tigecycline were both less than 15%, and the accuracy was between 88.21% and 108.24%. The extraction had good stability, the extraction recovery rate was 87.75–91.22%, and the matrix effect was 99.40–105.26%. Conclusion This study successfully established a method for simultaneously detecting colistin sulfate and tigecycline plasma concentrations. The method was simple, rapid, and highly sensitive and could be applied for therapeutic medication monitoring.
ISSN:2661-801X